Open Access
1 May 2014 Noninvasive optical cytochrome c oxidase redox state measurements using diffuse optical spectroscopy
Jangwoen Lee, Jae Gwan Kim, Sari B. Mahon, David Mukai, David Yoon, Gerry R. Boss, Steven E. Patterson, Gary Rockwood, Gary Isom, Matthew Brenner M.D.
Author Affiliations +
Abstract
A major need exists for methods to assess organ oxidative metabolic states in vivo. By contrasting the responses to cyanide (CN) poisoning versus hemorrhage in animal models, we demonstrate that diffuse optical spectroscopy (DOS) can detect cytochrome c oxidase (CcO) redox states. Intermittent decreases in inspired O 2 from 100% to 21% were applied before, during, and after CN poisoning, hemorrhage, and resuscitation in rabbits. Continuous DOS measurements of total hemoglobin, oxyhemoglobin, deoxyhemoglobin, and oxidized and reduced CcO from muscle were obtained. Rabbit hemorrhage was accomplished with stepwise removal of blood, followed by blood resuscitation. CN treated rabbits received 0.166  mg/min NaCN infusion. During hemorrhage, CcO redox state became reduced concurrently with decreases in oxyhemoglobin, resulting from reduced tissue oxygen delivery and hypoxia. In contrast, during CN infusion, CcO redox state decreased while oxyhemoglobin concentration increased due to CN binding and reduction of CcO with resultant inhibition of the electron transport chain. Spectral absorption similarities between hemoglobin and CcO make noninvasive spectroscopic distinction of CcO redox states difficult. By contrasting physiological perturbations of CN poisoning versus hemorrhage, we demonstrate that DOS measured CcO redox state changes are decoupled from hemoglobin concentration measurement changes.
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jangwoen Lee, Jae Gwan Kim, Sari B. Mahon, David Mukai, David Yoon, Gerry R. Boss, Steven E. Patterson, Gary Rockwood, Gary Isom, and Matthew Brenner M.D. "Noninvasive optical cytochrome c oxidase redox state measurements using diffuse optical spectroscopy," Journal of Biomedical Optics 19(5), 055001 (1 May 2014). https://doi.org/10.1117/1.JBO.19.5.055001
Published: 1 May 2014
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications and 1 patent.
Advertisement
Advertisement
KEYWORDS
Oxygen

Tissues

Absorption

Blood

Tissue optics

Scattering

Near infrared

Back to Top