1 January 1986 Velocity Modulated Scanning Technique For The Elimination Of Ring Artifacts In X-Ray Computed Tomography
K. S. Hong, Orhan Nalcioglu, Zang-Hee Cho
Author Affiliations +
Abstract
Rotate-rotate fan beam configuration commonly used in the third generation x-ray computed tomography scanners is susceptible to ring artifacts resulting from non-uniform detector response. A method which incorporates the relative variation of the angular velocities of the x-ray tube and the detector array is proposed. Simulation results utilizing the new technique are presented and compared with the third and fourth generation scanners.
© (1986) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
K. S. Hong, Orhan Nalcioglu, and Zang-Hee Cho "Velocity Modulated Scanning Technique For The Elimination Of Ring Artifacts In X-Ray Computed Tomography", Proc. SPIE 0671, Physics and Engineering of Computerized Multidimensional Imaging and Processing, (1 January 1986); https://doi.org/10.1117/12.966679