1 September 2009 Reduction of background in optoacoustic image sequences obtained under tissue deformation
Michael Jaeger, Lea Siegenthaler, Michael Kitz, Martin Frenz
Author Affiliations +
Abstract
For real-time optoacoustic imaging of the human body, a linear array transducer and reflection mode optical irradiation is preferably used. Experimental outcomes however revealed that such a setup results in significant image background, which prevents imaging structures at the ultimate depth limited only by the optical attenuation of the irradiating light and the signal noise level. Various sources of image background, such as bulk tissue absorption, reconstruction artifacts, and backscattered ultrasound, could be identified. To overcome these limitations, we developed a novel method that results in significantly reduced background and increased imaging depth. For this purpose, we acquire, in parallel, a series of optoacoustic and echo-ultrasound images while the tissue sample is gradually deformed by an externally applied force. Optoacoustic signals and background signals are differently affected by the deformation and can thus be distinguished by image processing. This method takes advantage of a combined optoacoustic/echo-ultrasound device and has a strong potential for improving real-time optoacoustic imaging of deep tissue structures.
©(2009) Society of Photo-Optical Instrumentation Engineers (SPIE)
Michael Jaeger, Lea Siegenthaler, Michael Kitz, and Martin Frenz "Reduction of background in optoacoustic image sequences obtained under tissue deformation," Journal of Biomedical Optics 14(5), 054011 (1 September 2009). https://doi.org/10.1117/1.3227038
Published: 1 September 2009
Lens.org Logo
CITATIONS
Cited by 46 scholarly publications and 7 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissues

Transducers

Tissue optics

Natural surfaces

Acoustics

Speckle

Signal attenuation

Back to Top