Paper
1 March 1992 Calibration of a Fourier transform spectrometer using three black body sources
Volker Tank, Erwin Lindermeir, H. Dietl
Author Affiliations +
Proceedings Volume 1575, 8th Intl Conf on Fourier Transform Spectroscopy; (1992) https://doi.org/10.1117/12.56304
Event: Eighth International Conference on Fourier Transform Spectroscopy, 1991, Lubeck-Travemunde, Germany
Abstract
A procedure to calibrate a Fourier transform spectrometer is presented. Blackbody sources of three different temperatures are used to eliminate errors in the calibration arising from the limited accuracy of the temperature measurement of the calibration sources. With three spectra of blackbodies it is possible to assume that the temperatures are unknown variables, as are the parameters of the functions describing the spectrometer. Nonlinear Gaussian balancing calculation is applied to determine these unknown variables and to minimize the influence of noise. A comparison between results obtained with this method and a conventional calibration procedure is presented.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Volker Tank, Erwin Lindermeir, and H. Dietl "Calibration of a Fourier transform spectrometer using three black body sources", Proc. SPIE 1575, 8th Intl Conf on Fourier Transform Spectroscopy, (1 March 1992); https://doi.org/10.1117/12.56304
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Black bodies

Spectroscopy

Temperature metrology

Fourier transforms

Gases

Mathematical modeling

Back to Top