12 October 2022 Detection and differentiation of semi-transparent materials simulating biological structures using optical coherence tomography: a phantom study
Author Affiliations +
Abstract

Significance

Lymphatic and peripheral nervous system imaging is of prime importance for monitoring various important pathologic processes including cancer development and metastasis, and response to therapy.

Aim

Optical coherence tomography (OCT) is a promising approach for this imaging task but is challenged by the near-transparent nature of these structures. Our aim is to detect and differentiate semi-transparent materials using OCT texture analysis, toward label-free neurography and lymphography.

Approach

We have recently demonstrated an innovative OCT texture analysis-based approach that used speckle statistics to image lymphatics and nerves in-vivo that does not rely on negative contrast. However, these two near-transparent structures could not be easily differentiated from each other in the texture analysis parameter space. Here, we perform a rigorous follow-up study to improve upon this differentiation in controlled phantoms mimicking the optical properties of these tissues.

Results

The results of the three-parameter Rayleigh distribution fit to the OCT images of six types of tissue-mimicking materials varying in transparency and biophysical properties demonstrate clear differences between them, suggesting routes for improved lymphatics-nerves differentiation.

Conclusions

We demonstrate a novel OCT texture analysis-based lymphatics-nerves differentiation methodology in tissue-simulating phantoms. Future work will focus on longitudinal in-vivo lymphangiography and neurography in response to cancer therapeutics toward adaptive personalized medicine.

1.

Introduction

Lymphatic vessels form a network responsible for transporting a colorless, watery fluid called lymph, consisting primarily of proteins and interstitial fluid, from tissue back into the bloodstream.1 Along with blood vessels, lymphatic vessels play an important role in the metastasis of cancer cells, which is the primary cause of death in cancer patients.2 The peripheral nervous system can also serve as a conduit for invading cancer cells, facilitating metastasis, and can modulate activity and growth of the tumors it innervates.3 In humans, the presence of metastasized tumor cells in lymph nodes is a strong determinant of a poor prognosis, thus emphasizing their importance in oncology.1,2

In-vivo imaging of lymphatic vessels and nerves is challenging because of the translucent/transparent nature of these structures.4,5 Typically, lymphatic vessels are detected in-vivo by the interstitial injection of an exogenous contrast agent which is preferentially absorbed into the lymphatic vessels as they uptake interstitial fluid. These dyes are easily observable by various imaging modalities ranging from low-resolution and high penetration depth magnetic resonance imaging and computed tomography, to high-resolution but low penetration depth optical imaging such as confocal fluorescence microscopy.4,6,7 Furthermore, visualization of lymphatic vessels with such methods is mostly confined to the vicinity of the contrast agent’s injection site.4

Optical coherence tomography (OCT) is a non-invasive, contrast-agent-free, volumetric imaging technique that has already shown promise toward in-vivo lymphography,810 and neurography.11,12 Owing to the aforementioned transparency of these structures, most OCT approaches rely on negative-contrast intensity-thresholding9,1315 for detection,8,9 whereby the absence of a signal in a region surrounded by an otherwise signal-rich region indicates a lymphatic vessel. However, as the signal-noise-ratio (SNR) drops with increasing imaging depth, it becomes progressively harder to differentiate absence-of-signal lymphatics from background noise. For instance, several studies have successfully implemented such methods for lymphography using swept-source OCT (ss-OCT) systems only up to a depth of 0.5  mm.9,14,15 Additionally, lymphatics have been detected utilizing vesselness models based on Hessian filters applied to B-scans, albeit with limited resolution and imaging depth challenges similar to negative-contrast methods.15,16 Lastly, another method applies depth-resolved attenuation coefficient distributions to visualize lymphatic networks with improved contrast and resolution when compared to intensity thresholding and Hessian filtering techniques,6,13,17 but the chances of false-positive lymphatic detections are higher. Utilizing OCT for non-invasive nerve visualization is a field yet to be thoroughly investigated, although there have been promising initial feasibility reports on OCT peripheral nerve identification.1012 In our previous studies,10,18 we presented a novel methodology for visualizing lymphatics (and somewhat unexpectedly also detecting nerves), based on texture analysis of spatial speckle statistics. However, the newly in-vivo detected lymphatics and nerves were not clearly differentiated.

In this controlled phantom study, we thus continue developing our innovative methodology to distinguish lymphatics from nerves. We perform a rigorous speckle statistical analysis based on three parameters of the Rayleigh probability distribution function (PDF) in phantoms that mimic different biological structures of interest. As the first step, we demonstrate the reliable use of the Rayleigh PDFs to differentiate between low-scattering structures and Intralipid® tissue-like media, using a goodness-of-fit metric. As the second innovation, our analysis now utilizes the Rayleigh PDFs fit parameters to further differentiate between the low-scattering structures. The potential of this technique to outperform classical methods is demonstrated. Future long-term studies, including applications to longitudinal treatment monitoring are briefly discussed.

2.

Methods

The Fourier-domain ss-OCT system used in this study has been previously described in detail.10 Briefly, it utilizes a laser (HS2000-HL, Santec, Japan) with a 20-kHz rotating polygon-based tunable filter, with a central wavelength of 1320 nm, a sweep range of 110 nm, and an average output power of 10 mW. The field-of-view of the volumetric images acquired in this study was 6  mm×6  mm×1.5  mm in depth, with 24 B-scans captured at each location (such repeats are needed in our analysis to ensure sufficient PDF speckle statistics). Each B-scan is acquired in two patches, with 400 A-scans per frame enabling the inter-frame interval of 25 ms. The distance between two adjacent B-scans is 3.75  μm with 800 A-scans per B-scan, and 1600 B-scans overall.

The Intralipid gel phantom was composed of water (89% by weight), gelatin (10%, G2500-500G; gel strength 300, Type A, Sigma-Aldrich Co, St. Louis, Missouri), and Intralipid (1%, Fresenius Kabi Canada Ltd., Richmond Hill, Ontario, Canada). The resultant optical properties are comparable to porcine skin with a transport mean free path of 1  mm.19 Inside the Intralipid phantom, we placed a polytetrafluoroethylene (PTFE) micro-tube (Masterflex RK-06417-11, Cole-Parmer Instrument Company, Vernon Hills, Illinois) with an inner diameter of 305  μm and outer diameter of 762  μm, and a 400μm diameter fishing line (Selizo Clear Fishing Wire) [see Fig. 1(a)]. To mimic lymphatic fluid, we used 1 ml water with two added drops of yellow food coloring (Club House, McCormick, London, Ontario, Canada). The yellow solution was injected by a syringe pump (NE-1000, New Era Pump Systems, Inc., Farmingdale, New York) into the PTFE tubing at a flow rate of 50  μl/min.

Fig. 1

Rayleigh distribution fitting to signal intensity histograms for selected ROIs on an OCT image. (a) Phantom used for this study, showing a fishing line and PTFE tube both embedded in Intralipid-gelatin mixture. (b) Representative B-scan of the phantom, showing six different ROIs for the speckle statistical analysis. (c)–(h) speckle histograms (green) for the six ROIs: Ar (R2>0.99), No (R2>0.99), dW (R20.985), tS (R20.98), stS (R20.96), and IL (R2<0.93), respectively. The red curves are the corresponding Rayleigh fits. Ar—air, No—noise, dW—dyed water, tS—transparent solid, stS—semi-transparent solid, and IL—Intralipid.

JBO_27_10_100501_f001.png

To perform a comprehensive analysis, six regions of interest (ROIs) were identified as air, noise, (phantom at depth in the low SNR area), fluid (inside the PTFE tube), transparent solid (fishing line), semi-transparent solid (PTFE tube wall), and Intralipid (tissue phantom). Each ROI had dimensions of 6×10×6 (fast lateral×slow lateral×depth) pixels, corresponding to physical dimensions of 45×37.5×45  μm3; with twenty-four repetitive B-scans, this yields a total of 8640 pixels. For each ROI, pixel intensity distributions were plotted as histograms and fitted with three-parameter Rayliegh PDF10

P(x;a,b,c)=a(xc)b2e(xc)22b2,
where x is the OCT signal intensity. The interpretation of the three fitting parameters comes primarily from quantitative ultrasound studies20 that have been adapted to OCT;10 roughly speaking, a is the amplitude normalization parameter, b is the scaling parameter, and c is the shifting parameter. Origin 2021 software was used to perform analysis of variance with multiple comparisons for the statistical analysis.

3.

Results

To demonstrate the feasibility of our improved technique to distinguish between lymphatics and nerves, six ROIs were chosen in the tissue-mimicking phantom for the Rayleigh PDF analysis as shown in Fig. 1(b). The resultant signal intensity histograms each contain 8640 pixels [green area in Figs. 1(c)1(h)] with the Rayleigh curve fitting in red. The highest goodness-of-fit range is observed in the ROIs corresponding to noise (No) and air (Ar) (R2>0.99), followed by dyed water (dW) and transparent solid (tS) with 0.99>R2>0.97, then semi-transparent solid (stS) with R20.96 and finally Intralipid (IL) with R2<0.93.

Next, we firmed up the fitting statistics by repeating the analysis 20 times. For each ROI, 20 observations were made (n=20) across two different phantoms and 10 ROIs selected for each of the six regions yielding R2 values of 0.993±0.0009 (air), 0.994±0.0009 (noise), 0.988±0.0011 (dW), 0.983±0.0014 (fishing line / transparent solid), 0.962±0.0135 (PTFE tubing / semi-transparent solid), and 0.910±0.0188 (Intralipid). The results are graphically summarized in Fig. 2, showing that this approach can clearly differentiate between the Intralipid (scattering tissue) and all else (transparent and semi-transparent structures, noise); this is similar to our previous in-vivo results.10,21 However, there is no significant difference between the two transparent/non-scattering media (dW and transparent solid); this is again consistent with our previous in-vivo result where the lymphatics and nerves could not be distinguished.10 Therefore, we extend our analysis by examining the three parameters of the Rayleigh PDF fits.

Fig. 2

Speckle statistical analysis via Rayleigh-fit distinguishes different simulated biological structures. The number of samples for each ROI is 20 (two different phantoms, 10 representative ROIs selected for each of the six regions). (p-values: *p0.05; ** p0.01; and *** p0.001).

JBO_27_10_100501_f002.png

The results for the a- and c-coefficients of the three-parameter Rayleigh PDF for the six ROIs are shown in Fig. 3. The a-coefficient in Fig. 3(a) is often referred to as the amplitude normalization parameter and appears to be the strongest differentiator between all types of regions: in decreasing order, a=69.1±5.0 (noise), 41.8±4.2 (air), 41.2±7.5 (transparent solid), 33.1±6.6 (dW), 11.2±5.2 (semi-transparent solid), and 3±1.7 (Intralipid). This is a noteworthy result as it demonstrates some potential differentiation between the dW (a33±7) and transparent solid regions (a42±7), the phantom analog of the lymphatics and nerves respectively (more specifically the myelin sheathing of the latter). Figure 3(b) shows the c-coefficient results; quantitatively, they are c=0.22±0.04 (air), 0.20±0.02 (noise), 0.22±0.03 (dW), 0.26±0.06 (transparent solid), 0.21±2.70 (semi-transparent solid), 7.56±5.26 (Intralipid). As seen, the only significant difference in c-coefficient values was between Intralipid and the rest of the five materials. Further, the b- fitting parameter showed minimal discrimination between any of the six types of ROIs (results not shown).

Fig. 3

Rayleigh function fit coefficients for the classification of six types of ROIs. (a) a- and (b) c-coefficients of the Rayleigh distribution function. A maximum a-coefficient was observed for the regions of noise. c-coefficient is significantly different between the Intralipid and the remaining five regions, similar to the R2 values trends in Fig. 2. (c) Differentiation based on a-coefficient between the tS (green color), dW (red color), and the stS (blue color). (d) The c-coefficient value ranges for the Intralipid. (p-values: *** p0.001).

JBO_27_10_100501_f003.png

Toward the quantification of the performance of classifiers based on the three Rayleigh-fit based features presented here (R2, a-, and c-coefficients), a receiver operating characteristic (ROC) curve analysis was performed. For a given B-scan, following Rayleigh-fit parameterization, 30 pixels in each of the six region types were randomly chosen for supervised training of classifiers with five-fold cross validation using the MATLAB R2021b software (The MathWorks, Inc., Natick, Massachusetts) Classifier Learner application. With all three parameters considered together, the optimal average validation group accuracy of (76.5±2.2)% was obtained in a fine Gaussian support vector machine (SVM) model (see Fig. 4). Despite this overall encouraging performance in the primary aim of detecting and distinguishing lymphatic-like (stS and dW) and nerve-like (tS) regions, Fig. 4 reveals that the SVM classifier underperforms in predicting regions representing air. This may be due to incomplete removal of the complex conjugate artifact, but further investigation is needed. Future work will improve on classifications by optimizing the pre-processing pipeline for the presented feature extraction and potentially introducing new features. For comparison, classifiers based on the OCT attenuation coefficient and gray-level-co-occurrence matrix texture analysis metrics were trained identically (data not shown) and were found to underperform with overall accuracies of (64.5±2.1)% and (45.3±0.8)%, respectively; importantly, they were also unable to differentiate between most of the inclusions.

Fig. 4

Training of classifiers based on the three proposed Rayleigh-fit metrics: R2, a-, and c- coefficients. The optimally performing classifier (in overall accuracy, and region-type specific sensitivity and specificity) was determined to be a fine Gaussian SVM. (a) ROC curves in the classification task for each of the six ROI types. The red dots indicate optimal selection points for high sensitivity and high specificity in detecting and differentiating stS, dW, and tS from one another and from the IL and No; adjacent bracketed numbers show false positive and true positive rates (accuracy). (b) Resultant confusion matrix. The overall accuracy is the six region averaged true positive rate, yielding (76.5±2.2)% (standard deviation from 10 randomizations of the training/validation datasets for all six classifications). The intensity of color shading represents the magnitude of TPR (blue) and FNR (brown).

JBO_27_10_100501_f004.png

To test the effects of changes in lymphatic flow rates, the water (±yellow food coloring) in the PTFE tubing under different flow conditions was examined. The three experiments involved dW flowing at a rate of 50  μl/min, non-flowing dW, and non-flowing water (the latter two conditions to focus on the effects of Brownian motion). As shown in Fig. 5(a), the R2 values were all very similar (0.98 for all three), suggesting that the absence/presence of flow does not affect our Rayleigh-fit analysis, at least at the goodness-of-fit level. Looking deeper at the a-b-c fitting parameter space, no significant differences were found [representative a-coefficient results shown in Fig. 5(b)]. This apparent insensitivity of our methodology to the lymphatic flow conditions may prove useful for in-vivo deployment, where this physiological variable is not controlled and in fact varies greatly in intra- and inter-patient settings.22

Fig. 5

Flow rate effects. (a) Similarity in the goodness-of-fit R2 values and (b) a-coefficient of Rayleigh distribution function of the fluid ROI, regardless of its flow conditions or presence/absence of dye.

JBO_27_10_100501_f005.png

4.

Discussion and Conclusion

In this study, we present a method to identify low-scattering structures within a tissue phantom based on OCT texture analysis utilizing speckle statistics. Based on the three-parameter Rayleigh distribution function fit to the pixel intensity distributions within selected image ROIs, this method successfully demonstrated the feasibility of not only detecting but also differentiating different types of transparent and semi-transparent inclusions amid tissue-like scattering background. Importantly, this method differentiates between optically translucent materials, namely, solid (fishing line) and fluid (dyed water inside the PTFE tube), which model nerves23 and the lymph fluid,8,14 respectively.

For the purpose of this study, axons within nerves can be considered simple fluid filled tubes where the fluid, similar to cytoplasm,24 is optically clear (because it is mainly composed of water and thus negligibly scatters light). Fishing line was selected as a low-scattering object to represent peripheral nerves because it has a low-scattering coefficient, being highly transparent due to its homogeneous composition.24 Thus, fishing line models optical properties of nerves well. Lymphatics are also difficult to detect due to their optical transparency in the visible and near-infrared spectrum,14 the underlying cause for their optical transparency is the lack of scatterers (and absorbers) contained in lymphatic fluid; only 6% are solid scattering components (cells, waste products, and/or excess proteins) while the remaining 94% is water.8,14 Thus, our dyed yellow water with minimal scatterer density is a realistic model for lymphatic fluid. The suitability of our phantoms for simulating the optical properties of nerves and lymphatics provides a useful testbed for optimizing methodology toward eventual in-vivo applications.

Imaging of lymphatics and nerves has historically been a challenge, however, recent techniques based on OCT imaging are showing some promise.811 This bodes well for detailed preclinical studies, and may have clinical relevance in sites with near-surface pathologies (e.g., skin, epithelial/mucosal lining of numerous body cavities). The OCT imaging depth is ~1–3 mm, therefore human lymphatic capillaries of ~10–60 μm25 diameter and peripheral nerves in the skin that branch directly into the sensory receptors with diameters of 100 to 500  μm11,24,25 may be detected in this range. Leveraging OCT’s limited penetration depth, this work may therefore lead to important applications ranging from early-detection of lymphoedema to compressed nerve diagnosis and surgical guidance.

The long-term objective of this study is to establish OCT as a valuable tool for investigating the full extent of the interactions of the lymphatics and nerves in the tumor microenvironment, which still remains largely unknown today. Due to the important role these semi-transparent structures are known to play in cancer metastasis, this tool may be vital to building a full picture of cancer treatment response (e.g., with radiotherapy) as we work toward adaptive personalized cancer medicine.

Disclosures

The authors declare that there are no conflicts of interest related to this article.

Acknowledgments

The OCT system was developed at the National Research Council of Canada with contribution from Drs. Costel Flueraru, Linda Mao, Shoude Chang, Sherif Sherif, and Erroll Murdock. This research was funded by the Canadian Institutes of Health Research, grant number 202010PJT.

References

1. 

A. W. Lambert, D. R. Pattabiraman and R. A. Weinberg, “Emerging biological principles of metastasis,” Cell, 168 (4), 670 –691 https://doi.org/10.1016/j.cell.2016.11.037 (2017). Google Scholar

2. 

F. Font-Clos, S. Zapperi and C. A. M. la Porta, “Blood flow contributions to cancer metastasis,” iScience, 23 101073 https://doi.org/10.1016/j.isci.2020.101073 (2020). Google Scholar

3. 

A. H. Zahalka and P. S. Frenette, “Nerves in cancer,” Nat. Rev. Cancer, 20 (3), 143 –157 https://doi.org/10.1038/s41568-019-0237-2 NRCAC4 1474-175X (2020). Google Scholar

4. 

A. K. Polomska and S. T. Proulx, “Imaging technology of the lymphatic system,” Adv. Drug Deliv. Rev., 170 294 –311 https://doi.org/10.1016/j.addr.2020.08.013 ADDREP 0169-409X (2021). Google Scholar

5. 

M. Ohana et al., “Current and future imaging of the peripheral nervous system,” Diagn. Interv. Imaging, 95 (1), 17 –26 https://doi.org/10.1016/j.diii.2013.05.008 (2014). Google Scholar

6. 

L. L. Munn and T. P. Padera, “Imaging the lymphatic system,” Microvasc. Res., 96 55 –63 https://doi.org/10.1016/j.mvr.2014.06.006 MIVRA6 0026-2862 (2014). Google Scholar

7. 

E. B. Brown et al., “In vivo measurement of gene expression, angiogenesis and physiological function in tumors using multiphoton laser scanning microscopy,” Nat. Med., 7 (7), 864 –868 https://doi.org/10.1038/89997 1078-8956 (2001). Google Scholar

8. 

P. Gong et al., “In vivo label-free lymphangiography of cutaneous lymphatic vessels in human burn scars using optical coherence tomography,” Biomed. Opt. Express, 7 (12), 4886 –4898 https://doi.org/10.1364/BOE.7.004886 BOEICL 2156-7085 (2016). Google Scholar

9. 

W. Qin, U. Baran and R. Wang, “Lymphatic response to depilation-induced inflammation in mouse ear assessed with label-free optical lymphangiography,” Lasers Surg. Med., 47 (8), 669 –676 https://doi.org/10.1002/lsm.22387 LSMEDI 0196-8092 (2015). Google Scholar

10. 

V. Demidov et al., “Analysis of low-scattering regions in optical coherence tomography: applications to neurography and lymphangiography,” Biomed. Opt. Express, 10 (8), 4207 https://doi.org/10.1364/BOE.10.004207 BOEICL 2156-7085 (2019). Google Scholar

11. 

A. E. Carolus et al., “Comparison between optical coherence tomography imaging and histological sections of peripheral nerves,” J. Neurosurg., 134 (1), 270 –277 https://doi.org/10.3171/2019.8.JNS191278 JONSAC 0022-3085 (2021). Google Scholar

12. 

A. E. Carolus et al., “High-resolution in vivo imaging of peripheral nerves using optical coherence tomography: a feasibility study,” J. Neurosurg., 132 (6), 1907 –1913 https://doi.org/10.3171/2019.2.JNS183542 JONSAC 0022-3085 (2020). Google Scholar

13. 

A. A. Moiseev et al., “Lymph vessels visualization from optical coherence tomography data using depth-resolved attenuation coefficient calculation,” J. Biophotonics, 14 (9), e202100055 https://doi.org/10.1002/jbio.202100055 (2021). Google Scholar

14. 

B. J. Vakoc et al., “Three-dimensional microscopy of the tumor microenvironment in vivo using optical frequency domain imaging,” Nat. Med., 15 (10), 1219 –1223 https://doi.org/10.1038/nm.1971 1078-8956 (2009). Google Scholar

15. 

U. Baran et al., “OCT-based label-free in vivo lymphangiography within human skin and areola,” Sci. Rep., 6 21122 https://doi.org/10.1038/srep21122 SRCEC3 2045-2322 (2016). Google Scholar

16. 

S. Yousefi et al., “Label-free optical lymphangiography: development of an automatic segmentation method applied to optical coherence tomography to visualize lymphatic vessels using Hessian filters,” J. Biomed. Opt., 18 (8), 086004 https://doi.org/10.1117/1.JBO.18.8.086004 JBOPFO 1083-3668 (2013). Google Scholar

17. 

M. S. Baleev et al., “Monitoring of blood and lymph circulation in the denervated small intestine based on multimodal optical coherence tomography data,” Proc. SPIE, 11968 75 –79 https://doi.org/10.1117/12.2609247 PSISDG 0277-786X (2022). Google Scholar

18. 

L. A. Matveev et al., “OCT lymphangiography based on speckle statistics evaluation,” Proc. SPIE, 11065 1106502 https://doi.org/10.1117/12.2523291 PSISDG 0277-786X (2019). Google Scholar

19. 

S. T. Flock et al., “Optical properties of Intralipid: a phantom medium for light propagation studies,” Lasers Surg. Med., 12 (5), 510 –519 https://doi.org/10.1002/lsm.1900120510 LSMEDI 0196-8092 (1992). Google Scholar

20. 

A. Sarti et al., “Maximum likelihood segmentation of ultrasound images with Rayleigh distribution,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 52 (6), 947 –960 https://doi.org/10.1109/TUFFC.2005.1504017 ITUCER 0885-3010 (2005). Google Scholar

21. 

L. A. Matveev et al., “Assessment of optical coherence tomography speckle patterns in low-scatterer-concentration regions: simulations for lymphatic vessels mapping,” Proc. SPIE, 11075 1107503 https://doi.org/10.1117/12.2526749 PSISDG 0277-786X (2019). Google Scholar

22. 

K. Alitalo, “The lymphatic vasculature in disease,” Nat. Med., 17 (11), 1371 –1380 https://doi.org/10.1038/nm.2545 1078-8956 (2011). Google Scholar

23. 

B. Young, G. O’Dowd and P. Woodford, Wheater’s Functional Histology, 6th ed.Elsevier( (2013). Google Scholar

24. 

Y. Grinberg et al., “Fascicular perineurium thickness, size, and position affect model predictions of neural excitation,” IEEE Trans. Neural Syst. Rehabil. Eng., 16 (6), 572 –581 https://doi.org/10.1109/TNSRE.2008.2010348 (2008). Google Scholar

25. 

K. N. Margaris and R. A. Black, “Modelling the lymphatic system: challenges and opportunities,” J. R. Soc. Interface, 9 (69), 601 –612 https://doi.org/10.1098/rsif.2011.0751 1742-5689 (2012). Google Scholar
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Muhammad Mohsin Qureshi, Nader Allam, Taylor Peters, Valentin V. Demidov, and I. Alex Vitkin "Detection and differentiation of semi-transparent materials simulating biological structures using optical coherence tomography: a phantom study," Journal of Biomedical Optics 27(10), 100501 (12 October 2022). https://doi.org/10.1117/1.JBO.27.10.100501
Received: 6 May 2022; Accepted: 14 September 2022; Published: 12 October 2022
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
KEYWORDS
Lymphatic system

Optical coherence tomography

Statistical analysis

Solids

Cancer

In vivo imaging

Speckle

Back to Top