2 March 2006 An accurate scatter measurement and correction technique for cone beam breast CT imaging using scanning sampled measurement (SSM)technique
Author Affiliations +
Abstract
We developed and investigated a scanning sampled measurement (SSM) technique for scatter measurement and correction in cone beam breast CT imaging. A cylindrical polypropylene phantom (water equivalent) was mounted on a rotating table in a stationary gantry experimental cone beam breast CT imaging system. A 2-D array of lead beads, with the beads set apart about ~1 cm from each other and slightly tilted vertically, was placed between the object and x-ray source. A series of projection images were acquired as the phantom is rotated 1 degree per projection view and the lead beads array shifted vertically from one projection view to the next. A series of lead bars were also placed at the phantom edge to produce better scatter estimation across the phantom edges. Image signals in the lead beads/bars shadow were used to obtain sampled scatter measurements which were then interpolated to form an estimated scatter distribution across the projection images. The image data behind the lead bead/bar shadows were restored by interpolating image data from two adjacent projection views to form beam-block free projection images. The estimated scatter distribution was then subtracted from the corresponding restored projection image to obtain the scatter removed projection images. Our preliminary experiment has demonstrated that it is feasible to implement SSM technique for scatter estimation and correction for cone beam breast CT imaging. Scatter correction was successfully performed on all projection images using scatter distribution interpolated from SSM and restored projection image data. The resultant scatter corrected projection image data resulted in elevated CT number and largely reduced the cupping effects.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xinming Liu, Chris C. Shaw, Tianpeng Wang, Lingyun Chen, Mustafa C. Altunbas, and S. Cheenu Kappadath "An accurate scatter measurement and correction technique for cone beam breast CT imaging using scanning sampled measurement (SSM)technique", Proc. SPIE 6142, Medical Imaging 2006: Physics of Medical Imaging, 614234 (2 March 2006); https://doi.org/10.1117/12.656655
Lens.org Logo
CITATIONS
Cited by 26 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-ray computed tomography

Lead

Breast

Scatter measurement

Sensors

Mammography

X-rays

Back to Top