Paper
9 December 1999 Raman process analyzer calibration transfer through analyzer standardization
Mohamed Shaheen, Ramasamy Manoharan
Author Affiliations +
Abstract
As Raman spectroscopy finds increased utilization for on- line measurement, the issues of analyzer stability and calibration transferability grow in importance. This paper present a method that allows calibrations developed on one analyzer to be transferred to others located at the same or different sites. This method also allows maintenance of calibrations in a single Raman analyzer over time even when critical components are altered. Calibration transfer eliminates the need to develop calibrations for different analyzers and upon analyzer repair, saving time and money in calibration model development and maintenance.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mohamed Shaheen and Ramasamy Manoharan "Raman process analyzer calibration transfer through analyzer standardization", Proc. SPIE 3859, Optical Online Industrial Process Monitoring, (9 December 1999); https://doi.org/10.1117/12.372938