Paper
18 July 2007 Non-ergodicity correction in laser speckle biomedical imaging
Author Affiliations +
Abstract
We discuss a new approach to laser speckle biomedical imaging with the goal to establish a quantitative link between the measured signal and the local dynamic properties of Brownian motion or blood flow. We demonstrate that the presence of a static component in laser speckle imaging signal can significantly complicate the quantitative interpretation of the imaging data. With Monte-Carlo simulations and model experiments we show that the error in the mean particles velocity extracted using traditional approaches can reach several orders of magnitude. With a proper data treatment on the other side the error can be substantially reduced. We suggest a simple data processing scheme that properly accounts for a static component in the scattered light intensity.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
P. Zakharov, A. C. Völker, F. Buck, B. Weber, and F. Scheffold "Non-ergodicity correction in laser speckle biomedical imaging", Proc. SPIE 6631, Novel Optical Instrumentation for Biomedical Applications III, 66310D (18 July 2007); https://doi.org/10.1117/12.727646
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Light scattering

Scattering

Speckle

Monte Carlo methods

Correlation function

Biomedical optics