18 March 2015 Physical performance testing of digital breast tomosynthesis
Takao Kuwabara, Kenji Yoshikawa
Author Affiliations +
Abstract
Digital breast tomosynthesis has become accepted in clinical use. It is important to physically evaluate a system to ensure that it is working at full performance. Non-linear reconstruction processing is proposed to improve interpretation of clinical images by enhancing the minute contrasts of breast tissue while suppressing metal artifacts. Because existing measuring methods assume a linear system, physical evaluation applied to images reconstructed with non-linear processing may result in unnatural values. We investigated the influence of different reconstruction methods on physical evaluations. We suggest using images reconstructed by back projection processing without a filter to ensure the device performance directly.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Takao Kuwabara and Kenji Yoshikawa "Physical performance testing of digital breast tomosynthesis", Proc. SPIE 9412, Medical Imaging 2015: Physics of Medical Imaging, 94123C (18 March 2015); https://doi.org/10.1117/12.2081079