14 November 2022 Optical calibration system for refractometers
Patrick O'Neal, Noam Sapiens
Author Affiliations +
Abstract

An optical calibration system for refractometers was designed and built using a singlet lens and a translatable detector that induces spherical error, making it a simple and easy-to-use system. The system was calibrated using spectacles and then validated by correcting blurred images created by a series of ophthalmic test lenses. A linear regression model of the correction of the ophthalmic test lenses showed nearly direct correlation with an offset of 0.12 D and an R2 value of 0.9998. An updated version of the system was designed in Zemax OpticStudio with a singlet lens, which could induce spherical error accurate to that of the human eye within ≤ 0.05 D. In future work, this system will be manufactured and tested in the lab.

© 2022 Society of Photo-Optical Instrumentation Engineers (SPIE)
Patrick O'Neal and Noam Sapiens "Optical calibration system for refractometers," Optical Engineering 61(12), 121807 (14 November 2022). https://doi.org/10.1117/1.OE.61.12.121807
Received: 6 June 2022; Accepted: 27 October 2022; Published: 14 November 2022