Paper
17 May 2016 Evaluation of the repeatability of a Landolt-C based automated sensor resolution assessment methodology
Alan R. Pinkus, David W. Dommett, H. Lee Task
Author Affiliations +
Abstract
This paper describes a novel sensor resolution assessment chart and procedure based on the Landolt C. This automated resolution assessment procedure does not rely on human vision as the primary means of interpreting resolution chart results. Seven sensors across four different spectral bands and several geometric resolutions were assessed a total of five times each to determine the repeatability (confidence interval) for this automated procedure. The results are presented and compared with previous studies.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alan R. Pinkus, David W. Dommett, and H. Lee Task "Evaluation of the repeatability of a Landolt-C based automated sensor resolution assessment methodology", Proc. SPIE 9842, Signal Processing, Sensor/Information Fusion, and Target Recognition XXV, 98420Z (17 May 2016); https://doi.org/10.1117/12.2222579