Paper
14 September 1993 Laboratory measurements of sensitometry, MTF, veiling glare, Wiener spectrum and DQE for image intensifier tubes
Robert M. Gagne, Charles N. West, Robert F. Wagner, P. W. Quinn
Author Affiliations +
Abstract
Several different test protocols were used to measure the sensitometric response, the modulation transfer function (MTF), the veiling glare and the Wiener spectrum of two x-ray image intensifier (II) tubes. These data provided the means to calculate summary measures of imaging performance, i.e., the noise equivalent quanta, NEQ, or the detective quantum efficiency, DQE, as a function of spatial frequency. Results are presented to show the differences between an older versus a newer generation II tube, i.e., DQE equals 0.3 +/- 0.03 and 0.6 +/- 0.06 at 0.5 lp/mm, respectively. The eventual goal of this work is to achieve some consensus on methodology for these measurements and to validate the use of algorithmic observers in quantitating imaging performance in the clinical environment.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert M. Gagne, Charles N. West, Robert F. Wagner, and P. W. Quinn "Laboratory measurements of sensitometry, MTF, veiling glare, Wiener spectrum and DQE for image intensifier tubes", Proc. SPIE 1896, Medical Imaging 1993: Physics of Medical Imaging, (14 September 1993); https://doi.org/10.1117/12.154596
Lens.org Logo
CITATIONS
Cited by 13 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation transfer functions

Point spread functions

X-rays

Imaging systems

Image intensifiers

Spatial frequencies

Medical imaging

Back to Top