2 July 2021 Sub-millimeter endoscope demonstrates feasibility of in vivo reflectance imaging, fluorescence imaging, and cell collection in the fallopian tubes
Ricky Cordova, Kelli Kiekens, Susan Burrell, William Drake, Zaynah Kmeid, Photini Rice, Andrew Rocha, Sebastian Diaz, Shigehiro Yamada, Michael Yozwiak, Omar L. Nelson, Gustavo C. Rodriguez, John Heusinkveld, Ie-Ming Shih, David S. Alberts, Jennifer K. Barton
Author Affiliations +
Abstract

Significance: Most cases of high-grade serous ovarian carcinoma originate as serous tubal intraepithelial carcinoma (STIC) lesions in the fallopian tube epithelium (FTE), enabling early endoscopic detection.

Aim: The cell-acquiring fallopian endoscope (CAFE) was built to meet requirements for locating potentially pathological tissue indicated by an alteration in autofluorescence or presence of a targeted fluorophore. A channel was included for directed scrape biopsy of cells from regions of interest.

Approach: Imaging resolution and fluorescence sensitivity were measured using a standard resolution target and fluorescence standards, respectively. A prototype was tested in ex vivo tissue, and collected cells were counted and processed.

Results: Measured imaging resolution was 88  μm at a 5-mm distance, and full field of view was ∼45  deg in air. Reflectance and fluorescence images in ex vivo porcine reproductive tracts were captured, and fit through human tracts was verified. Hemocytometry counts showed that on the order of 105  cells per scrape biopsy could be collected from ex vivo porcine tissue.

Conclusions: All requirements for viewing STIC in the FTE were met, and collected cell counts exceeded input requirements for relevant analyses. Our benchtop findings suggest the potential utility of the CAFE device for in vivo imaging and cell collection in future clinical trials.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Ricky Cordova, Kelli Kiekens, Susan Burrell, William Drake, Zaynah Kmeid, Photini Rice, Andrew Rocha, Sebastian Diaz, Shigehiro Yamada, Michael Yozwiak, Omar L. Nelson, Gustavo C. Rodriguez, John Heusinkveld, Ie-Ming Shih, David S. Alberts, and Jennifer K. Barton "Sub-millimeter endoscope demonstrates feasibility of in vivo reflectance imaging, fluorescence imaging, and cell collection in the fallopian tubes," Journal of Biomedical Optics 26(7), 076001 (2 July 2021). https://doi.org/10.1117/1.JBO.26.7.076001
Received: 14 December 2020; Accepted: 15 June 2021; Published: 2 July 2021
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Endoscopes

Fallopian tube

In vivo imaging

Reflectivity

Luminescence

Tissues

Biopsy

Back to Top