Paper
9 November 1977 Short Wavelength Rocketborne Infrared Spectrometer
Clair L. Wyatt, D. Gary Frodsham
Author Affiliations +
Abstract
An ir spectrometer, used to obtain measurements of faint atmospheric emission spectra from a rocket carrier, is described. The sensor scans the 2.07 to 5.47 μm spectral region at the rate of 2 scan/s. The spectral resolution of the sensor, which employs a circular variable filter (CVF), ranges from 1.0 to 3.0%. The entire optical subsection, indium anti-monide detector, CVF, silicon lens, baffle, and removable cover are cooled below 80° K in a high-vacuum Dewar system. The noise equivalent spectral sterance [radiance] of the spectrometer at 5.3 μm is 1.93 x 10-10 (W cm-2sr-1μm-1).
© (1977) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Clair L. Wyatt and D. Gary Frodsham "Short Wavelength Rocketborne Infrared Spectrometer", Proc. SPIE 0124, Modern Utilization of Infrared Technology III, (9 November 1977); https://doi.org/10.1117/12.955865