Paper
26 November 1996 Method for near-real-time continuous air monitoring of phosgene, hydrogen cyanide, and cyanogen chloride
Frank Gary Lattin, Donald G. Paul
Author Affiliations +
Abstract
A sorbent-based gas chromatographic method provides continuous quantitative measurement of phosgene, hydrogen cyanide, and cyanogen chloride in ambient air. These compounds are subject to workplace exposure limits as well as regulation under terms of the Chemical Arms Treaty and Title III of the 1990 Clean Air Act amendments. The method was developed for on-sit use in a mobile laboratory during remediation operations. Incorporated into the method are automated multi-level calibrations at time weighted average concentrations, or lower. Gaseous standards are prepared in fused silica lined air sampling canisters, then transferred to the analytical system through dynamic spiking. Precision and accuracy studies performed to validate the method are described. Also described are system deactivation and passivation techniques critical to optimum method performance.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Frank Gary Lattin and Donald G. Paul "Method for near-real-time continuous air monitoring of phosgene, hydrogen cyanide, and cyanogen chloride", Proc. SPIE 2835, Advanced Technologies for Environmental Monitoring and Remediation, (26 November 1996); https://doi.org/10.1117/12.259771
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Cyanide

Hydrogen

Sensors

Instrument modeling

Capillaries

Gases

Back to Top