Paper
19 February 2009 3D in vivo swept source endoscopic OCT based on rigid GRIN lens rod probe
Author Affiliations +
Abstract
In this manuscript, we report a fast speed swept source endoscopic OCT system utilizing a 4.5mm diameter rigid GRIN lens rod probe. The rigid probe has a tunable working distance with tuning range 0~7mm. The system could take front scanning images with speed of 40frames per second (512 A line per frame). The probe can work in contact and non contact mode. Fast speed contact and non-contact imaging was demonstrated.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gangjun Liu, Tuqiang Xie, Jun Zhang, George M. Peavy, Matthew Brenner, and Zhongping Chen "3D in vivo swept source endoscopic OCT based on rigid GRIN lens rod probe", Proc. SPIE 7168, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XIII, 716812 (19 February 2009); https://doi.org/10.1117/12.809624
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

GRIN lenses

In vivo imaging

Endoscopy

3D image processing

Mirrors

Signal to noise ratio

RELATED CONTENT


Back to Top