1 January 2009 Combination of different noninvasive measuring techniques: a new approach to increase accuracy of peripheral near infrared spectroscopy
Gerhard Pichler, Karin Grossauer, Elisabeth Peichl, Andrea Gaster, Andrea Berghold, Gerold Schwantzer, Heinz Zotter, William Müller, Berndt Urlesberger
Author Affiliations +
Abstract
One of the problems of near-infrared-spectroscopy (NIRS) measurements is low reproducibility. The aim of the present study was to introduce quality criteria to increase reproducibility of peripheral NIRS measurements. In a prospective cohort study in 40 neonates, repeated NIRS measurements were performed on the calf. During five "reapplication" periods (of NIRS optodes), five "measurements" (venous occlusions) were performed. Tissue oxygenation index (TOI), mixed venous oxygenation (SvO2), fractional oxygen extraction (FOE), hemoglobin flow (Hbflow), oxygen delivery (DO2), and oxygen consumption (VO2) were assessed. Measurements with linear changes during venous occlusions were included for further analysis (first quality criterion: R2>0.95). The second quality criterion was the equation 0 ≤ TOI-SvO2 ≤(SaO2-SvO2)×0.2. Variance components and mean standard deviations were analyzed after introduction of the quality criteria. Variance components of reapplication and measurement decreased after introduction of the second quality criterion (TOI: 46.6–35.0%, SvO2: 76.8–38.2%, FOE: 73.1–37.5%, Hbflow: 70.3–51.9%, DO2: 71.5–52.7%, and VO2: 70.9–63.8%). Mean standard deviations of TOI (6.6±3.0 to 4.7±3.2%), SvO2 (11.1±4.8 to 5.7±3.9%), FOE (11.3±4.8 to 5.9±4.0%), Hbflow (4.3±2.0 to 2.9±1.6 μmol/100 mL/min), and DO2 (17.8±7.6 to 11.4±6.2 μmol/100 mL/min) decreased significantly, too. Only 12% of measurements fulfilled both quality criteria. With the introduction of two quality criteria, test–retest variability of peripheral NIRS measurements decreased significantly and reproducibility increased significantly.
©(2009) Society of Photo-Optical Instrumentation Engineers (SPIE)
Gerhard Pichler, Karin Grossauer, Elisabeth Peichl, Andrea Gaster, Andrea Berghold, Gerold Schwantzer, Heinz Zotter, William Müller, and Berndt Urlesberger "Combination of different noninvasive measuring techniques: a new approach to increase accuracy of peripheral near infrared spectroscopy," Journal of Biomedical Optics 14(1), 014014 (1 January 2009). https://doi.org/10.1117/1.3076193
Published: 1 January 2009
Lens.org Logo
CITATIONS
Cited by 21 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Near infrared spectroscopy

Oxygen

Tissues

Sensors

Capillaries

Oximetry

Skin

Back to Top