1 May 2010 Modeling the corneal birefringence of the eye toward the development of a polarimetric glucose sensor
Author Affiliations +
Abstract
Optical polarimetry for monitoring glucose concentration in the aqueous humor of the eye as a potential noninvasive means of assessing blood glucose has promise, but the realization of such an approach has been limited by noise from time-varying corneal birefringence due to motion artifact. Modeling the corneal birefringence of the eye is critically important toward understanding the overall effect of this noise source compared to other changes in the signal, and can aid in design of the polarimetric system. To this end, an eye model is introduced in this work that includes spatially varying birefringence properties of the cornea. The degree of birefringence and the fast axis orientation is calculated as a function of beam position on the anterior chamber. It is shown that the minimum change in polarization vector orientation occurs for beam position near the midpoint between the corneal apex and limbus. In addition, the relative wavelength independence of motion artifact is shown in the same region. The direct consequence of these findings are that a multiwavelength polarimetric system can potentially be utilized to eliminate the effect of time-varying corneal birefringence, and that eye coupling is optimal at the midpoint between the apex and limbus.
©(2010) Society of Photo-Optical Instrumentation Engineers (SPIE)
Bilal H. Malik and Gerard L. Coté "Modeling the corneal birefringence of the eye toward the development of a polarimetric glucose sensor," Journal of Biomedical Optics 15(3), 037012 (1 May 2010). https://doi.org/10.1117/1.3447923
Published: 1 May 2010
Lens.org Logo
CITATIONS
Cited by 34 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Birefringence

Glucose

Eye

Cornea

Eye models

Polarization

Polarimetry

Back to Top