Paper
2 June 2003 Rotation-induced measurement error by a CD-SEM
Author Affiliations +
Abstract
As feature sizes shrink, metrology tools are challenged to deliver higher precision. To meet the demands of the semiconductor industry it is critical to determine and eliminate erros induced by metrology tools. It is also critical to investigate error-inducing factors neglected in the past. Image rotation is regularly experienced on a CD SEM, however error induced by it is assumed to be negligible and neglected. Experience shows an inline SEM often has image rotation of as much as ±26.3% of the edge thickness. For the range of rotation often experienced, experimental data fits the theoretical model suggested. Current tool monitoring techniques rely on pitch accuracy to monitor the day to day performance of the tool. Pitch is also used for tool calibration. However, pitch measurements are not significantly affected by image rotation hence using pitch as a standard for CD SEM monitoring is inadequate when dealing wiht a rotated image.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tejas K. Jhaveri and Rand Cottle "Rotation-induced measurement error by a CD-SEM", Proc. SPIE 5038, Metrology, Inspection, and Process Control for Microlithography XVII, (2 June 2003); https://doi.org/10.1117/12.485030
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scanning electron microscopy

Critical dimension metrology

Data modeling

Metrology

Semiconductors

Calibration

Manufacturing

Back to Top