19 September 1985 Comparison Of Chest Filters
G. Schiller, A. Olson, H. Nguyen
Author Affiliations +
Abstract
Recently, anatomically shaped lead acrylic filters have been introduced for chest radiography. Two of these filters were compared to several (Al, Cu, Y, and Pb acrylic) uniform filters. Phototimed exposures at 80, 100, 120 and 140 kVp were made on a realistic chest phantom. The optical density in the lung field was kept constant for all filters and kVp's. Exposure time and entrance exposures to the mediastinum and lungs were measured. When compared to standard (3.2 mm Al HVL) aluminum filtration a reduction of (44% - 60%) of lung exposure and better visualization of the mediastinum and retrocardiac areas were noted. However, a significant (80 - 350%) increase in exposure time was required and mediastinum exposure increased by 40 - 100 percent. When using uniform filters, in addition to the standard aluminum filter, entrance exposures to the lung and mediastinum were reduced by 30 -50% for Yttrium and Copper and by 27 to 35% for lead acrylic. Exposure times increased by up to 36%, 64%, and 52% respectively. When using spatially shaped filters, improved image quality and reduced lung exposure results, however, one must be aware of the significant increase in exposure time especially at low (80) kVp's. There is also an increase of medias-tinum exposure and possibility of positioning artifacts.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. Schiller, A. Olson, and H. Nguyen "Comparison Of Chest Filters", Proc. SPIE 0555, Medical Imaging and Instrumentation '85, (19 September 1985); https://doi.org/10.1117/12.949466
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Aluminum

Lung

Lead

Image filtering

Chest

Copper

Yttrium

Back to Top