Paper
21 June 2004 Use of adaptive optics for improved multiphoton imaging
John M. Girkin, Paul N. Marsh
Author Affiliations +
Abstract
We report on the integration of active optical elements in a multiphoton microscope to improve the resolution and overall image quality when imaging deeply into biological samples. Optical models were generated of sample systems and these have been compared with the performance of the complete imaging system. The active elements used were commercially available flexible membrane mirrors controlled by custom, home written, software. Significant improvements in image quality have been demonstrated using a range of optimisation routines based on the analysis of the images produced by the system, rather than with a wavefront sensor. A three-fold increase in the resolution 100μm into the sample was achieved.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John M. Girkin and Paul N. Marsh "Use of adaptive optics for improved multiphoton imaging", Proc. SPIE 5323, Multiphoton Microscopy in the Biomedical Sciences IV, (21 June 2004); https://doi.org/10.1117/12.525659
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Mirrors

Adaptive optics

Imaging systems

Objectives

Water

Image quality

Point spread functions

RELATED CONTENT


Back to Top