Paper
18 April 2000 Practical quality control standards for digital display monitors
John F. Copeland, Christopher S. Melhus, Roxann C. Horton, Vaidehi Venkatakrishnan, Robert G. Zamenhof, Jeffrey B. Mendel
Author Affiliations +
Abstract
We present two-tiered QC standards for grayscale monitors. Materials: SMPTE test pattern, line-pair object, radio- opaque ruler, photometer, and magnifier with scale. Results: The following standards are based on our institutional experience in association with a PACS vendor and the literature. Radiologists monitor: small artifacts, disregard <EQ 0.25 mm, 0.25 mm < 2 < 0.35 mm, 0 >= 0.35 mm; no groupings of artifacts; no visible distortion in the 'usable' area; background luminance 0.1 Fl; 'bright' >= 90 Fl, no fine-detail 'blooming'; brightness uniformity <EQ +/- 15 percent of mean value, no visible gradients, measured in at least 9 positions; scales 98 percent; 2.5 linepair/mm visible; 0/5-95/100 SMPTE contrast patches visible; illuminance 50 lux. Clinician monitor; small artifacts disregard <EQ 0.35 mm < 4 < 0.50 mm, 0 >= 0.50 mm; luminance 50 Fl; uniformity <EQ +/- 20 percent, if > +/- 20 percent at only 2 points, these locations can be tested to +/- 20 percent within the local sector; 1 linepair/mm visible; else same. Conclusions: We believe this approach is reasonably achievable.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John F. Copeland, Christopher S. Melhus, Roxann C. Horton, Vaidehi Venkatakrishnan, Robert G. Zamenhof, and Jeffrey B. Mendel "Practical quality control standards for digital display monitors", Proc. SPIE 3976, Medical Imaging 2000: Image Display and Visualization, (18 April 2000); https://doi.org/10.1117/12.383071
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Picture Archiving and Communication System

Standards development

Mammography

Radiology

Calibration

Control systems

Data centers

RELATED CONTENT


Back to Top