1.IntroductionDiabetes is a metabolic disease characterized by high blood glucose levels, making it one of the most serious chronic diseases in the world.1–3 Among the non-invasive glucose measurement methods, near-infrared (NIR) diffuse reflectance spectroscopy has been well developed with a variety of devices and research results.4 The living skin’s blood glucose affects NIR diffuse light intensity by changing the refractive index5 and absorption of interstitial fluid, thus the measured diffuse light intensity can monitor the glucose in interstitial fluid, which is highly correlated with the glucose in the blood. In addition, the skin’s scattering also changes when the refractive index of interstitial fluid is altered. Therefore, glucose affects both the absorption and scattering of the skin tissue.6,7 In this paper, we discuss the NIR diffuse light signals of glucose in the 1000 to 1700 nm band. The measured light signal combines the absorption effect and the scattering effect. The scattering can influence the whole waveband of 1000 to 1700 nm, but the absorption mainly affects the signal of 1400 to 1700 nm. In 1400 to 1700 nm, glucose has relatively obvious absorption (mainly caused by bonds and bonds). However, for the 1000 to 1400 nm band, the absorption is very weak and almost none, so the glucose’s response is dominated by scattering. Moreover, even for 1400 to 1700 nm, the glucose-caused absorption coefficient change is much smaller than the change in scattering coefficient, thus the scattering still dominates the glucose signal here.8,9 According to the study of Amerov et al., the absorption coefficient in aqueous solution can change by at the wavelength range of 1500 to 1600 nm for every 1 mM increase in glucose, whereas it is almost zero for 1000 to 1300 nm.10 According to Liu et al., the reduced scattering coefficient at 1000 nm will be approximately for 1 mM glucose, which is much greater than the change in absorption coefficient.11 According to optical coherence tomography (OCT) results, the scattering coefficient change caused by glucose in living tissue is significantly greater than that in phantom solutions, and the coherent signal response caused by glucose in living tissue can reach tens or hundreds of times that of phantom solution.12 In this paper, we discuss the contributions of absorption and scattering of glucose, respectively, to the glucose signal in Intralipid solution. Except for glucose, many other factors, such as the ions (, , etc.) in the interstitial fluid, the density of tissue particles, and the size of tissue particles, can affect tissue scattering and lead to similar spectra to the glucose’s scattering spectrum, thus reducing the accuracy of glucose prediction.13 Unfortunately, chemometrics methods, such as partial least squares regression,14 still have no good solution for this problem. In fact, there are slight differences between the spectra of the above scattering variables. The concentrations of ions (, , etc.) and glucose in the interstitial fluid affect the fluid’s refractive index; the density and size of particles in tissue can change the particles’ distribution. Both the fluid’s refractive index and the particles’ distribution can change the scattering coefficient of tissue.15 In addition, glucose has additional absorption in the NIR band, particularly in the range of 1400 to 1700 nm. This paper aims to analyze the spectral differences between the above factors, in order to provide theoretical support to distinguish glucose from other factors, especially in terms of finding reasonable wavelengths. We use the theory of Mie scattering to discuss the scattering difference between the factors.16,17 Mie scattering theory is often based on the equations proposed by Van.18 According to Van’s theory, Graaff et al. obtained the approximate calculation formula of the reduced scattering cross section using the relative refractive index, particle size, and particle density, thus promoting the application of Mie scattering theory in biological tissues.19 Kohl et al. provided a simplified equation of the reduced scattering coefficient for tissues to estimate the effect of glucose on tissue scattering.20 Liu et al. investigated the scattering alteration of Intralipid solution and ex vivo tissue when carbohydrate was added to the media.21 In the 1000 to 1700 nm range, we have selected six significant wavelengths for glucose measurement, namely 1050, 1219, 1314, 1409, 1550, and 1609 nm. Glucose absorption is minimal at 1050, 1219, and 1314 nm, mainly used to reflect glucose scattering, whereas 1409, 1550, and 1609 nm provide more apparent absorption information for glucose and produce comprehensive results of glucose absorption and scattering.4 The 1050 nm wavelength can exhibit hemoglobin’s absorption22 and can be used to monitor blood flow. Fat has strong absorption at 1219 nm,23 which can reflect changes in subcutaneous fat. Water has higher absorption at 1409 nm, so it is greatly influenced by skin water content and temperature.24 In addition, 1314 nm is a compromise wavelength that is relatively insensitive to hemoglobin, fat, water, and even temperature changes.25,26 Therefore, multiple wavelengths are necessary to acquire enough information. Han et al. reported the glucose measurement results using the six wavelengths on human subjects,27 which showed that the 1409, 1550, and 1609 nm wavelengths exhibit excellent glucose measurement sensitivity. The source–detector separations (SDSs) selected in this paper are also reasonable. Liu et al. discovered that there are corresponding suitable SDS ranges for different wavelengths, where shows an approximately linear change with SDS.11 For shorter wavelengths, SDS can be 0.8 to 5 mm.28,29 The SDSs used in this paper, 1.25 to 2.3 mm, can meet the measurement requirements of the six wavelengths. The small SDSs of 1.25 to 1.7 mm could be better than other SDSs for 1409, 1550, and 1609 nm. 2.Theoretical Analysis of Scattering Factors2.1.Reduced Scattering Coefficient in Mie Scattering TheoryAccording to Graaff’s study,19 when the wavelength and particle’s radius a meet the case of , the reduced scattering cross-section can be approximated as where is the scattering cross-section, is the anisotropy factor, is the refractive indices inside the particles, is for that outside the particles, i.e., the refractive indices of the solution in the tissue, and is the relative refractive index.When the scatterer is spherical, the reduced scattering coefficient of tissues can be expressed as30 where is the volume fraction of the scattered particles.In a 3% Intralipid solution, soybean oil particles (oil) are the main scattering particles, with the background solution being predominantly aqueous.31–33 The volume fraction of scattered particles (soybean oil) was about 0.03237. Christian et al. and Xu et al. determined the average particle size of the Intralipid solution is .34,35 In the 1000 to 1660 nm band, the and of Intralipid solution are shown in Fig. 1(a).36 Fig. 1Optical properties of 3% Intralipid solution: (a) The refractive index and (b) absorption coefficient and reduced scattering coefficient. ![]() Using the typical values of (), (0.03237), , and , the reduced scattering coefficient can be calculated by using Eq. (2), as shown in Fig. 1(b). The absorption coefficient of the solution also is shown in Fig. 1(b). 2.2.Scattering Change Caused by Solution’s Refractive IndexAccording to Eq. (2), the derivative of the scattering coefficient with respect to the background refractive index can be as Using Eq. (3), we can estimate the and , i.e., caused by 1 mM glucose concentration () and 1 mM ion concentration (, as we use NaCl), where for aqueous solution is set to about , and for aqueous solution is set to about .37,38 Figure 2(a) shows the results. 2.3.Scattering Change Caused by Particle DensityAccording to Eq. (2), the derivative of the scattering coefficient with respect to the particle density (i.e., volume fraction ) can be as where the volume fraction of scattered particles in 3% Intralipid solution is about 0.03237.2.4.Scattering Coefficient Change Caused by Particle SizeAccording to Eq. (2), the derivative of the scattering coefficient with respect to the particle size (i.e., particle radius ) can be as The radius of the particles in Intralipid solution is . The average particle radius of is used in this paper. Our analysis shows that the caused by particle concentration and particle size are contained within the previous coefficients, which remain identical when is 0.00137% and is 0.336 nm. At 1000 nm, their values are the same as those caused by 1 mM glucose, but they differ at other wavelengths: the absolute value of caused by glucose is slightly larger than that caused by particle density or size. These findings are presented in Fig. 2(b). 3.Differential AbsorbanceAccording to the first-order approximation of the radiative transport equation, for a semi-infinite medium, at a SDS of , the diffuse reflection light intensity can be expressed as where is the incident light intensity, is the effective attenuation coefficient (EAC), , is the diffusion coefficient, , , , , and .The diffuse light absorbance at the SDS of is defined by the natural logarithm as Differential absorbance between two SDSs ( and ) can be For the given SDSs and , differential absorbance is approximated as the variable related to EAC,39 which can reduce the light source drift, instrument drift, etc. Thus, the change in EAC can be approximated as The above derivation roughly explains the characteristic that absorbance change shows an approximately linear change with SDS when the component changes. The above derivation may not be suitable for wavelengths near 1450 nm with too strong water absorption and SDSs too close or too far away. Nevertheless, the characteristic that absorbance shows a monotonic change with SDS will remain unaffected, and its non-linearity can be corrected as a systematic error in measurement. 4.Monte Carlo SimulationThe MCMLGO program written by Lihong V. Wang was used to simulate the diffuse reflection intensity of the scattering media (3% Intralipid solution). The number of photons was set to , and the wavelengths are 1000 to 1660 nm with an interval of 10 nm. The media’s optical parameters (, , , ) were set according to Fig. 1(b), and their changes , and caused by glucose, NaCl, and Intralipid solution concentration were set according to Fig. 2. (1) The first group includes the solutions with 0 and 100 mM glucose. was set to ; Furthermore, glucose-induced and were independently set in two simulations in order to illustrate their separate effects. (2) The second group simulates the diffuse reflection light intensity of the solution with 100 mM NaCl, is . (3) The third group simulated the diffuse reflection intensity of the two Intralipid solutions (3%, 3.5%). According to Eqs. (7) and (8), the absorbance of the SDSs of 0.01 to 0.25 cm for the three groups of simulation can be obtained [Fig. 3(b)], and then the differential absorbance and EAC can be calculated using Eqs. (8) and (9). Two SDSs (0.08 and 0.2 cm) were selected for differential processing. In the first group, the spectrum of 1 mM glucose was estimated from the spectrum of 100 mM glucose, the spectrum of NaCl of 2.68 mM was estimated from the spectrum of 100 mM NaCl, and the spectrum of 0.001525% Intralipid solution concentration was estimated from the data of 0.5% Intralipid solution concentration change (3% and 3.5%). After the data processing, the caused by 1 mM glucose, 2.68 mM NaCl, and 0.001525% Intralipid solution concentration can be the same at 1000 nm, and the three spectra were compared in Fig. 3(c). Fig. 3MC simulation results. (a) of SDSs of 0.01 to 0.25 cm caused by 100 mM glucose at six wavelengths; (b) of SDSs of 0.01 to 0.25 cm caused by 100 mM glucose, 100 mM NaCl, and 0.5% Intralipid solution concentration (1600 nm); (c) caused by 1 mM glucose, 2.68 mM NaCl, and 0.001525% Intralipid solution; (d) caused by 1 mM glucose, glucose’s scattering, and glucose’s absorption. ![]() Figure 3(d) shows the results caused by the absorption change of glucose, the scattering change of glucose, and the two effects together. It is evident that scattering change is the predominant factor in the glucose signal, and glucose absorption can only be observed between 1400 and 1700 nm. Further, caused by glucose’s scattering and absorption at 1400 to 1700 nm shows the opposite results, then the absolute value of is reduced here. 5.Experiment5.1.Measurement SystemThe measurement system shown in Fig. 4 is composed of a light source, an optical switch, a sensing probe, and a data processing unit. The light source consists of six superluminescent diodes (Inphenix) with center wavelengths of 1050, 1219, 1314, 1409, 1550, and 1609 nm (bandwidth 60 to 130 nm). Data sampling at each wavelength takes . The incident light is guided by a silicon fiber with a core diameter of . The photosensitive area of the detector (U-Science Co., Ltd.) is made of a ring-shaped InGaAs material with a width of 0.2 mm. The detector includes four separate photosensitive areas, in the middle of which the SDSs are 1.25, 1.7, 2.0, and 2.3 mm, respectively. Each pair of two adjacent sensing areas is used for differential processing. At the farthest SDS of 2.3 mm, the absorbance fluctuations of six wavelengths within 1 min are respectively , , , , , and a.u., whereas the fluctuation amplitudes under the other closer SDSs are smaller. 5.2.Materials30% Intralipid solution (Fresenius Kabihuarui Pharmaceutical Co., Ltd.) was used in the experiment. First, the 30% Intralipid solution was diluted to 3% with pure water. Then, the 3% Intralipid solution with glucose and NaCl were prepared, respectively, by adding the anhydrous glucose powder or NaCl powder. The 3.5% Intralipid solution was made by mixing the 3% and 30% Intralipid solution in the proper ratio. For each test, the 3% Intralipid solution was measured first, as the baseline of the other solutions. 5.3.Experimental ResultsThe diffuse light intensity was recorded at the four SDSs, and the differential absorbance of two adjacent SDSs was obtained using Eqs. (8) and (9). The experimental results are presented in Figs. 5 and 6. The absorbance caused by a 100 mM glucose, 200 mM NaCl, and 0.5% Intralipid solution concentration change at SDSs of 1.25, 1.7, 2.0, and 2.3 mm are shown in Fig. 5. The EAC spectra were calculated using Eq. (10), as shown in Fig. 6. Due to the differences in the coverage and effectiveness of the four photosensitive regions, there are systematic biases in the EAC values. Fig. 5The experimentally measured changes at the SDSs of 1.25, 1.7, 2.0, and 2.3 mm, respectively. (a) The absorbance caused by 100 mM glucose; (b) the absorbance caused by 200 mM NaCl; and (c) the absorbance caused by 0.5% Intralipid solution concentration change. ![]() Fig. 6The experimentally measured changes in the concentration of 100 mM glucose, 200 mM NaCl, and Intralipid solution by 0.1%, respectively. (a) 1.25 and 1.7 mm differential; (b) 1.7 and 2.0 mm differential; (c) 2.0 and 2.3 mm differential. ![]() From Fig. 5, it can be observed that varies approximately linearly with SDS, but this linearity is not consistent for every wavelength and SDS. The theoretical results of versus SDS from MC simulation are shown in Fig. 3(a). Overall, the linearity of the three shorter wavelengths (1050, 1219, and 1314 nm) is better than that of the three longer wavelengths (1409, 1550, and 1609 nm), possibly due to the shorter wavelengths being less absorbed by the skin, having longer optical paths, deeper penetration depths, and better adhering to diffusion equation. In a more detailed analysis, the linearity of the three shorter wavelengths is worse for smaller SDSs compared to larger SDSs due to scattering effects, whereas the linearity of the three longer wavelengths deteriorates as SDS becomes too large, and light attenuates sharply due to the stronger absorption of water. Therefore, each wavelength has a suitable linear range of SDS for measurement, neither too close nor too far. Due to limitations in the measurement system and the maximum tolerable light intensity by the skin, it is not possible to infinitely increase the energy of the light sources, resulting in systemic differences between the differential results. The use of different SDSs for differential processing will also produce varying EACs, as demonstrated in Fig. 6. In addition to the aforementioned reasons, other system errors may also exist in the measurement process, such as the angle of incidence of the light source not being completely vertical, leading to eccentricity, and resulting in sensing deviation for each SDS. As shown in Fig. 6, for a given pair of SDSs, the spectra of the three variables appear different. First, the glucose’s spectra show great differences to the spectra of NaCl and Intralipid solution concentration change since it shows the maximum of is around 1409 nm, which are more likely around 1550 nm for the spectra of NaCl and Intralipid solution concentration change. Second, the glucose’s spectra at 1050 to 1314 nm is different from the others, as they always appear a great change with wavelength. However, for the at 1050 to 1314 nm of NaCl and Intralipid solution concentration change, they are similar or just have little difference between the wavelengths. In the MC simulation results of Fig. 3, we can also see the differences in glucose and the others. The maximum of glucose’s also appear around 1400 nm, but the maximums of caused by NaCl or Intralipid solution concentration are more likely around 1450 nm. Meanwhile, there are differences for the spectra of 1000 to 1300 nm, the glucose spectrum is slightly upward due to the glucose absorption’s effect. It is different from the experimental results, in which the glucose’s spectrum at 1000 to 1300 nm is slightly downward. In our theoretical analysis, the water concentration change was ignored for the particle concentration change and particle size change, thus the experiment may show the true results. 6.ConclusionIn this paper, the spectral analysis based on the diffuse equation, Monte Carlo simulation, and measurement experiment was used to investigate the differences between the spectra caused by glucose and other scattering factors in the 1000 to 1700 nm band. The investigated scattering-induced factors include the medium’s refractive index, particle concentration, and particle size. The theoretical analysis using the diffuse equation and Monte Carlo simulation showed that the particle concentration and particle size have similar spectral characteristics. In the experiment, the spectra of glucose concentration, NaCl concentration, and particle concentration in solutions were tested. The NaCl concentration and particle concentration are related to the refractive index change and particle density change, respectively. The two scattering factors showed similar spectrum characteristics in the experiment. Moreover, we have found that the glucose has definite differences from the other scattering factors both in the theoretical analysis and the experimental results: in the 1000 to 1300 nm band, the spectrum of glucose changes more with wavelength while the spectra of the scattering factors (such as NaCl concentration and particle concentration) are gently here; in the 1400 to 1700 nm band, due to the glucose’s absorption, the glucose’s spectrum increases significantly and the maximum of the spectral curve of the move to around 1400 nm as it is around the longer wavelength (1450 nm for MC results and 1550 nm for experiment results) for the spectra of NaCl and particle concentration. The above spectral characteristics can be well applied to distinguish the glucose’s spectrum and the spectra caused by the other scattering-induced factors. In this paper, we explore the combined effects of glucose absorption and scattering in intralipid solution. Our findings could be useful in identifying glucose signals in vivo, despite the fact that scattering signals from glucose are much stronger in the human body than in vitro or in the Intralipid solution.15 To monitor glucose scattering in the human body, wavelengths in the range of 1000 to 1300 nm can be employed, whereas the combined effect of scattering and absorption in the human body can be obtained using wavelengths of 1400 to 1700 nm. Because multiple wavelengths may be necessary for accurate glucose signal identification and multivariate calibration, our results highlight the importance of considering these combined effects when developing new methods for glucose monitoring. DisclosuresThe authors declare that they have no affiliations with or involvement in any organization or entity with any financial interest in the subject matter or materials discussed in this paper. AcknowledgmentsThe research leading to these results was funded by the National Natural Science Foundation of China (81971657 and 81871396). ReferencesC. M. Mooshage et al.,
“Diametrical effects of glucose levels on microvascular permeability of peripheral nerves in patients with type 2 diabetes with and without diabetic neuropathy,”
Diabetes, 72
(2), 290
–298 https://doi.org/10.2337/db22-0548 DIAEAZ 0012-1797
(2023).
Google Scholar
S. Demir et al.,
“Emerging targets in type 2 diabetes and diabetic complications,”
Adv. Sci., 8
(18), 2100275 https://doi.org/10.1002/advs.202100275
(2021).
Google Scholar
S. Lim, J. W. Kim and G. Targher,
“Links between metabolic syndrome and metabolic dysfunction-associated fatty liver disease,”
Trends Endocrinol. Metab., 32
(7), 500
–514 https://doi.org/10.1016/j.tem.2021.04.008
(2021).
Google Scholar
J. Yadav et al.,
“Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy,”
Biomed. Signal. Process., 18 214
–227 https://doi.org/10.1016/j.bspc.2015.01.005
(2015).
Google Scholar
K. Maruo and Y. Yamada,
“Near-infrared noninvasive blood glucose prediction without using multivariate analyses: introduction of imaginary spectra due to scattering change in the skin,”
J. Biomed. Opt., 20
(4), 047003 https://doi.org/10.1117/1.JBO.20.4.047003 JBOPFO 1083-3668
(2015).
Google Scholar
S. Delbeck et al.,
“Non-invasive monitoring of blood glucose using optical methods for skin spectroscopy: opportunities and recent advances,”
Anal. Bioanal. Chem., 411 63
–77 https://doi.org/10.1007/s00216-018-1395-x ABCNBP 1618-2642
(2019).
Google Scholar
J. Huang, Y. Zhang and J. Wu,
“Review of non-invasive continuous glucose monitoring based on impedance spectroscopy,”
Sens. Actuator A-Phys., 311 112103 https://doi.org/10.1016/j.sna.2020.112103
(2020).
Google Scholar
M. Zhao et al.,
“Activatable fluorescence sensors for in vivo bio-detection in the second near-infrared window,”
Chem. Sci., 12
(10), 3448
–3459 https://doi.org/10.1039/D0SC04789A 1478-6524
(2021).
Google Scholar
F. Dong et al.,
“A combination of near-infrared hyperspectral imaging with two-dimensional correlation analysis for monitoring the content of alanine in beef,”
Biosensors, 12
(11), 1043 https://doi.org/10.3390/bios12111043 BISSED 0265-928X
(2022).
Google Scholar
A. K. Amerov et al.,
“Molar absorptivities of glucose and other biological molecules in aqueous solutions over the first overtone and combination regions of the near-infrared spectrum,”
Appl. Spectrosc., 58
(10), 1195
–1204 https://doi.org/10.1366/0003702042336136 APSPA4 0003-7028
(2004).
Google Scholar
J. Liu et al.,
“Specialized source-detector separations in near-infrared reflectance spectroscopy platform enable effective separation of diffusion and absorption for glucose sensing,”
Biomed. Opt. Express, 10
(9), 4839
–4858 https://doi.org/10.1364/BOE.10.004839 BOEICL 2156-7085
(2019).
Google Scholar
K. V. Larin et al.,
“Potential application of optical coherence tomography for noninvasive monitoring of glucose concentration,”
Proc. SPIE, 4263 83
–90 https://doi.org/10.1117/12.429342 PSISDG 0277-786X
(2001).
Google Scholar
S. R. Arridge, M. Cope and D. T. Delpy,
“The theoretical basis for the determination of optical pathlengths in tissue: temporal and frequency analysis,”
Phys. Med. Biol., 37
(7), 1531 https://doi.org/10.1088/0031-9155/37/7/005 PHMBA7 0031-9155
(1992).
Google Scholar
K. Meacham-Hensold et al.,
“High-throughput field phenotyping using hyperspectral reflectance and partial least squares regression (PLSR) reveals genetic modifications to photosynthetic capacity,”
Remote Sens. Environ., 231 111176 https://doi.org/10.1016/j.rse.2019.04.029
(2019).
Google Scholar
J. S. Maier et al.,
“Possible correlation between blood glucose concentration and the reduced scattering coefficient of tissues in the near infrared,”
Opt. Lett., 19
(24), 2062
–2064 https://doi.org/10.1364/OL.19.002062 OPLEDP 0146-9592
(1994).
Google Scholar
J. M. Steinke and A. P. Shepherd,
“Comparison of Mie theory and the light scattering of red blood cells,”
Appl. Opt., 27
(19), 4027
–4033 https://doi.org/10.1364/AO.27.004027 APOPAI 0003-6935
(1988).
Google Scholar
B. Beauvoit et al.,
“Characterization of absorption and scattering properties for various yeast strains by time-resolved spectroscopy,”
Cell Biophys., 23 91
–109 https://doi.org/10.1007/BF02796508 CBIODE 0163-4992
(1993).
Google Scholar
H. Hulst, Light Scarttering by Small Particles, 9th ed.Dover Publications, Inc., New York
(1957). Google Scholar
R. Graaff et al.,
“Reduced light-scattering properties for mixtures of spherical particles: a simple approximation derived from Mie calculations,”
Appl. Opt., 31
(10), 1370
–1376 https://doi.org/10.1364/AO.31.001370 APOPAI 0003-6935
(1992).
Google Scholar
M. Kohl et al.,
“Influence of glucose concentration on light scattering in tissue-simulating phantoms,”
Opt. Lett., 19
(24), 2170
–2172 https://doi.org/10.1364/OL.19.002170 OPLEDP 0146-9592
(1994).
Google Scholar
H. Liu et al.,
“Dependence of tissue optical properties on solute-induced changes in refractive index and osmolarity,”
J. Biomed. Opt., 1
(2), 200
–211 https://doi.org/10.1117/12.231370 JBOPFO 1083-3668
(1996).
Google Scholar
S. Prahl, Optical Absorption of Hemoglobin, Oregon Medical Laser Center, Portland, Oregon
(1999). Google Scholar
R. H. Wilson et al.,
“Review of short-wave infrared spectroscopy and imaging methods for biological tissue characterization,”
J. Biomed. Opt., 20
(3), 030901 https://doi.org/10.1117/1.JBO.20.3.030901 JBOPFO 1083-3668
(2015).
Google Scholar
P. S. Jensen,
“Influence of temperature on water and aqueous glucose absorption spectra in the near-and mid-infrared regions at physiologically relevant temperatures,”
Appl. Spectrosc., 57
(1), 28
–36 https://doi.org/10.1366/000370203321165179 APSPA4 0003-7028
(2003).
Google Scholar
I. V. Meglinski and S. J. Matcher,
“Quantitative assessment of skin layers absorption and skin reflectance spectra simulation in the visible and near-infrared spectral regions,”
Physiol. Meas., 23
(4), 741
–753 https://doi.org/10.1088/0967-3334/23/4/312 PMEAE3 0967-3334
(2002).
Google Scholar
K. Maruo et al.,
“New methodology to obtain a calibration model for noninvasive near-infrared blood glucose monitoring,”
Appl. Spectrosc., 60
(4), 441
–449 https://doi.org/10.1366/000370206776593780 APSPA4 0003-7028
(2006).
Google Scholar
T. Han et al.,
“In vivo measurement strategy for near-infrared noninvasive glucose detection and human body verification,”
Appl. Spectrosc., 76
(9), 1100
–1111 https://doi.org/10.1177/00037028221092474 APSPA4 0003-7028
(2022).
Google Scholar
G. Han et al.,
“Preliminary clinical validation of a differential correction method for improving measurement accuracy in noninvasive measurement of blood glucose using near-infrared spectroscopy,”
Appl. Spectrosc., 71
(9), 2177
–2186 https://doi.org/10.1177/0003702816685335 APSPA4 0003-7028
(2017).
Google Scholar
M. Althobaiti and I. Al-Naib,
“Optimization of dual-channel near-infrared non-invasive glucose level measurement sensors based on Monte-Carlo simulations,”
IEEE Photonics J., 13
(3), 1
–9 https://doi.org/10.1109/JPHOT.2021.3079408
(2021).
Google Scholar
B. Beauvoit, T. Kitai and B. Chance,
“Contribution of the mitochondrial compartment to the optical properties of the rat liver: a theoretical and practical approach,”
Biophys. J., 67
(6), 2501
–2510 https://doi.org/10.1016/S0006-3495(94)80740-4 BIOJAU 0006-3495
(1994).
Google Scholar
J. Linford et al.,
“Development of a tissue-equivalent phantom for diaphanography,”
Med. Phys., 13
(6), 869
–875 https://doi.org/10.1118/1.595948 MPHYA6 0094-2405
(1986).
Google Scholar
S. T. Flock et al.,
“Optical properties of Intralipid: a phantom medium for light propagation studies,”
Laser Surg. Med., 12
(5), 510
–519 https://doi.org/10.1002/lsm.1900120510
(1992).
Google Scholar
I. Driver et al.,
“The optical properties of aqueous suspensions of Intralipid, a fat emulsion,”
Phys. Med. Biol., 34
(12), 1927 https://doi.org/10.1088/0031-9155/34/12/015 PHMBA7 0031-9155
(1989).
Google Scholar
C. J. M. Moes et al.,
“Measurements and calculations of the energy fluence rate in a scattering and absorbing phantom at 633 nm,”
Appl. Opt., 28
(12), 2292
–2296 https://doi.org/10.1364/AO.28.002292 APOPAI 0003-6935
(1989).
Google Scholar
T. Xu et al.,
“Measurement and analysis of light distribution in Intralipid-10% at 650 nm,”
Appl. Opt., 42
(28), 5777
–5784 https://doi.org/10.1364/AO.42.005777 APOPAI 0003-6935
(2003).
Google Scholar
S. Kedenburg et al.,
“Linear refractive index and absorption measurements of nonlinear optical liquids in the visible and near-infrared spectral region,”
Opt. Mater. Express, 2
(11), 1588
–1611 https://doi.org/10.1364/OME.2.001588
(2012).
Google Scholar
R. C. Weast, CRC Handbook of Chemistry and Physics, 55th ed.CRC, Cleveland, Ohio
(1974). Google Scholar
K. V. Larin et al.,
“Specificity of noninvasive blood glucose sensing using optical coherence tomography technique: a pilot study,”
Phys. Med. Biol., 48
(10), 1371 https://doi.org/10.1088/0031-9155/48/10/310 PHMBA7 0031-9155
(2003).
Google Scholar
T. J. Farrell, M. S. Patterson and B. Wilson,
“A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,”
Med. Phys., 19
(4), 879
–888 https://doi.org/10.1118/1.596777 MPHYA6 0094-2405
(1992).
Google Scholar
BiographyLu Yue is a graduate student at Tianjin University. She received her bachelor’s degree in engineering from Tianjin University in 2021. Her current research includes near-infrared spectroscopy, biomedical engineering, and instrumentation. Han Tongshuai is an assistant researcher at Tianjin University. He received his doctoral degree from Tianjin University in 2020. His current research interests are in the areas of near-infrared spectroscopy, biomedical engineering, and instrumentation. |