Paper
11 September 2009 Dual anode contact geometries for x-ray and gamma-ray spectroscopy and 3D localization
J. W. Martin, A. B. Garson III, Q. Li, K. Lee, M. Groza, V. Buliga, A. Burger, H. Krawczynski
Author Affiliations +
Abstract
We report on the continued development and testing of unique types of Cadmium Zinc Telluride (CZT) detectors. Using large volume (10×20×20 mm3) CZT crystals, we contact various "dual anode" detector designs. We incorporate a segmented cathode with five regions so that the charge on all seven contacts can be used to determine the energy and the 3-D interaction location of detected X-ray and gamma-ray photons. We describe the status of the detector development program, emphasize strengths and weaknesses of the different contact configurations, and discuss possible applications of Dual Anode Detectors in radiation detection applications.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. W. Martin, A. B. Garson III, Q. Li, K. Lee, M. Groza, V. Buliga, A. Burger, and H. Krawczynski "Dual anode contact geometries for x-ray and gamma-ray spectroscopy and 3D localization", Proc. SPIE 7449, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics XI, 744904 (11 September 2009); https://doi.org/10.1117/12.830115