Paper
12 March 2009 Clinical validation of a medical grade color monitor for chest radiology
J. Jacobs, F. Zanca, J. Verschakelen, G. Marchal, H. Bosmans
Author Affiliations +
Abstract
Until recently, the specifications of medical grade monochrome LCD monitors outperformed those of color LCD monitors. New generations of color LCD monitors, however, show specifications that are in many respects similar to those of monochrome monitors typically used in diagnostic workstations. The aim of present study was to evaluate the impact of different medical grade monitors in terms of detection of simulated lung nodules in chest x-ray images. Specifically, we wanted to compare a new medical grade color monitor (Barco Coronis 6MP color) to a medical grade grayscale monitor (Barco Coronis 3MP monochrome) and a consumer color monitor (Philips 200VW 1.7MP color) by means of an observer performance experiment. Using the free-response acquisition data paradigm, seven radiologists were asked to detect and locate lung nodules (170 in total), simulated in half of the 200 chest X-ray images used in the experiment. The jackknife free-response receiver operating characteristic (JAFROC) analysis of the data showed a statistically significant difference between at least two monitors, F-value=3.77 and p-value =0.0481. The different Figure of Merit values were 0.727, 0.723 and 0.697 for the new color LCD monitor, the medical grade monitor and the consumer color monitor respectively. There was no difference between the needed reading times but there was a difference between the mean calculated Euclidian distances between the position marked by the observers and the center of the simulated nodule, indicating a better accuracy with both medical grade monitors. Present data suggests that the new generation of medical grade color monitors could be used as diagnostic workstations.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Jacobs, F. Zanca, J. Verschakelen, G. Marchal, and H. Bosmans "Clinical validation of a medical grade color monitor for chest radiology", Proc. SPIE 7263, Medical Imaging 2009: Image Perception, Observer Performance, and Technology Assessment, 726308 (12 March 2009); https://doi.org/10.1117/12.813702
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
LCDs

Lung

Chest

Diagnostics

Chest imaging

Radiology

Statistical analysis

Back to Top