Paper
1 June 1998 Modification of the Heidelberg retinal laser tomograph and flowmeter to allow measurements in supine patients
Karen Margaret Joos M.D., Chasidy Singleton, Jin-Hui Shen, Wensi Sun, Denis M. O'Day M.D.
Author Affiliations +
Proceedings Volume 3246, Ophthalmic Technologies VIII; (1998) https://doi.org/10.1117/12.309437
Event: BiOS '98 International Biomedical Optics Symposium, 1998, San Jose, CA, United States
Abstract
Purpose: To determine if the Heidelberg Retinal Laser Tomograph (HRT) and Flowmeter (HRF) can be modified to obtain images in supine patients. Methods: A mount was customized to securely attach the Heidelberg scanning head to an operating microscope stand. This mount was designed to allow rotation for viewing of the macula or optic nerve head region in either eye. The HRT was used to acquire 3 consecutive images of the optic nerve head in supine subjects to obtain a mean topographic image. The HRF was then used to obtain capillary flow measurements in supine subjects. Results: The optic nerve area in either the right or left eye can be safely and easily visualized with the modified Heidelberg system in supine patients for evaluation of optic nerve head topography or capillary flow. However, the configuration of the Heidelberg scanning head requires the images to be taken 180 degrees from the normal orientation of the scanning head to the patient's body. Therefore, the images are rotated 180 degrees from those taken in upright subjects. This must be considered when analyzing the data. Conclusion: The Heidelberg Retinal Tomograph and Flowmeter are capable of acquiring images in supine subjects. However, these images are rotated 180 degrees.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Karen Margaret Joos M.D., Chasidy Singleton, Jin-Hui Shen, Wensi Sun, and Denis M. O'Day M.D. "Modification of the Heidelberg retinal laser tomograph and flowmeter to allow measurements in supine patients", Proc. SPIE 3246, Ophthalmic Technologies VIII, (1 June 1998); https://doi.org/10.1117/12.309437
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Head

Optic nerve

Capillaries

Eye

Microscopes

Visualization

Blood circulation

Back to Top