Paper
11 February 2011 Stable absolute flow estimation with Doppler OCT based on virtual circumpapillary scans
Amardeep S. G. Singh, Christoph Kolbitsch, Tilman Schmoll, R. A. Leitgeb
Author Affiliations +
Proceedings Volume 7885, Ophthalmic Technologies XXI; 78850G (2011) https://doi.org/10.1117/12.874825
Event: SPIE BiOS, 2011, San Francisco, California, United States
Abstract
We propose an algorithm to extract the angles of vessels for the correction of flow measurements in circumpapillary Doppler OCT scans. Firstly, we register a volume to two reference scans in order to determine the physiologically correct structure of the volume. Then, vessels are segmented in the volume and the angles are calculated and stored in a look-up table. After having registered the circular scan to the volume by using the projection along the z-axis, the angles can be extracted from the look-up table. Repeatability measurements of flow parameters on 5 vessels of a healthy subject are presented.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Amardeep S. G. Singh, Christoph Kolbitsch, Tilman Schmoll, and R. A. Leitgeb "Stable absolute flow estimation with Doppler OCT based on virtual circumpapillary scans", Proc. SPIE 7885, Ophthalmic Technologies XXI, 78850G (11 February 2011); https://doi.org/10.1117/12.874825
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Retina

Doppler effect

Doppler tomography

Image segmentation

Arteries

Veins

Back to Top