1.IntroductionOptical coherence tomography (OCT), which dates back to 1991,1 provides cross-sectional views of biological tissues’ subsurface microstructures2,3 and has become a widely used diagnostic technique in the medical field due to its noninvasive nature. For example, ophthalmology has benefited from OCT, as it can image the retina and thus aid in the diagnosis.4 Different from common CCD imaging, OCT is an interferometric technique, and typically uses near-infrared laser light to penetrate into the scattering medium before capturing the backscattered optical waves for the final imaging.5 With the development of ultrahigh resolution OCT6 and Fourier-domain OCT,7,8 it is feasible to visualize biological tissues at a cellular level and up to the depth of nearly 1 mm below the surface with high sensitivity and image quality. Furthermore, the image acquisition speed of OCT systems has greatly improved with the development of high-speed sensors and tunable lasers with MHz scanning rates, which allow real-time imaging of in vivo tissues.8 One of the important challenges limiting high-speed OCT’s development is its unsatisfying image quality caused by speckle noise. Due to the coherence of optical waves, speckle noise arises under limited spatial-frequency bandwidth of the interference signals.5 The generation mechanism of OCT determines that the properties of speckle noise are related not only to the laser source, but also to the tissue’s structural properties,9–11 and thus results in nonuniform speckle noise over the entire image. Due to the significance of high precision in medical diagnosis, it is vital to remove speckle noise from OCT images for image quality enhancement. Much effort has been exerted for denoising OCT images, and various approaches are reported. These methods mainly fall into two categories, namely single-frame methods and multiframe methods. Single-frame methods often assume a prior model (either parametric or nonparametric) for the latent signal and noise, and then remove the noise determinatively or probabilistically from the input single image. Filtering is a widely used strategy, and there are many OCT denoising filters. For example, Ozcan et al.12 apply various digital filters for denoising OCT images, and the results indicate that the nonorthogonal wavelet filter together with the enhanced Lee and the adaptive Wiener filters can significantly reduce speckle noise. Based on the wavelet filter, Yue et al.13 utilize the iterative edge enhancement feature of nonlinear diffusion to improve the denoising results. Similarly, Zhang et al.14 use the nonlinear diffusion in the Laplacian pyramid domain to filter ultrasonic images. The Kovesi Nw filtering technique and the Laplacian pyramid nonlinear diffusion technique are unified together in Ref. 15 to remove both shot noise and speckle noise from OCT images. In addition to the filtering techniques, there are also some other approaches such as regularization and Bayesian inference. The work in Ref. 16 uses a regularization method to minimize the Csiszar’s I-divergence measurement, which would extrapolate additional details from the input noisy images to improve the visual effects. Wong et al.17 and Cameron et al.18 use a statistic Bayesian least square model to reduce OCT speckle noise in logarithmic space. Xie et al.19 take image contents into consideration and propose a salient structure extraction algorithm combined with an adaptive speckle suppression term to enhance ultrasound images. Different from the above filtering or statistic methods, based on sparse coding, the work in Ref. 20 learns an overcomplete dictionary from high signal-to-noise ratio (SNR) images and then utilizes this dictionary to reconstruct low-SNR OCT images and achieve significant noise suppression. In all, making use of the intrinsic redundancy within a single OCT frame can help noise removal to a large extent. Benefiting from the development of high-speed OCT imaging systems, the correlation between adjacent frames increases, and researchers gradually begin to use multiple OCT frames to attenuate speckle noise and propose various multiframe methods. Technically these methods could be further classified into hardware methods and algorithmic methods. The general idea of hardware methods is to change the parameters of OCT imaging systems to decorrelate speckle noise in different frames, and then compound these frames after image registration to get a noise-free OCT image. For example, Refs. 10 and 2122.–23 alter the angle of incident light while capturing different frames,24,25 change the detection angle of backreflected light,26 and change the frequency of the laser beam. Using the dynamic focus OCT facility, the work in Ref. 27 compares the performance of several spatial compounding methods and shows their respective pros and cons. The methods include median filtering, random weighted averaging, and random pixel selection. The main disadvantage of these hardware methods is the complex procedure of data acquisition, which would greatly increase the design complexity of OCT imaging systems.18 In addition, the performance of spatial compounding methods is not satisfactory. There are several recently reported algorithmic methods for denoising multiple OCT images that make use of the redundancy of latent sharp OCT images in the frequency domain. For example, the work in Ref. 28 performs the three-dimensional (3-D) curvelet transform to the volume data, then thresholds the coefficients, and finally does the inverse 3-D curvelet transform to realize the noise removal. Under the same framework, Mayer et al.4 choose the wavelet domain for coefficient thresholding. In spite of the promising performance, these denoising algorithms run the risk of losing crucial details by directly truncating the coefficients. Making use of both the intraframe and interframe redundancies of OCT volume data, this paper proposes an algorithmic multiframe optimization method to denoise OCT images. Within each single frame, an OCT image is statistically similar to a natural image and its pixel gradient map tends to be sparse. This serves as the intraframe prior. To avoid piecewise constant artifacts by simply using the total variation constraint, and considering the excellent detail-preservation performance of the low-rank prior in matrix completion29,30 and image reconstruction,31,32 we make use of the interframe redundancy by registering the OCT images along the image count dimension to form a low-rank volume. Subjecting the images to both the image formation model and the nonparametric bound constraint of nonuniform speckle noise, we build a preliminary nonconvex optimization model which jointly minimizes the rank of temporally registered OCT volume and forces the sparsity of its spatial gradient. To solve the above model, we first perform some mathematical transformations and approximations for convexification. Then, considering the superior convergence property of the augmented Lagrange multiplier (ALM) method33 for solving constrained optimization problems, as utilized in Ref. 34, we derive a numeric algorithm based on the ALM method to solve the convexified optimization model. Experiments on a pig eye, human retina, and orange OCT data show that our denoising technique could effectively reduce speckle noise while preserving rich details and that it exhibits superior performance to the other popular methods. The remainder of this paper is organized as follows: Sec. 2 sequentially describes the preprocessing operations—frame registration, noise variance estimation, model construction, and algorithm derivation. Then in Sec. 3, we apply our method to real-captured OCT images including pig eye, human retina, and orange data and compare our approach with several other popular methods in terms of both visual quality and quantitative evaluation. Finally, we conclude this paper with some conclusions and discussions in Sec. 4. 2.MethodIn this section, we explain the whole operation framework of our approach as diagramed in Fig. 1. Our method mainly includes three steps: (1) preprocessing step including pixel logarithm, frame registration, and noise variance estimation, (2) modeling step, and (3) solving step based on the ALM method. The implementation of each step is detailed in the following subsections. 2.1.PreprocessingIn this section, we conduct three preprocessing operations on the captured OCT frames including pixel logarithm, frame registration, and noise variance estimation. Due to the interference between the reference beam and the backscattering beam in OCT facilities, the speckle noise in OCT images is multiplicative5,9,10 and can be described as where denotes the captured value at location , while and , respectively, denote the ground truth and measurement noise at the same location. To convert the correlation between and from multiplication to addition, we take the logarithm to both sides of Eq. (1) and get In the following, we assume that all the variables have been logarithmically transformed.The second preprocessing operation is frame registration. Although the OCT equipment usually has a high capture speed, there always tends to be a slight misalignment among OCT image sequences in vivo imaging due to the object motion and other system factors.4 Here, we adopt the registration method in Ref. 4, where a powell optimizer is utilized for minimizing the sum of squared distances among multiple registered images. Specifically, the approach applies translations and rotations to warp the pixels describing the same tissue position in different frames to the same image coordinate. The third requisite preprocessing operation is the variance estimation of speckle noise. Considering the satisfying performance of the median absolute deviation (MAD) method for noise estimation,35 we here utilize the MAD method as described in Ref. 18. Due to similar tissue properties and light directions in the same neighborhood, we assume a uniform noise variance for each pixel within a small patch. Therefore, the MAD within a small neighborhood of pixel is first computed in logarithmic space as18 where denotes the median value over ’s neighborhood , and is the ’th neighboring pixel of . To make the estimated deviation more precise, we choose a larger neighborhood and calculate the local standard deviation of its subneighborhood . Then, we regard the mode of these as the preliminary noise deviation at position :Finally, to force the smoothness of noise variance among adjacent pixels, we conduct a cubic spline fitting process to amend and get the final standard deviation estimation of the OCT noise (the corresponding noise variance can be calculated as the square of the estimated standard deviation). Empirical studies in Ref. 18 state that the noise estimation performs best when the pixel numbers of and are and , respectively. 2.2.ModelingIn this section, we build our optimization model incorporating both of the interframe and intraframe priors. Suppose that there are frames in the OCT volume, and the pixel number of each frame is . We denote the temporally registered noisy OCT images, their latent sharp version, and the measurement noise as , , and , respectively. Mathematically, the image formation equation can be written as By representing each frame as a column vector, the dimensions of , , and are all . After frame registration, theoretically the entries in one specific row of should be exactly the same, as shown in Fig. 1. Therefore, we treat as a low-rank matrix and use the minimization of its nuclear norm , which calculates the sum of ’s singular values34 as the interframe prior constraint. According to the statistical studies,36,37 the adjacent pixels in natural images have similar intensities. Thus, the image gradient centers around zero and follows a heavy tailed distribution, i.e., the gradient of natural images is sparse. Although captured via a different imaging mechanism from the usual CCD imaging methods, the OCT images still follow similar statistics, and we impose the gradient sparsity of the latent OCT images as the intraframe prior. Specifically, the -norm, which counts the number of nonzero entries in a matrix, can model the sparseness quite well, thus we can minimize as the intraframe constraint, where is the gradient calculating operator. Adopting the same representation in Ref. 38, here we use matrix multiplication for the gradient calculation, namely , where and are, respectively, the horizontal and vertical gradient operators and are defined as the diagonal matrices of corresponding high-pass filters and . As mentioned before, speckle noise includes both the image information and the zero-mean noise. Since what we are most concerned with is the removal of the latter component, we treat the first component as a part of the latent sharp image and concentrate on attenuating the zero-mean noise. According to the three sigma rule, which indicates that nearly all (99.73%) of the instances of a random variable lie within three times the standard deviation from its mean, we can approximatively formulate the noise constraint as where is the standard deviation matrix whose dimension is . By introducing a non-negative matrix variable , we can transform the above inequality into an equality as in which is the entry-wise product, i.e., for two matrices and , .Based on the above notations, the optimization model for denoising is defined as with being a positive weighting parameter to balance different objective regulation terms.2.3.Solving the ModelIn this section, we derive our optimization algorithm based on the ALM method to solve the above model in Eq. (8). The model is obviously nonconvex, so we first conduct several convexification transformations to the model. As shown in Refs. 39 and 40, replacing the -norm with the -norm is one typical convexification transformation. Here, we replace with , where denotes the sum of the matrix entries’ absolute values. Further, it is hard to directly utilize the ALM method to solve the convexified objective function due to its high nonlinearity. To address this problem, we replace the variables whose nuclear norm or -norm needs minimization with two introduced auxiliary variables and . In addition, we pack as for computation simplicity, where . With the above substitutions, we can rewrite the model as Here, are supposed to be in theory.As stated before, we utilize the ALM method to solve the above model. This method adopts an iterative optimization strategy, and successively updates every variable within each iteration. In the following, we derive the updating rules for each variable. First, the augmented Lagrangian function of Eq. (9) is where denotes the inner product, defines the Lagrangian multipliers (in matrix form), and refers to the Frobenius norm that calculates the root of all the square entries’ sum in a matrix. Here, is a penalty parameter balancing the four equation constraints in Eq. (9) and follows the standard ALM updating rule as , where and are both user-defined parameters and indexes the iteration. The updating rules of the other variables including , , , , and are derived as follows.Optimize . By removing all the items irrelevant to in , we can get According to the ALM algorithm, we can get the updating rule of as where is the singular value decomposition (SVD) of , andSimilarly, keeping only the items related to in yields and we can get the updating rule of asOptimize and . By keeping only the items related to , is simplified as and the partial derivative of with respect to isSimilarly, the partial derivative of with respect to is Obviously, it is hard to get the closed-form solution to either or , so we resort to the gradient descent method to approximatively update these two variables as Here, stands for the learning rate.Optimize . The derivative of with respect to is Under the non-negative assumption on , we can get its updating rule as All the algorithm parameters are set as follows: , , , , , and . These constant parameters are statistically set by testing the algorithm on a series of OCT data to obtain the best denoising performance and least running time. In addition, all the parameters are fixed across all the experiments in this paper. We take the average of input preregistered frames as the initialization of the denoised frame. For more clarity, the entire iterative algorithm based on the above derivations is summarized in Algorithm 1. Algorithm 1The proposed multiframe algorithm for OCT denoising.
2.4.Evaluation CriteriaTo quantitatively evaluate the denoising performance of different approaches, we utilize three widely used image quality criteria including the peak SNR (PSNR), the structure similarity (SSIM),41 and the figure of merit (FOM)42,43 as evaluation metrics.
3.Experimental ResultsIn this section, we test our denoising approach on three OCT image sets and compare our denosing results with those of several previously reported popular methods. In addition to the visual comparison, we also perform quantitative comparison in terms of PSNR, SSIM, FOM, and running time. 3.1.Results on the Pig Eye Data and Performance ComparisonIn this experiment, we use the public pig eye OCT dataset in Ref. 45, which is also used as the source data in Ref. 4. The dataset is acquired using a Spectralis HRA & OCT (Heidelberg Engineering) to scan a pig eye in the high-speed mode with 768 A-scans. There are a total of 455 images (each frame contains ) in the dataset including 35 sets, and 13 frames sharing the same imaging position are included in each set. All the 35 positions correspond to a complete 0.384-mm shift in the transversal direction. For each frame, the pixel spacing is in the axial direction and in the transversal direction. To assess the quality of the recovered images, we need a noise-free benchmark image for reference. However, due to the high-imaging speed, the captured images’ SNR is very low. Therefore, we utilize the same technique as in Ref. 4, in which the averaged image of all the 455 preregistered frames is used as the latent noise-free image. Since the proposed method is multiframe based, we first investigate the effect of the most important parameter—the number of input frames—on our algorithm. Fixing all the other parameters, we run a MATLAB implementation of our proposed method on an Intel E7500 2.93 GHz CPU computer with 4GB RAM and 64 bit Windows 7 system and compare the performance of different numbers of input frames, ranging from 2 to 13. From the result in Fig. 2, we can see that our algorithm still works using only 2 frames, and the reconstruction quality gradually improves using an increasing number of frames from 2 to 8. This trend is much less obvious with more than 8 frames. Based on this observation, we use 8 input frames in the following experiments to compare our approach with other methods. Note that there exists some slight drop when the frame number grows larger than 8. This is because the frame registration gets more difficult for a longer sequence and thus slightly hampers the reconstruction. Next, we run our algorithm and the other four popular denoising methods on the OCT dataset for comparison. The four methods include the complex diffusion method,46 the Bayesian method,17 the non-stationary speckle compensation method (NSC),18 and the multiframe wavelet OCT denoising method.4 To validate the superior effectiveness of our approach, we compare all the algorithms’ performances visually and quantitatively. What should be noticed is that the complex diffusion method, the Bayesian method, and NSC are all single-frame methods, so we take the average image of the registered input frames as their single-frame input. In addition, the first two methods assume a spatially invariant noise parameter (standard deviation). Correspondingly, we use the maximum in the estimated deviation matrix as the input standard deviation of noise. By random selection, the serial numbers of the input 8 sequential frames are from “35_6” to “35_13.” The recovered images are shown in Fig. 3, where only the first frame is presented for each method. We can see that the recovered images of the anisotropic diffusion method, the Bayesian method, and NSC still contain undesired noise, which largely degenerates the image quality and makes these three methods less competitive than the other two techniques. On the whole, the multiframe wavelet method and our method are both superior to the three single-frame approaches. This superiority is attributed to the fact that the multiframe methods utilize the interframe correlation and redundancy, which would offer more information of the latent noise-free data. Comparing the results of these two multiframe methods, we can see that in the smooth regions, the wavelet method leaves out more noise than our method. In the textured regions, the result of our method maintains a higher color contrast which would improve the visual effects. A closer comparison is presented in the closeups. For example, in the green-rectangle-highlighted region, the wavelet method nearly blurs out the details of the white spot on the left side, while our method still contains gray value changes which would provide important information for diagnosis. Numerical assessments are shown in Table 1. For the entire image, we can see that our method could raise the noisy images’s PSNR from 17.19 to 31.74 dB and the SSIM from 0.13 to 0.91. In terms of all the three evaluation criteria including PSNR, SSIM, and FOM, our method consistently has an advantage over the other methods. Comparing the two multiframe methods, namely the multiframe wavelet method and our method, we can see that our approach is superior in PSNR and SSIM by around 1 dB and 0.1, respectively. Our superior performance is mainly attributed to two factors: the combined constraints from both the interframe and intraframe priors and the good convergence of the derived algorithm. Comparing the numerical evaluation results of the two selected regions of interests, we can also see the clearer advantage of our method over the other methods. What is more, the running time comparison is also provided in Table 1 (the noise estimation time is also included for all the four algorithms). We run the MATLAB codes of our algorithm and the other three methods except for NSC on our computer, while the running time of NSC is provided by its proposers who run their MATLAB and C++ implementation on a different platform. We can see that our approach needs around 36 s to process one frame and is of similar efficiency to the Bayesian method which is the fastest algorithm among the popular methods except for NSC. Table 1Quantitative comparisons among different denoising methods.
3.2.Analyzing Optical Coherence Tomography Images of Human RetinaTo test the practical denoising effectiveness of our method, we conduct a denoising experiment on human retinal OCT images. We use the same public dataset as that used in Ref. 47, which is acquired by an spectral domain (SD)-OCT imaging system from Bioptigen Inc. with axial resolution, 500 A-scans per B-scan, and 5 azimuthally repeated B-scans in each volume. Referring to the processing progress described in Sec. 2.1, we first register the OCT frames and then use different methods to denoise these frames. Considering that the anisotropic diffusion method, the Bayesian method, and NSC leave too much noise on the recovered images, and thus show little competitiveness compared with the other two multiframe methods, here we only present the denoising results of the multiframe wavelet method and our method for clearer comparison. The results are shown in Fig. 4, which exhibit a similar performance ranking to that of the pig eye data. Comparing the denoising results produced by the multiframe wavelet method and our proposed method carefully, we can see that the result of the wavelet method contains undesired edge burrs, while our result presents clearer layer boundaries (such as the horizontal layer edges in the two selected regions), which would greatly help in follow-up analysis of the denoised images, such as OCT layer segmentation and diagnosis. 3.3.Phantom Study Using the Orange DataTo further validate the advantages of the proposed method in preserving image details while removing noise, we use the same orange dataset as in Ref. 48 to conduct a phantom study. The dataset is acquired by an SD-OCT system with the axial resolution being and the transversal resolution being . This data include 100 aligned frames and contain many thin structures corresponding to the tangerine pith. Similar to the experiment on human retina, considering the multiframe methods largely outperform the single-frame methods in both noise attenuation and detail preservation, here we only show the results of the multiframe wavelet method as well as ours. We still use 8 input frames and the results are shown in Fig. 5. For performance comparison between the two methods, we consider the average of 100 frames as benchmark, as shown in Fig. 5(c). From the results in Fig. 5(d), we can clearly see that some fine structural details are smoothed out by the wavelet method, especially in the region highlighted by the red rectangle. This is because the wavelet method truncates the wavelet coefficients of the input frames to attenuate noise, and some entries describing fine structures tend to be treated as speckle noise and removed. On the contrary, our method can preserve the thin structures quite well during noise suppression, as shown in Fig. 5(e). Compared against the results in Fig. 5(c), one can see clearly that our approach is more competitive in such cases with rich thin structures. 4.Conclusions and Discussions4.1.Summary and ConclusionsIn this study, we propose a multiframe OCT denoising method utilizing the constraints from both interframe and intraframe priors. Specifically, the interframe prior refers to the low rank of registered OCT frames and the intraframe prior is the sparsity of the image gradient. Benefiting from the proper convexification transformations and usage of ALM, the derived algorithm converges well on different data. In addition, by incorporating a nonparametric and nonuniform noise description, our approach is applicable for different noise models. On the adopted benchmark data, our approach could improve the OCT image’s quality by raising the image’s PSNR from 17.19 to 31.74 dB and SSIM from 0.12 to 0.91, in around 36 s for each frame. The comparisons with the other four popular methods on the three datasets reveal that our method has advantages mainly in two aspects: (1) being able to attenuate speckle noise effectively and preserve crucial image details; (2) with efficiency comparable with the reported fastest approaches. Such a high performance of the proposed method is mainly the result of the combined prior modeling and effective optimization algorithm. 4.2.Limitations and Future ExtensionsThe performance of our method depends on the registration accuracy because the low-rank prior in the objective function does not hold for an unaligned frame stack. This is also a challenge for other multiframe denoising methods and needs to be addressed by the progress of noise robust matching techniques. In addition, a larger frame number is favorable to take advantage of the low-rank prior. Therefore, one needs to set the system’s frame rate to balance the noise level and the number of available frames in practical applications. Besides, the widely used anisotropic total variation is utilized as the intraframe prior, which penalizes the diagonal gradients more significantly than the horizontal and vertical ones. This means that the utilized nonuniform constraint on image intensity changes along different directions, which may introduce undesired artifacts in the denoised images. Thus, exploring an isotropic intraframe prior would be one of our future extensions. In addition, the running time of the proposed method can be further reduced. There are two very time-consuming calculations in the current algorithm including the requisite relatively large iteration number and the SVD. To decrease the iterations, we can utilize an accelerated gradient descent scheme to speed up the method such as using adaptive learning rates instead of a constant rate. For a faster SVD, we can replace the currently adopted default implementation in MATLAB with an accelerated SVD algorithm such as the block Lanczos algorithm.49 Furthermore, we can also implement the proposed algorithm with C or C++ and use graphics processing unit accelerating techniques48,50 to further speed up the method. AcknowledgmentsThis work was supported by the National Natural Science Foundation of China, Nos. 61171119, 61120106003, and 61327902. The authors thank Professor Yong Huang and Professor Guohai Situ for their valuable suggestions and help, and also wish to thank Professor Alexander Wong for providing their running results to help with the experiments. ReferencesD. Huang et al.,
“Optical coherence tomography,”
Science, 254
(5035), 1178
–1181
(1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar
J. M. Schmitt,
“Optical coherence tomography (OCT): a review,”
IEEE J. Sel. Top. Quantum Electron., 5
(4), 1205
–1215
(1999). http://dx.doi.org/10.1109/2944.796348 IJSQEN 1077-260X Google Scholar
J. Welzel,
“Optical coherence tomography in dermatology: a review,”
Skin Res. Technol., 7
(1), 1
–9
(2001). http://dx.doi.org/10.1034/j.1600-0846.2001.007001001.x 0909-752X Google Scholar
M. A. Mayer et al,
“Wavelet denoising of multiframe optical coherence tomography data,”
Biomed. Opt. Express, 3
(3), 572
–589
(2012). http://dx.doi.org/10.1364/BOE.3.000572 BOEICL 2156-7085 Google Scholar
J. M. Schmitt, S. Xiang and K. M. Yung,
“Speckle in optical coherence tomography,”
J. Biomed. Opt., 4
(1), 95
–105
(1999). http://dx.doi.org/10.1117/1.429925 JBOPFO 1083-3668 Google Scholar
M. Wojtkowski et al.,
“Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,”
Opt. Express, 12
(11), 2404
–2422
(2004). http://dx.doi.org/10.1364/OPEX.12.002404 OPEXFF 1094-4087 Google Scholar
A. Fercher et al.,
“Measurement of intraocular distances by backscattering spectral interferometry,”
Opt. Commun., 117
(1C2), 43
–48
(1995). http://dx.doi.org/10.1016/0030-4018(95)00119-S OPCOB8 0030-4018 Google Scholar
M. Wojtkowski et al.,
“In vivo human retinal imaging by Fourier domain optical coherence tomography,”
J. Biomed. Opt., 7
(3), 457
–463
(2002). http://dx.doi.org/10.1117/1.1482379 JBOPFO 1083-3668 Google Scholar
J. Schmitt and A. Knüttel,
“Model of optical coherence tomography of heterogeneous tissue,”
J. Opt. Soc. Am. A, 14
(6), 1231
–1242
(1997). http://dx.doi.org/10.1364/JOSAA.14.001231 JOAOD6 0740-3232 Google Scholar
M. Bashkansky and J. Reintjes,
“Statistics and reduction of speckle in optical coherence tomography,”
Opt. Lett., 25
(8), 545
–547
(2000). http://dx.doi.org/10.1364/OL.25.000545 OPLEDP 0146-9592 Google Scholar
B. Karamata et al.,
“Speckle statistics in optical coherence tomography,”
J. Opt. Soc. Am. A, 22
(4), 593
–596
(2005). http://dx.doi.org/10.1364/JOSAA.22.000593 JOAOD6 0740-3232 Google Scholar
A. Ozcan et al.,
“Speckle reduction in optical coherence tomography images using digital filtering,”
J. Opt. Soc. Am. A, 24
(7), 1901
–1910
(2007). http://dx.doi.org/10.1364/JOSAA.24.001901 JOAOD6 0740-3232 Google Scholar
Y. Yue et al.,
“Nonlinear multiscale wavelet diffusion for speckle suppression and edge enhancement in ultrasound images,”
IEEE Trans. Med. Imaging, 25
(3), 297
–311
(2006). http://dx.doi.org/10.1109/TMI.2005.862737 ITMID4 0278-0062 Google Scholar
F. Zhang et al.,
“Nonlinear diffusion in Laplacian pyramid domain for ultrasonic speckle reduction,”
IEEE Trans. Med. Imaging, 26
(2), 200
–211
(2007). http://dx.doi.org/10.1109/TMI.2006.889735 ITMID4 0278-0062 Google Scholar
M. Gargesha et al.,
“Denoising and 4D visualization of OCT images,”
Opt. Express, 16
(16), 12313
–12333
(2008). http://dx.doi.org/10.1364/OE.16.012313 OPEXFF 1094-4087 Google Scholar
D. L. Marks, T. S. Ralston and S. A. Boppart,
“Speckle reduction by I-divergence regularization in optical coherence tomography,”
J. Opt. Soc. Am. A, 22
(11), 2366
–2371
(2005). http://dx.doi.org/10.1364/JOSAA.22.002366 JOAOD6 0740-3232 Google Scholar
A. Wong et al.,
“General Bayesian estimation for speckle noise reduction in optical coherence tomography retinal imagery,”
Opt. Express, 18
(8), 8338
–8352
(2010). http://dx.doi.org/10.1364/OE.18.008338 OPEXFF 1094-4087 Google Scholar
A. Cameron et al.,
“Stochastic speckle noise compensation in optical coherence tomography using non-stationary spline-based speckle noise modelling,”
Biomed. Opt. Express, 4
(9), 1769
–1785
(2013). http://dx.doi.org/10.1364/BOE.4.001769 BOEICL 2156-7085 Google Scholar
J. Xie et al.,
“Boundary enhancement and speckle reduction for ultrasound images via salient structure extraction,”
IEEE Trans. Biomed. Eng., 53
(11), 2300
–2309
(2006). http://dx.doi.org/10.1109/TBME.2006.878088 IEBEAX 0018-9294 Google Scholar
L. Fang et al.,
“Sparsity based denoising of spectral domain optical coherence tomography images,”
Biomed. Opt. Express, 3
(5), 927
–942
(2012). http://dx.doi.org/10.1364/BOE.3.000927 BOEICL 2156-7085 Google Scholar
N. Iftimia, B. E. Bouma and G. J. Tearney,
“Speckle reduction in optical coherence tomography by path length encoded angular compounding,”
J. Biomed. Opt., 8
(2), 260
–263
(2003). http://dx.doi.org/10.1117/1.1559060 JBOPFO 1083-3668 Google Scholar
L. Ramrath et al.,
“Towards multi-directional OCT for speckle noise reduction,”
Medical Image Computing and Computer-Assisted Intervention, 815
–823 Springer, New York
(2008). Google Scholar
M. Hughes, M. Spring and A. Podoleanu,
“Speckle noise reduction in optical coherence tomography of paint layers,”
Appl. Opt., 49
(1), 99
–107
(2010). http://dx.doi.org/10.1364/AO.49.000099 APOPAI 0003-6935 Google Scholar
A. Desjardins et al.,
“Speckle reduction in OCT using massively-parallel detection and frequency-domain ranging,”
Opt. Express, 14
(11), 4736
–4745
(2006). http://dx.doi.org/10.1364/OE.14.004736 OPEXFF 1094-4087 Google Scholar
A. Desjardins et al.,
“Angle-resolved optical coherence tomography with sequential angular selectivity for speckle reduction,”
Opt. Express, 15
(10), 6200
–6209
(2007). http://dx.doi.org/10.1364/OE.15.006200 OPEXFF 1094-4087 Google Scholar
M. Pircher et al.,
“Speckle reduction in optical coherence tomography by frequency compounding,”
J. Biomed. Opt., 8
(3), 565
–569
(2003). http://dx.doi.org/10.1117/1.1578087 JBOPFO 1083-3668 Google Scholar
M. R. Avanaki et al.,
“Spatial compounding algorithm for speckle reduction of dynamic focus OCT images,”
IEEE Photonics Technol. Lett., 25
(15), 1439
–1442
(2013). http://dx.doi.org/10.1109/LPT.2013.2266660 IPTLEL 1041-1135 Google Scholar
Z. Jian et al.,
“Three-dimensional speckle suppression in optical coherence tomography based on the curvelet transform,”
Opt. Express, 18
(2), 1024
–1032
(2010). http://dx.doi.org/10.1364/OE.18.001024 OPEXFF 1094-4087 Google Scholar
E. Candès and B. Recht,
“Exact matrix completion via convex optimization,”
Commun. ACM, 55
(6), 111
–119
(2012). http://dx.doi.org/10.1145/2184319 CACMA2 0001-0782 Google Scholar
Y. Deng et al.,
“Noisy depth maps fusion for multiview stereo via matrix completion,”
IEEE J. Sel. Top. Signal Process., 6
(5), 566
–582
(2012). http://dx.doi.org/10.1109/JSTSP.2012.2195472 1932-4553 Google Scholar
H. Ji et al.,
“Robust video denoising using low rank matrix completion,”
in IEEE Conf. Computer Vision and Pattern Recognition (CVPR),
1791
–1798
(2010). Google Scholar
J. Suo et al.,
“Joint non-Gaussian denoising and superresolving of raw high frame rate videos,”
IEEE Trans. Image Process., 23
(3), 1154
–1168
(2014). http://dx.doi.org/10.1109/TIP.2014.2298976 IIPRE4 1057-7149 Google Scholar
D. P. Bertsekas, Constrained Optimization and Lagrange Multiplier Methods, Academic Press, Waltham, Massachusetts
(1982). Google Scholar
Z. Lin, M. Chen and Y. Ma,
“Linearized alternating direction method with adaptive penalty for low-rank representation,”
612
–620
(2010). Google Scholar
A. Leigh et al.,
“Comprehensive analysis on the effects of noise estimation strategies on image noise artifact suppression performance,”
in IEEE Int. Symp. Multimedia (ISM),
97
–104
(2011). Google Scholar
T. S. Cho et al.,
“A content-aware image prior,”
in IEEE Conf. Computer Vision and Pattern Recognition,
169
–176
(2010). Google Scholar
T. S. Cho et al.,
“Image restoration by matching gradient distributions,”
IEEE Trans. Pattern Anal., 34
(4), 683
–694
(2012). http://dx.doi.org/10.1109/TPAMI.2011.166 ITPIDJ 0162-8828 Google Scholar
F. Heide et al.,
“High-quality computational imaging through simple lenses,”
ACM Trans. Graphics., 32
(5), 149:1
–149:14
(2013). http://dx.doi.org/10.1145/2516971 ATGRDF 0730-0301 Google Scholar
Y. Deng, Q. Dai and Z. Zhang,
“An overview of computational sparse models and their applications in artificial intelligence,”
Artif. Intell. Evol. Comput. and Metaheuristics, 345
–369 Springer, Berlin Heidelberg
(2013). Google Scholar
H. Lee et al.,
“Efficient sparse coding algorithms,”
Advances in Neural Information Processing Systems, 801
–808 NIPS, San Diego, California
(2006). Google Scholar
Z. Wang et al.,
“Image quality assessment: from error visibility to structural similarity,”
IEEE Trans. Image Process., 13
(4), 600
–612
(2004). http://dx.doi.org/10.1109/TIP.2003.819861 IIPRE4 1057-7149 Google Scholar
Y. Yu and S. Acton,
“Speckle reducing anisotropic diffusion,”
IEEE Trans. Image Process., 11
(11), 1260
–1270
(2002). http://dx.doi.org/10.1109/TIP.2002.804276 IIPRE4 1057-7149 Google Scholar
Y. Yue et al.,
“Nonlinear multiscale wavelet diffusion for speckle suppression and edge enhancement in ultrasound images,”
IEEE Trans. Med. Imaging, 25
(3), 297
–311
(2006). http://dx.doi.org/10.1109/TMI.2005.862737 ITMID4 0278-0062 Google Scholar
W. K. Pratt, Digital Image Processing, 2ndWiley, New York
(1991). Google Scholar
M. A. Mayer et al.,
“Image denoising algorithms archive,”
(2013) http://www5.cs.fau.de/en/research/software/idaa/ ( September ). 2013). Google Scholar
R. Bernardes et al.,
“Improved adaptive complex diffusion despeckling filter,”
Opt. Express, 18
(23), 24048
–24059
(2010). http://dx.doi.org/10.1364/OE.18.024048 OPEXFF 1094-4087 Google Scholar
L. Fang et al.,
“Fast acquisition and reconstruction of optical coherence tomography images via sparse representation,”
IEEE Trans. Med. Imaging, 32
(11), 2034
–2049
(2013). http://dx.doi.org/10.1109/TMI.2013.2271904 ITMID4 0278-0062 Google Scholar
D. Xu, Y. Huang and J. U. Kang,
“Real-time compressive sensing spectral domain optical coherence tomography,”
Opt. Lett., 39
(1), 76
–79
(2014). http://dx.doi.org/10.1364/OL.39.000076 OPLEDP 0146-9592 Google Scholar
S. Wei and Z. Lin,
“Accelerating iterations involving eigenvalue or singular value decomposition by block Lanczos with warm start,”
(2010). Google Scholar
Y. Huang, X. Liu and J. U. Kang,
“Real-time 3D and 4D Fourier domain Doppler optical coherence tomography based on dual graphics processing units,”
Biomed. Opt. Express, 3
(9), 2162
–2174
(2012). http://dx.doi.org/10.1364/BOE.3.002162 BOEICL 2156-7085 Google Scholar
BiographyLiheng Bian is currently pursuing a PhD degree with the Department of Automation, Tsinghua University, Beijing, China. He received his BE degree in electronic information and engineering from Xi’dian University, Xi’an, China, in 2013. His research interests include computational imaging and optical bioimaging. Jinli Suo received her BS degree in computer science from Shandong University, Shandong, China, in 2004, and the PhD degree from the Graduate University of the Chinese Academy of Sciences, Beijing, China, in 2010. She is currently a lecturer with the Department of Automation, Tsinghua University, Beijing. Her research interests mainly include computer vision, computational photography, and statistical learning. Feng Chen received his BS and MS degrees in automation from St. Petersburg State Polytechnical University, St. Petersburg, Russia, in 1994 and 1996, respectively, and the PhD degree in automation from Tsinghua University, Beijing, China, in 2000. He is a professor at Tsinghua University. His research interests are probabilistic graphical models and computer vision. Qionghai Dai received his PhD degree in automation from Northeastern University, Shenyang, China, in 1996. He has been a faculty member since 1997 and a professor since 2005 with the Department of Automation, Tsinghua University, Beijing, China. He has published more than 120 conference and journal papers, and holds 67 patents. His current research interests include computational photography, high resolution microscopy, compressed sensing and vision. |