Paper
26 June 2017 Experimental light scattering by small particles: system design and calibration
Göran Maconi, Ivan Kassamakov, Antti Penttilä, Maria Gritsevich, Edward Hæggström, Karri Muinonen
Author Affiliations +
Abstract
We describe a setup for precise multi-angular measurements of light scattered by mm- to μm-sized samples. We present a calibration procedure that ensures accurate measurements. Calibration is done using a spherical sample (d = 5 mm, n = 1.517) fixed on a static holder. The ultimate goal of the project is to allow accurate multi-wavelength measurements (the full Mueller matrix) of single-particle samples which are levitated ultrasonically.

The system comprises a tunable multimode Argon-krypton laser, with 12 wavelengths ranging from 465 to 676 nm, a linear polarizer, a reference photomultiplier tube (PMT) monitoring beam intensity, and several PMT:s mounted radially towards the sample at an adjustable radius. The current 150 mm radius allows measuring all azimuthal angles except for ±4° around the backward scattering direction. The measurement angle is controlled by a motor-driven rotational stage with an accuracy of 15’.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Göran Maconi, Ivan Kassamakov, Antti Penttilä, Maria Gritsevich, Edward Hæggström, and Karri Muinonen "Experimental light scattering by small particles: system design and calibration", Proc. SPIE 10329, Optical Measurement Systems for Industrial Inspection X, 103292S (26 June 2017); https://doi.org/10.1117/12.2269518
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Light scattering

Particle systems

Particles

Linear polarizers

Optical testing

Photomultipliers

Back to Top