Paper
19 February 2018 Measurement accuracy of a stressed contact lens during its relaxation period
Author Affiliations +
Proceedings Volume 10474, Ophthalmic Technologies XXVIII; 1047423 (2018) https://doi.org/10.1117/12.2290435
Event: SPIE BiOS, 2018, San Francisco, California, United States
Abstract
We examine the dioptric power and transmitted wavefront of a contact lens as it releases its handling stresses. Handling stresses are introduced as part of the contact lens loading process and are common across all contact lens measurement procedures and systems.

The latest advances in vision correction require tighter quality control during the manufacturing of the contact lenses. The optical power of contact lenses is one of the critical characteristics for users. Power measurements are conducted in the hydrated state, where the lens is resting inside a solution-filled glass cuvette. In a typical approach, the contact lens must be subject to long settling times prior to any measurements. Alternatively, multiple measurements must be averaged. Apart from potential operator dependency of such approach, it is extremely time-consuming, and therefore it precludes higher rates of testing.

Comprehensive knowledge about the settling process can be obtained by monitoring multiple parameters of the lens simultaneously. We have developed a system that combines co-aligned a Shack-Hartmann transmitted wavefront sensor and a time-domain low coherence interferometer to measure several optical and physical parameters (power, cylinder power, aberrations, center thickness, sagittal depth, and diameter) simultaneously. We monitor these parameters during the stress relaxation period and show correlations that can be used by manufacturers to devise methods for improved quality control procedures.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David C. Compertore and Filipp V. Ignatovich "Measurement accuracy of a stressed contact lens during its relaxation period", Proc. SPIE 10474, Ophthalmic Technologies XXVIII, 1047423 (19 February 2018); https://doi.org/10.1117/12.2290435
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Contact lenses

Manufacturing

Monochromatic aberrations

Wavefront sensors

Eye

Optical properties

Optics manufacturing

RELATED CONTENT

Interference test procedures for telescopic mirrors
Proceedings of SPIE (February 26 2004)
Aberration mapping for sight correction
Proceedings of SPIE (June 01 1998)
Improving metrology for micro-optics manufacturing
Proceedings of SPIE (November 10 2003)
Ophthalmic Lens Design With Splines
Proceedings of SPIE (May 09 1986)

Back to Top