Paper
2 September 2008 Pump-probe photothermal lens experiment for measuring the thermal diffusivity coefficient of organic solvents
A. Marcano O., N. Melikechi, A. Zelinskas, M. Benitez
Author Affiliations +
Abstract
We report on a pump-probe photothermal lens experiment aimed for the measurement of thermal diffusivity coefficient of organic samples in the CW regime. We show that when the probe beam is collimated and the pump beam is focused the time dependence of the signal does not depend on the sample position. This effect allows simple determination of the thermal diffusivity coefficient and its effective calibration using reference samples. We use this method to determine the thermal diffusivity of distilled water, methanol, chloroform, nitrobenzene and other organic solvents.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
A. Marcano O., N. Melikechi, A. Zelinskas, and M. Benitez "Pump-probe photothermal lens experiment for measuring the thermal diffusivity coefficient of organic solvents", Proc. SPIE 7049, Linear and Nonlinear Optics of Organic Materials VIII, 70490C (2 September 2008); https://doi.org/10.1117/12.794172