1 May 1994 New method of data acquisition in k-space in magnetic resonance imaging
Ramesh Chandra, Henry Rusinek
Author Affiliations +
Abstract
We propose a new MRI acquisition method where data are collected from multiple directions in k-space. Its principal advantages are the reduction in acquisition time and the improvement of the signal to noise ratio. The potential disadvantage of the method is a loss in resolution. We have implemented the method and compared it with conventional truncated Fourier imaging using simulated objects, physical phantoms and human brains. The results show noise reduction ranging 25 - 40%. For brain images the loss of resolution is difficult to detect. The use of the method is recommended in applications where the signal to noise is poor, such as image subtraction in functional imaging.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ramesh Chandra and Henry Rusinek "New method of data acquisition in k-space in magnetic resonance imaging", Proc. SPIE 2163, Medical Imaging 1994: Physics of Medical Imaging, (1 May 1994); https://doi.org/10.1117/12.174278
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Data acquisition

Brain

Neuroimaging

Magnetic resonance imaging

Image resolution

Medical imaging

Physics

Back to Top