19 September 1985 Acceptance Testing And Calibration Of Digital Subtraction Angiography Systems
Pei-Jan Paul Lin, Arthur Olson
Author Affiliations +
Abstract
Based on an acceptance testing procedure by means of the American Association of Physicists ini Medicine (AAPM) digital subtraction angiography (DSA) phantom described previously , experiences were gained with respect to the expected level of system performance and the image quality reproducible for typical DSA systems currently installed in the clinical environment. There were more than 20 DSA systems involved in this study and encompassed several DSA manufacturers' products. A DSA system is employed as an example to highlight the importance of the initial calibration of the DSA imaging chain, and to demonstrate the differences in the image quality obtainable for the same system.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Pei-Jan Paul Lin and Arthur Olson "Acceptance Testing And Calibration Of Digital Subtraction Angiography Systems", Proc. SPIE 0555, Medical Imaging and Instrumentation '85, (19 September 1985); https://doi.org/10.1117/12.949488