26 May 2011 The effects of two noises from passive FTIR (Fourier transform infrared) spectroscopy system on the detection probability of stand-off hazardous compounds with passive Fourier transform infrared spectroscopy system using Michelson interferometer
Author Affiliations +
Abstract
An Fourier Transform Infra-Red(FTIR) spectrometry has played an significant role in a variety of fields in recent years. In particular, FTIR spectrometer technology has been adopted in passive remote sensing system to predict detection probabilities of stand-off hazardous compounds. There are three steps to detect hazardous compounds. Firstly, MODerate spectral resolution atmospheric TRANsmittance(MODTRAN) algorithm is used to calculate spectral radiances of background and atmosphere affected by hazardous compounds. It transfers a difference of spectral radiance between background and hazardous compounds existing region to FTIR spectroscopy system. Secondly, FTIR spectrometry system collects an interferogram which represents spectral radiances respective to given time intervals (reciprocal of wavenumber) and sends it to signal processing part. Lastly, the signal processing part performs Fourier Transformation on the interferogram and identifies the spectral radiance with reference data from gas library by using Spectral Angle Mapper(SAM) algorithm which results in visualizing the hazardous gases. However, there are some noises which affect the interferogram and the spectrum in practice. Specifically, there are two main noises which have critical effects on the interferogram and the spectrum by reducing its Signal to Noise Ratio(SNR) such as detector noise, jitter of moving mirror and optics misalignment. In this paper, a theory of the effects of the detector noise and the jitter of moving mirror on the interferogram and its demonstration through simulation are presented.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hee Kyung Ahn, Hong Jin Kong, and Kang Sup Shim "The effects of two noises from passive FTIR (Fourier transform infrared) spectroscopy system on the detection probability of stand-off hazardous compounds with passive Fourier transform infrared spectroscopy system using Michelson interferometer", Proc. SPIE 8024, Advanced Environmental, Chemical, and Biological Sensing Technologies VIII, 80240V (26 May 2011); https://doi.org/10.1117/12.882820
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal to noise ratio

FT-IR spectroscopy

Spectroscopy

Sensors

Mirrors

Interference (communication)

Michelson interferometers