Paper
25 April 2000 Cone-beam CT with a flat-panel imager: noise considerations for fully 3D computed tomography
Jeffrey H. Siewerdsen, David A. Jaffray
Author Affiliations +
Abstract
Application of flat-panel imagers (FPIs) in cone-beam computed tomography (CBCT) offers a promising new modality for full three-dimensional (3-D) x-ray imaging. Understanding the potential performance and fundamental limitations of such technology, however, requires knowledge of the noise characteristics of the 3-D imaging system. The noise performance of a prototype flat-panel cone-beam CT (FPI-CBCT) system is investigated empirically and theoretically in terms of voxel noise, noise-power spectrum (NPS), and detective quantum efficiency (DQE). Methods for NPS analysis common in characterizing 2-D imagers are extended to the fully 3-D case, and a general framework for NPS analysis in n dimensions is presented, including the important considerations of NPS convergence and normalization within the constraints of system linearity and stationarity. Factors affecting imager noise and NPS are numerous, including exposure, number of views, image blur, additive noise, reconstruction filter, and sampling matrix. The applicability of existing theoretical descriptions of CT voxel noise is examined. A theoretical cascaded systems model that accurately predicts the 2-D noise characteristics of the FPI is extended to describe the signal and noise transfer characteristics of the fully 3-D FPI-CBCT system employing filtered back-projection. Analysis of the fully 3-D NPS reveals features of the 3-D imaging system that might otherwise be missed and shows the significant effect of 3-D noise aliasing. Furthermore, it quantifies the performance of various FPI designs in CBCT (e.g., direct and indirect detectors, and variations therein) and provides a guide for the development of high-performance FPI-CBCT systems suited to specific clinical objectives.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jeffrey H. Siewerdsen and David A. Jaffray "Cone-beam CT with a flat-panel imager: noise considerations for fully 3D computed tomography", Proc. SPIE 3977, Medical Imaging 2000: Physics of Medical Imaging, (25 April 2000); https://doi.org/10.1117/12.384515
Lens.org Logo
CITATIONS
Cited by 22 scholarly publications and 21 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Imaging systems

3D image processing

Computed tomography

Sensors

X-ray computed tomography

Systems modeling

Therapeutics