1.IntroductionTemperature measurement by infrared thermography is liable to several sources of uncertainty both internal and external to the imager.1,2 These uncertainties somehow combine and propagate to affect the result of the measurement system, i.e., the measured temperature of the inspected object. In other words, a temperature measured by thermography has an uncertainty, which is a combination of all sources of uncertainties.3 Internal sources of uncertainty are well understood and their influence is presented in a condensed way by manufacturers as the performance characteristics of the imager, for example, in the expected range of uncertainty for the measurements, such as thermal sensitivity and others.4 Alternatively, these internal sources of uncertainty can be estimated from the calibration certificate.5 Proper calibration methods can provide the total measurement uncertainty of the thermal imager due to its several internal sources of uncertainty.6 Temperature measurement by thermography also involves other instruments besides the thermal imager, such as the thermometer, hygrometer, and measuring tape (or a laser distance meter). The atmospheric temperature (propagation medium temperature), relative humidity, the distance between the imager and the inspected object, and the temperature of objects located in the ambient (ambient temperature) also have to be measured. The thermographer estimates the emissivity of the inspected object. These measurements are considered here as sources of uncertainty outside the imager. The works presented in Refs. 7 and 8 present a method for calculating the uncertainty of temperature measurement considering the uncertainties of transmittance of the propagating medium, of the object’s emissivity, and of the thermal imager. However, some information is not generally available to the thermographer. This includes (1) the transmittance of the propagating medium and its associated uncertainty and (2) the mathematical model, i.e., the spectral sensitivity of the thermal imager employed. This information is generally not available.9,10 Thus, the works presented in Refs. 7 and 8 describe a method for calculating the measurement uncertainty associated with the temperature measured by thermography. However, they are not useful as they require some information generally not available in the data sheet or in the user’s manuals of the measurement instruments. The search for methods for calculating the uncertainty of the temperature measurement by thermography has encouraged us to contact manufacturers of thermal imagers. We found that the thermal processing image and analysis software provided by the manufacturers does not calculate the measurement uncertainty, in spite of “when reporting the result of a measurement of a physical quantity, it is obligatory that some quantitative indication of the quality of the result be given so that those who use it can assess its reliability.”3 It would be appropriate for such software to estimate the uncertainty of the temperature measurement, but the determination of the uncertainty due to sources outside the imager is the user’s responsibility.6 In general, the calculation models for uncertainty are reserved by the manufacturers.8 Occasionally, only upon request, the manufacturer will provide the output data (uncertainty associated with the temperature measurement by thermography) from the input data (uncertainty of measurement instruments) informed by the client/customer.9,10 However, the calculation method is not available. The work presented in Ref. 11 suggests a laboratorial procedure to determine the function that relates the internal camera signal to the temperature of the inspected object. Unfortunately, this procedure is not practical for the regular thermographer because the resources needed are not common in service shops and because thermal imagers may use different internal calibration models.11 In addition to the recommendation that all measurements should be accompanied by their associated uncertainty,3 the estimation of measurement temperature uncertainty helps the thermographer to make more accurate diagnoses. For instance, in electrical maintenance, the criteria of temperature difference indicative of faults is ,12–15 a value for which the measurement uncertainty should be a maximum of 1 to 2°C (approximately one third of the tolerance of the measurement process).16,17 This paper presents a calculation method to estimate the uncertainty of temperature measurement by infrared thermography in predictive maintenance due to the uncertainties of the instruments used and measurements taken by the thermographer from data usually available to him/her. These are the distance between the inspected object and the thermal imager, the atmospheric temperature and the relative humidity, the emissivity of the object, the temperature of the objects located in the ambient, and the imager itself. It is assumed that the thermographer performs the thermographic inspection according to the recommendations of the thermal imager manufacturer, respecting the angle of view limits, the minimum inspected object dimensions, etc. In addition to estimating measurement uncertainty, we estimate the contribution from each source of uncertainty for the final uncertainty measurement so as to enable the thermographer to analyze and prioritize investment to improve the quality of temperature measurement. 2.CalculationAs mathematical models of thermal imagers that correlate the spectral excitance received and the temperature of the inspected object are not available, we used general phenomenological and empirical models available in the literature. Of course, these models do not consider the aberrations and nonlinearities of each imager;1 however, they provide a “quantitative indication of the quality of the outcome”3 of the measured temperature, which nowadays is not made available by the manufacturers of thermal imagers and software developers. 2.1.Quantities Involved in Temperature Measurement by ThermographyIn a survey of four Internet sites of manufacturers of imagers, we found that the imagers recommended for predictive maintenance operate approximately in the range of 7.5 to 14 μm.18–21 This paper, therefore, focuses on this wavelength range. For this range transmittance, atmospheric absorption, which is the predominant factor for atmospheric attenuation, is virtually independent of the wavelength.22 In the operating wavelength range of thermal imagers used in predictive maintenance,2 7.5 to 14 μm, the method available in Ref. 22 can be used to calculate atmospheric transmittance. The calculating method available in Ref. 22 considers that concentrations of absorbers like nitrous oxide, carbon monoxide, oxygen, methane, and nitrogen can be considered constant and their contributions irrelevant to practical problems of thermography. The concentration of water vapor varies widely according to the location and climate and is relevant in calculation of attenuation. The concentration of carbon dioxide () can be considered almost constant at and, in principle, it should be considered in the calculation of atmospheric attenuation. However, when considering the works published in Refs. 23 and 24, it is noted that the atmospheric transmittance due to the concentration of is almost unitary in the range of 7.5 to 14 μm for propagation distances , common in thermography applied to predictive maintenance. Thus, the calculation of atmospheric attenuation is summarized in calculating the attenuation due to water molecules existing in the propagation path. Thus, the method consists in calculating the atmospheric internal transmittance, as only their absorption effect is relevant.2 Considering the uncertainty sources of interest, as previously defined in Sec. 1 (the distance between the inspected object and the imager, atmospheric temperature, relative humidity, emissivity of the object, ambient temperature, and the thermal imager), the calculation method is presented as a flow chart in Fig. 1. The definition of the calculation methods and calculation results provide the basis for calculating the propagation of uncertainty from these sources to the measured temperature for the inspected object. Absolute humidity is a characteristic of the propagation medium that is not readily available. However, it can be calculated from the relative humidity and atmospheric temperature, quantities easily measurable by the thermographer, with Eq. (1)25 where is the relative humidity of the propagation medium (dimensionless), is the propagation medium temperature (unit: K); and is the absolute humidity of air in the propagation path (unit: ).Once the absolute humidity is obtained, the amount of precipitable water in the propagation medium is calculated according to Eq. (2)25,26 where is the amount of precipitable water (unit: mm) and is the length of the propagation path of the electromagnetic field from the object to the thermal imager (unit: m).The transmittance of the propagating medium can be obtained by consulting the Passman and Larmore tables available in Refs. 22, 23, and 27. The required input data are the amount of precipitable water and the wavelength of interest. It is noteworthy that within the wavelength range of interest in predictive maintenance (7.5 to 14 μm), atmospheric attenuation may be considered independent of wavelength22 and dependent only on the amount of water precipitable, as the influence of is considered irrelevant. Therefore, the transmittance can be computed from three ascertainable variables: distance between the imager and the inspected object, temperature, and relative humidity of the propagation medium. The total radiation received by the thermal imager is made up of radiation from the inspected object, radiation from the other surfaces located in the environment and reflected on the inspected surface, and radiation from the column of air between the inspected surface and the thermal imager.28 For short distances, the radiation from the column of air is negligible.28 As a result, the radiation that will achieve the thermal imager comes only from the inspected object: radiation emitted by itself and reflected from the other surfaces in the ambient. Not considering the atmospheric transmittance, it can be computed as Eq. (3)28 where is the response of the thermal imager; is the absolute temperature of the object under inspection (unit: K); is the absolute temperature of the objects in the ambient (unit: K); is a coefficient, which is a function of the thermal imager, and includes some constants present in Planck’s law; is the inspected object’s spectral emissivity (dimensionless); represents the Planck radiation equation integrated over the spectral band of the imager, at temperature ; and is the wavelength analyzed (unit: μm).28The transmittance, in this case due to atmospheric absorption, is considered in Eq. (4)1,25 where is the power from the inspected object at a distance from the imager, is the radiation received by the imager, and is the transmittance of the medium between the imager and the object (dimensionless).For wavelengths from 7.5 to 14 μm and temperatures , a common situation in predictive maintenance, the function , can be expressed as Eq. (5)28 As the user informs all the parameters on the right side of Eq. (3), the coefficient is considered hypothetically unitary and a hypothetical response of the thermal imager is calculated. As a multiplying factor, this coefficient does not influence the calculation. Emissivity will be considered constant in the wavelength range and not temperature dependent over the measuring range.11 The nondependence of temperature considers that no significant superficial changes, which could affect its emissivity, occur to the inspected object.11 2.2.Uncertainty of the Temperature Measured by ThermographyTo calculate the uncertainty output, that is, the uncertainty of the temperature measured, a knowledge of the mathematical models involved in this process is required. With these models, one can apply the law of propagation of uncertainty from the inputs to the output. If the input variables are uncorrelated, Eq. (6) is applied. All variables are considered uncorrelated because this consideration yields negligible errors.10,11 where are the estimates of the input variables; is the combined standard uncertainty associated to the output variable ; and is the function that correlates input variables to output variable .The standard uncertainty of each measurement is calculated considering its uncertainty and resolution intervals, yielding its type B uncertainty. A rectangular (or uniform) distribution of possible values in these intervals is considered.3 The uncertainty of absolute humidity is calculated through Eq. (1). Uncertainty of precipitable water in turn is computed through Eq. (2). A function that correlates atmospheric transmittance to the amount of precipitable water is obtained from regression of the data in Passman and Larmore tables. The uncertainty of the transmittance is calculated by applying the law of propagation of uncertainty, Eq. (6), to this function. The power received by the thermal imager is estimated through Eqs. (3) and (5). The power from the inspected object is calculated from Eq. (4).1,25 The uncertainty of the power from the object is estimated by propagating the uncertainty of the transmittance through Eq. (4) and by propagating the uncertainty of emissivity through Eq. (3). The temperature uncertainty is estimated from the power from the inspected object and its uncertainty through Eqs. (3) and (5). Finally, the thermal imager’s uncertainty should be associated to this result by Eq. (6). 2.3.Contribution from Each Source to Temperature Measurement UncertaintyThe calculation of the contribution from each source of uncertainty to the measuring temperature uncertainty by thermography is carried out using the method available in Ref. 29, which is shown in Eq. (7). It should be remembered that the sources of uncertainty of interest are relative humidity, atmospheric temperature, propagation distance, surface emissivity of the inspected object, and the thermal imager. where is the Kessel coefficient of contribution of the uncertainty of input variable to the uncertainty of the output variable (dimensionless); is the partial derivate of the function with respect to input variable ; and is the coefficient of correlation between the output variable and the input variable .The coefficient of correlation co can be calculated according to Eq. (8):29 3.ResultsTo evaluate the quality of the results of the uncertainty calculation method proposed in this paper, a comparison is made with the adopted input parameters and the obtained output results available in Refs. 10 and 11. In these works, the authors had access to the function that relates the radiation received by the thermal imager to the temperature of the inspected object. They also had access to the calculation model of atmospheric transmittance adopted by the manufacturer of the thermal imager. As the method presented here does not consider any model reserved by camera manufacturers, it is expected that the results are similar in magnitude but not exactly the same. In order to compare our results to the ones available in Ref. 10, the values adopted for the input data in the simulation are listed in Table 1. Table 1Adopted values for simulations.
To evaluate the influence of the uncertainty of each input data in the uncertainty of the output, and to compare them to results shown in Ref. 10, the variation ranges of input uncertainty presented in Table 2 were adopted. Table 2Ranges of standard uncertainties of the inputs adopted for simulations.
The results of the sensitivity tests of the standard uncertainty of the temperature measured from the standard uncertainty of the input variables are shown in Figs. 2, 3, 4, 5, and 6. Next, we assess the contribution of each source of uncertainty to the final uncertainty of the temperature measurement considering the typical characteristics of the measurements performed by the thermographer. These characteristics are shown in Table 3 and were taken from catalogs of instruments commonly used in thermography. The standard uncertainty of each measurement is calculated considering its uncertainty and resolution intervals, yielding its type B uncertainty. A rectangular (or uniform) distribution of possible values in these intervals3 was considered. The input values adopted were the same as shown in Table 1, considering 60°C for the temperature of the object. The adopted standard uncertainty of emissivity values evaluated by the thermographer is 0.0015 or 0.030. These standard uncertainty of emissivity values are in accordance with expected uncertainty of emissivities close to the unity, between 1 and 4% of emissivity.30,31 Table 3Instruments used and measurements performed by the thermographer.
The results are shown in Tables 4 and 5. Table 4Contribution from each measurement at 0.015 standard uncertainty of emissivity.
Table 5Contribution from each measurement at 0.030 standard uncertainty of emissivity.
This calculation procedure was compiled in MATLAB 2011b®, producing the output data (variables listed in Table 4) from the instrument characteristics informed by the user (variables listed in Table 3). Comparison between our results and the results presented in Ref. 11 are presented in Table 6. The values adopted for the simulation are11 atmospheric and ambient temperatures 20°C; emissivity 0.8; relative humidity 50%; distance 50 m; and object’s temperature 90°C. Table 6Comparison of results for emissivity and ambient temperature uncertainties.
4.DiscussionThe results of the sensitivity tests for the uncertainty of the measured temperature by propagation from uncertainties of the input data (Figs. 2, 3, 4, 5, and 6), shown in Sec. 3, demonstrate that the method proposed in this paper provides results close to those shown in Ref. 10. They are developed from phenomenological and empirical models available to the customer and are similar to those developed from models reserved by manufacturers of thermal imagers and not available. Furthermore, the comparisons made to the work presented in Ref. 11 show very similar results for contributions from emissivity uncertainty. A discrepancy can be seen between the results for the uncertainty of emissivity of the inspected object when compared to the work presented in Ref. 10, despite being of the same order of magnitude. However, this result is in line with results presented in Ref. 11. As the calculation models used in Ref. 10 are not available, we are not able to discuss this difference. Sensitivity analysis also shows that the influence of distance, atmospheric and ambient temperatures, and relative humidity measurements is very low in comparison to the influence of the measurement of the emissivity of the inspected object. In fact, the results of the contribution from such sources shown in Tables 4 and 5 seem irrelevant to the uncertainty of the measurement temperature by thermography, under the specified conditions. Emissivity measurement and the intrinsic uncertainty of the thermal imager are the two major factors in the total uncertainty. From Tables 4 and 5, it can be seen that an increase in the standard uncertainty of emissivity from 0.015 to 0.030 nearly doubles the standard uncertainty of the measured temperature from 1.6 to 2.8°C if the imager has 1°C uncertainty. Less impactful but still significant, a thermal imager with 2°C uncertainty compared to a 1°C uncertainty thermal imager increases the uncertainty of the final result from 1.6 to 1.9°C, by , if the emissivity standard uncertainty is 0.015. 5.ConclusionThe alternative calculation method of uncertainty of temperature measurement by infrared thermography presented in this work showed results in line with those available in the literature. The proposed method can be used to give an indication of the quality of the temperature measurement performed from an estimate of its uncertainty for diagnosis of the equipment under inspection. Sensitivity analysis of the uncertainty of the temperature measurement from the uncertainties of the input data showed that of all sources considered, the uncertainties of emissivity and of the thermal imager are the most important factors. Emissivity, which is usually not measured but estimated by the thermographer according to data available in the literature, is the factor with the greatest impact. Therefore, the use of a thermal imager with low uncertainty and the careful and thoughtful evaluation of emissivity will lead to temperature measurements of a better quality, i.e., with lower uncertainty. The results indicate that under normal conditions the standard uncertainty of the temperature measured can vary from to 3.0°C. Therefore, the use of a thermal imager with a high level of uncertainty (2°C) or high standard uncertainty of emissivity (0.03), which yields a standard uncertainty of the measured temperature in the range of 1.9 to 3.0°C in established conditions, may lead to misdiagnosis. The acceptance criterion is in predictive maintenance. As estimating the uncertainty of temperature measurement is shown here to be practicable, and in compliance with technical recommendations,3 manufacturers of thermal imagers and software developers should provide this function for users. ReferencesM. VollmerK.-P. Möllmann, Infrared Thermal Imaging, Wiley-VCH, Weinheim
(2010). Google Scholar
K. Chrzanowski, Non-Contact Thermometry: Measurement Errors, Warsaw(2001). Google Scholar
“Evaluation of measurement data—guide to the expression of uncertainty in measurement,”
Google Scholar
K. ChrzanowskiJ. FischerR. Matyszkiel,
“Testing and evaluation of thermal cameras for absolute temperature measurement,”
Opt. Eng., 39
(9), 2535
–2544
(2000). http://dx.doi.org/10.1117/1.1287832 OPEGAR 0091-3286 Google Scholar
“International vocabulary of metrology: basic and general concepts and associated terms,”
Google Scholar
G. GrgićI. Pušnik,
“Analysis of thermal imagers,”
Int. J. Thermophys., 32
(1–2), 237
–247
(2011). http://dx.doi.org/10.1007/s10765-011-0932-z IJTHDY 0195-928X Google Scholar
K. ChrzanowskiR. Matyszkiel,
“Software package for uncertainty calculations of temperature measurements with thermal cameras,”
97
–102
(2000) http://qirt.gel.ulaval.ca/archives/qirt2000/papers/055.pdf Google Scholar
K. Chrzanowskiet al.,
“Uncertainty of temperature measurement with thermal cameras,”
Opt. Eng., 40
(6), 1106
–1114
(2001). http://dx.doi.org/10.1117/1.1367351 OPEGAR 0091-3286 Google Scholar
W. MinkinaS. Dudzik, Infrared Thermography: Errors and Uncertainties, 1st ed.Wiley, Noida
(2009). Google Scholar
W. MinkinaS. Dudzik,
“Simulation analysis of uncertainty of infrared camera measurement and processing path,”
Measurement, 39
(8), 758
–763
(2006). http://dx.doi.org/10.1016/j.measurement.2006.03.001 MSRMDA 0263-2241 Google Scholar
B. Laneet al.,
“Uncertainty of temperature measurements by infrared thermography for metal cutting applications,”
Metrologia, 50
(6), 637
–653
(2013). http://dx.doi.org/10.1088/0026-1394/50/6/637 MTRGAU 0026-1394 Google Scholar
M. S. JadinS. Taib,
“Recent progress in diagnosing the reliability of electrical equipment by using infrared thermography,”
Infrared Phys. Technol., 55
(4), 236
–245
(2012). http://dx.doi.org/10.1016/j.infrared.2012.03.002 IPTEEY 1350-4495 Google Scholar
B. R. Lyon Jr.G. L. OrloveD. L. Peters,
“The relationship between current load and temperature for quasi-steady state and transient conditions,”
Proc. SPIE, 4020 62
–70
(2000). http://dx.doi.org/10.1117/12.381580 PSISDG 0277-786X Google Scholar
“ANSI/NETA MTS-2011 standard for maintenance testing specifications for electrical power equipment and systems,”
(2011). Google Scholar
FLIR Systems, User’s Manual FLIR InfraCAM SD, 2008,
(2014) http://www.alpine-components.co.uk/files/manuals-downloads/FLIR-InfraCAM-SD-Manual.pdf June ). 2014). Google Scholar
“ANSI/AAMI DF80—medical electrical equipment,”
(2010). Google Scholar
N. D. S. F. Duarte Jr.,
“Management system of measurement: important, but not always recognized,”
Metrol. Instrumentação, 7
(56),
(2008). Google Scholar
FLIR Systems Inc., “FLIR customer support center,”
(2013) http://flir.custhelp.com/ June ). 2013). Google Scholar
Fluke Corporation., “Fluke thermal imagers,”
(2013) http://www.fluke.com/fluke/usen/products/categoryti.htm June ). 2013). Google Scholar
A. G. Testo,
“Measuring instruments for the maintenance of electrical installations,”
(2014) http://www.testo.us/en/home/products/thermography/maintenance/electrical_maintenance/electrical_maintenance.jsp June ). 2014). Google Scholar
InfraRed Integrated Systems Ltd., “Thermal cameras—thermography products,”
(2013) http://www.irisys.co.uk/thermal-imaging/products/ June ). 2013). Google Scholar
G. Gaussorgues, Infrared Thermography, 3rd ed.Chapman & Hall, London
(1994). Google Scholar
S. PassmanL. Larmore, Atmospheric Transmission, The RAND Corporation, Santa Monica
(1956). Google Scholar
V. R. StullP. J. WyattG. N. Plass, The Infrared Absorption of Carbon Dioxide, Ford Motor Company, New Port Beach
(1963). Google Scholar
R. SabatiniM. A. Richardson, Airborne Laser Systems Testing and Analysis, North Atlantic Treaty Organisation, Virginia
(2010). Google Scholar
D. J. C. Coura,
“Analysis of the major effects of the propagation of optical signals in free space, coated with a numerical efficient platform,”
INATEL,
(2004). Google Scholar
P. J. WyattV. R. StullG. N. Plass,
“The infrared absorption of water vapor,”
(2013) http://www.dtic.mil/dtic/tr/fulltext/u2/297458.pdf May ). 2013). Google Scholar
F. J. M. MecaF. J. R. SanchezP. M. Sanchez,
“Calculation and optimisation of the maximum uncertainty in infrared temperature measurements taken in conditions of high uncertainty in the emissivity and environment radiation values,”
Infrared Phys. Technol., 43
(6), 367
–375
(2002). http://dx.doi.org/10.1016/S1350-4495(02)00125-1 IPTEEY 1350-4495 Google Scholar
R. KesselR. KackerM. Berglund,
“Coefficient of contribution to the combined standard uncertainty,”
Metrologia, 43
(4), S189
–S195
(2006). http://dx.doi.org/10.1088/0026-1394/43/4/S04 MTRGAU 0026-1394 Google Scholar
J. IshiiA. Ono,
“Uncertainty estimation for emissivity measurements near room temperature with a Fourier transform spectrometer,”
Meas. Sci. Technol., 12
(12), 2103
–2112
(2001). http://dx.doi.org/10.1088/0957-0233/12/12/311 MSTCEP 0957-0233 Google Scholar
M. R. MillerG. MulhollandC. Anderson,
“Experimental cutting tool temperature distributions,”
J. Manuf. Sci. Eng., 125
(4), 667
(2003). http://dx.doi.org/10.1115/1.1621425 1087-1357 Google Scholar
Minipa, “MT-240 thermohygrometers,”
(2014) http://www.minipa.com.br/5/21/159-Minipa-Termo-higrometros-MT-240 March ). 2014). Google Scholar
Fluke Corporation, “56X infrared thermometer,”
(2014) http://media.fluke.com/documents/56x_____gseng0000.pdf March ). 2014). Google Scholar
Fluke Corporation, “Fluke 424D, 419D and 414D laser distance meters,”
(2014) http://media.fluke.com/documents/3028711_0000_ENG_E_W.PDF March ). 2014). Google Scholar
FLIR Systems Inc, “FLIR® ex-series infrared camera specifications,”
(2014) http://www.flir.com/thermography/americas/us/view/?id=61194&collectionid=830&col=61196 March ). 2014). Google Scholar
FLIR, “Technical data FLIR P660,”
(2014) http://support.flir.com/DsDownload/Assets/40402-3001_en_50.pdf May ). 2014). Google Scholar
BiographyPablo Rodrigues Muniz received his BSc and MSc degrees in electrical engineering and mechanical engineering, respectively, from the Universidade Federal do Espírito Santo, Brazil, in 2002 and 2006, respectively. He is currently working toward his Dr. degree at Universidade Federal da Bahia, Salvador, BA, Brazil. He is with Instituto Federal do Espírito Santo, Vitória, ES, Brazil, as a professor. He had worked as maintenance specialist at electrical shop and engineering division at ArcelorMittal Tubarão. Ricardo de Araújo Kalid received his BSc and MSc degrees from Universidade Federal da Bahia in 1988 and 1991, respectively, and received his PhD degree in 1999 from Universidade de São Paulo, all in chemical engineering. He has experience in industrial and chemical engineering. He coordinates several research efforts concerning industrial processes. He is currently professor at Universidade Federal da Bahia and dean Pro Tempore of Universidade Federal do Sul da Bahia. Shirley P. N. Cani received her BSc, MSc, and PhD degrees, all in electrical engineering, from the Universidade Federal do Espírito Santo, Vitória, Brazil, in 2001, 2003, and 2007, respectively. During her MSc and PhD studies, she was involved in optical communication research. Since 2008, she has been with the Electrotechnical Department, Instituto Federal do Espírito Santo. Robson da Silva Magalhães received his BSc, MSc, and DSc degrees in mechanical engineering, electrical engineering, and industrial engineering, respectively, from the Universidade Federal da Bahia, Salvador, Brazil, in 1985, 2005, and 2010, respectively. He has experience in industrial maintenance, emphasizing on vibration and motor current signature analysis. He is currently a professor with the Chemical Engineering Department, Universidade Federal da Bahia. |