Paper
5 November 2018 Calibration of probe misalignment in point-diffraction interferometer
Chao Wang, Daodang Wang, Zhongming Xie, Ming Kong, Jun Zhao, Lihua Lei, Yanhua Zeng, Wentao Zhang
Author Affiliations +
Abstract
The fiber point-diffraction interferometer provides a feasible method for the measurement of three-dimensional absolute displacement, in which the fiber point-diffraction sources generating coherent spherical reference waves are required to be parallel placed in the testing probe. However, the misalignment in the actual probe could introduce significant error in the displacement measurement result, and it also places high requirement on the adjustment of testing probe. A method based on phase difference in the axial displacement is proposed to calibrate the probe misalignment, and the numerical simulation is carried out to demonstrate the feasibility of the proposed method. The proposed method lowers the requirement on the processing of testing probe.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Chao Wang, Daodang Wang, Zhongming Xie, Ming Kong, Jun Zhao, Lihua Lei, Yanhua Zeng, and Wentao Zhang "Calibration of probe misalignment in point-diffraction interferometer", Proc. SPIE 10815, Optical Design and Testing VIII, 1081505 (5 November 2018); https://doi.org/10.1117/12.2501154
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Wavefronts

Charge-coupled devices

Interferometers

Spherical lenses

Numerical simulations

Reconstruction algorithms

RELATED CONTENT