3 May 2024 Assessing changes in regional cerebral hemodynamics in adults with a high-density full-head coverage time-resolved near-infrared spectroscopy device
Author Affiliations +
Abstract

Significance

Cerebral oximeters have the potential to detect abnormal cerebral blood oxygenation to allow for early intervention. However, current commercial systems have two major limitations: (1) spatial coverage of only the frontal region, assuming that surgery-related hemodynamic effects are global and (2) susceptibility to extracerebral signal contamination inherent to continuous-wave near-infrared spectroscopy (NIRS).

Aim

This work aimed to assess the feasibility of a high-density, time-resolved (tr) NIRS device (Kernel Flow) to monitor regional oxygenation changes across the cerebral cortex during surgery.

Approach

The Flow system was assessed using two protocols. First, digital carotid compression was applied to healthy volunteers to cause a rapid oxygenation decrease across the ipsilateral hemisphere without affecting the contralateral side. Next, the system was used on patients undergoing shoulder surgery to provide continuous monitoring of cerebral oxygenation. In both protocols, the improved depth sensitivity of trNIRS was investigated by applying moment analysis. A dynamic wavelet filtering approach was also developed to remove observed temperature-induced signal drifts.

Results

In the first protocol (28±5 years; five females, five males), hair significantly impacted regional sensitivity; however, the enhanced depth sensitivity of trNIRS was able to separate brain and scalp responses in the frontal region. Regional sensitivity was improved in the clinical study given the age-related reduction in hair density of the patients (65±15 years; 14 females, 13 males). In five patients who received phenylephrine to treat hypotension, different scalp and brain oxygenation responses were apparent, although no regional differences were observed.

Conclusions

The Kernel Flow has promise as an intraoperative neuromonitoring device. Although regional sensitivity was affected by hair color and density, enhanced depth sensitivity of trNIRS was able to resolve differences in scalp and brain oxygenation responses in both protocols.

1.

Introduction

Surgical procedures requiring general anesthesia can lead to impaired oxygen supply to the brain, increasing the risk of cerebral injury. Cerebral desaturation occurs in up to 25% to 37% of all cardiac patients and 69% to 76% of patients undergoing high-risk procedures, such as carotid endarterectomy.13 It is particularly a risk during shoulder surgery due to low blood pressure caused by performing the procedure in a semi-upright position (i.e., beach chair position).4 Notably, up to 80% of patients undergoing surgery in the beach chair position will require blood pressure interventions to mitigate the risk of cerebral desaturation.3,5 Interventions include re-positioning patients, administering vasopressors to increase mean arterial blood pressure (MAP),6 and transfusion in the case of anemia. Detecting evidence of cerebral desaturation is, therefore, a critical component of clinical management to enable prompt administration of treatments, thereby improving patient outcomes.6,7

Cerebral oximeters are now widely used intraoperatively to assess tissue oxygen saturation (StO2) and detect desaturation events.6,8 However, commercial cerebral oximeters typically have two channels for bilateral monitoring of the frontal watershed regions.9 This limited spatial coverage may result in undetected regional cerebral desaturation events.10 A further concern with such systems, which use continuous-wave (cw) near-infrared spectroscopy (NIRS) technology, is their susceptibility to extracerebral signal contamination and limited sensitivity to the adult brain. Advances in NIRS technology are providing promising alternatives that could overcome the abovementioned limitations. For example, Kernel Flow11,12 is a new NIRS device with full-head coverage, high density of optodes, and is based on time-resolved NIRS (trNIRS) technology,13 which provides greater sensitivity to the adult brain compared to cwNIRS.1416 However, the feasibility of using this device for neuromonitoring during surgery is unknown. Therefore, we investigated the possibility of using Kernel flow11,12 to monitor regional cerebral blood oxygenation during surgery.

The feasibility study was conducted using two hemodynamic protocols to (1) investigate the ability of the Flow system to simultaneously measure cerebral oxygenation across multiple cortical regions and (2) compare cerebral hemodynamic responses derived using depth-sensitive trNIRS parameters. The first protocol involved applying digital carotid compression (DCC) to healthy volunteers to assess the device’s ability to detect transient decreases in oxygenation. This simple technique induces a decrease in middle cerebral artery blood flow, similar to an occlusion of the internal carotid artery, and is commonly used to assess collateral blood flow during neurovascular surgeries.17 Based on previous studies monitoring StO2 during DCC,18,19 it was hypothesized that all probes on the ipsilateral hemisphere would exhibit a decrease in StO2.18

The second protocol assessed the feasibility of using the device for intraoperative neuromonitoring of patients undergoing elective shoulder surgery. Specifically, the effects of phenylephrine (PE) on cerebral oxygenation were investigated. PE is a vasopressor that has been shown to reduce StO2.20,21 Importantly, PE is known to cause systemic vasoconstriction without directly affecting cerebral vasomotor tone.22 Therefore, we hypothesize that the effects of PE would be similar across the entire cortex. Additionally, analysis conducted using various depth-sensitive trNIRS parameters was expected to reveal differences in scalp and brain responses to PE administration.

2.

Methods

2.1.

Instrumentation

Kernel Flow11,12 is a trNIRS device that provides high-density, full-head coverage. The system is equipped with 52 hexagonal modules (Fig. 1), each having 2 laser sources in the middle, emitting at 690 and 850 nm, and 6 detectors at the vertices located 10 mm from the sources. There are 52 source locations and 312 detectors. The system provides up to 2206 potential channels with varying source–detector distances (SDDs); 312 are short-distance channels (SDD = 10 mm), and 1894 are long-distance channels (SDD between 20 and 60 mm) across the head. For this study, intramodular channels were classified as short channels (SDD = 10 mm), and cross-modular channels with SDD25  mm were classified as long channels. The placement of the optical probes is divided into eight plates that roughly cover the brain’s four lobes and will be referred to as frontal, temporal, occipital, and sensorimotor [Fig. 1(d)]. Additionally, the system uses spring-loaded light pipes for transmission and detection to improve scalp coupling. The device weighs 2.2 kg and has a sampling rate of 7.1 Hz with a built-in temperature sensor and accelerometer.

Fig. 1

(a) Kernel Flow system showing the inside of the helmet. (b) One of the Kernel Flow modules with the light source in the middle and detectors around the perimeter. (c) Kernel Flow on a participant’s head. (d) Kernel Flow probe placement on a standard head model.23 Red dots depict laser sources, and blue dots are detectors. The black circles show the division of the eight brain regions (two hemispheres, each divided into frontal, sensorimotor, temporal, and occipital region). Images in panels (a) and (b) are adapted from Ref. 24.

JBO_29_S3_S33302_f001.png

2.2.

Experimental Protocols

2.2.1.

Protocol 1: digital carotid compression

All procedures were approved by the Health Sciences Research Ethics Board at Western University, which adheres to the guidelines of the Tri-Council Policy Statement for research involving humans. Participants provided written informed consent following written and verbal explanations of the experimental procedures. Ten healthy, young volunteers (Table 1) were recruited to participate in the study. Fitzpatrick Skin Type25 and hair color were recorded. The former is based on skin tone and how reactive the skin is to sunlight, ranging from I (light and easily sunburns) to VI (dark and does not typically sunburn). The exclusion criteria included pre-existing neurological, psychiatric, or vascular disease diagnoses. Participants were seated upright with the Kernel Flow on their head and instructed to remain as still as possible during the experiment. The experimental protocol started with a 30-s baseline period, followed by 30-s carotid compression, and ended with a 30-s recovery period. DCC involved applying light pressure to the right common carotid artery at a location 1  cm superior to the clavicle.18 DCC was repeated twice in each participant, and the trial showing the most pronounced hemodynamic change was used for data analysis.

Table 1

Healthy volunteer demographics.

Volunteer IDAgeSexFitzpatrick skin typeHair colorClassificationOverall active channels (%)
130MIVBaldingLight39
229FIIIDark brownDark6
323FIVBlackDark5
438MIIBlondeLight44
522FIIIDark brownDark8
631FIIDark brownDark6
728MIIILight brownLight16
825MVIBlackDark9
927FIIILight brownLight10
1022MIVBlackDark9

2.2.2.

Protocol 2: Phenylephrine administration during shoulder surgery

Adult patients undergoing elective shoulder surgery at St. Joseph’s Hospital (London, Ontario) were prospectively recruited between October 2022 and March 2023. All procedures were approved by the Health Sciences Research Ethics Board at Western University (REB# 115021), which adheres to the guidelines of the Tri-Council Policy Statement for research involving humans. Participants (Table 2) provided written informed consent following written and verbal explanations of the protocol. Once the patient was anesthetized, intubated, and placed in the beach chair position, the Kernel Flow was placed on the head and stabilized with a head strap (Fig. 2). To avoid overheating, a forced-air cooling system (Bair Hugger, 3M Medical) was used to blow cool air at the Flow device. Data were collected continuously throughout the surgical procedure, and instances of PE administration were noted. MAP and heart rate (HR) measurements were acquired using hospital equipment every 5 min. To investigate the effect of PE, 1 min prior to the PE administration was used as the baseline, and 5 min post-injection were included in the analysis.

Table 2

Demographics of the patients included in the analysis.

Patient IDSurgery sideSexAgeWeight (kg)Height (cm)Shoulder surgery typeNumber of PE bolus injections
5RF7569160Arthroplasty reverse6
7LM6992173Arthroplasty reverse4
12RM6692.5175Shoulder arthroplasty total2
13RM68115180Arthroplasty reverse3
23RF6067.5147Arthroplasty reverse1

Fig. 2

A shoulder surgery patient with the Kernel flow system on the head. The device was stabilized with a head strap.

JBO_29_S3_S33302_f002.png

2.3.

Data Processing

Raw data were uploaded to Kernel’s cloud-based pre-processing pipeline, which starts by removing poor-contact channels based on the shape of the distribution of times-of-flight (DTOF), photon counts, scalp coupling index, and HR peak power.12 After preprocessing, we received the pruned data containing only “good” quality channels. The pipeline then detects the peak of each DTOF and defines an integration window for computing the statistical moments: 10% of the peak on the rising edge and 1% on the trailing edge of the histogram. Thereafter, three statistical moments are calculated: (1) the total number of photons (N, i.e., the zeroth moment); (2) the mean time of flight (t, i.e., the first moment); and (3) the variance (V, i.e., the second moment). Note that higher moments are more sensitive to the brain, whereas lower moments are more sensitive to the scalp.15,26 For the remainder of the paper, the total number of photons (N) at a short SDD will be referred to as Nshort, and the total number of photons at a long SDD labeled Nlong. The same nomenclature will be applied to the mean time of flight (t) and the variance (V). The least depth-sensitive measurement is Nshort, which predominantly reflects the scalp response. Likewise, measurements with the greatest depth sensitivity are derived from Vlong, which mostly reflects the brain response.16,27

Data were further processed using an in-house pipeline developed in MATLAB (Mathworks Inc., United States). Changes in each moment relative to its baseline value were calculated to generate three different time courses for each wavelength. Time courses were then converted into corresponding absorption changes (Δμa(λ)) using sensitivity analysis28,29 [Eq. (1)], where ΔS is the signal change for a given moment and SFS represents the corresponding sensitivity factor (which also dependents on the source-detector distance)

Eq. (1)

Δμa(λ)=ΔSSFS.

Δμa was then converted to oxyhemoglobin (HbO) and deoxyhemoglobin (Hb) concentration changes (Δ[HbO] for HbO, and Δ[Hb] for Hb) using Eq. (2), where εHbO(λ) and εHbO(λ) are the molar extinction coefficients for HbO and Hb, respectively

Eq. (2)

Δμa(λ)=εHbO(λ)×Δ[HbO]+εHb(λ)×Δ[Hb].

Baseline values of [HbO] and [Hb] (i.e., [HbO]b and [Hb]b) from the literature were assumed for all participants (33 and 24  μM,30,31 respectively). Relative concentration changes, calculated from the different moments and SDDs, were used to estimate StO2 changes using the following equation:

Eq. (3)

ΔStO2=((Δ[HbO]+[HbO]b)(Δ[HbO]+[HbO]b)+(ΔHb+[Hb]b)[HbO]b[HbO]b+[Hb]b)×100%.

Detection channels were sorted into eight brain regions based on the four brain lobes of each hemisphere, and regional averages were calculated. Time courses were filtered with a zero-phase filter using a window size of approximately 1.5 s.

2.4.

Correction for Temporal Drift

The time course of Δ[HbO] and Δ[Hb] exhibited a strong dependence on the device’s temperature, which manifested as temperature-dependent signal drifts. This was particularly evident during the longer monitoring periods in the surgery study (protocol 2). To mitigate the confounding effects of the temperature drift, dynamic bandpass wavelet filtering was implemented. The filtering consisted of applying a continuous wavelet transform to decompose the temperature time course of each module to identify the frequency of temperature drift with maximum power across the monitoring period. This frequency was then used as the lower limit of the bandpass filter for the Δ[HbO] and Δ[Hb] time courses. Across all patients, the lowest frequency of the bandpass filter was 0.008 Hz (period of 125 s). The upper limit of the bandpass filter was consistently 3.1 Hz, as limited by the device sampling rate (7.1 Hz). Each channel was filtered with its module-specific, time-varying bandpass filter. This resulted in 52 unique filters used across the head. This approach could not be applied to the DCC time courses (protocol 1) because the measurements were only 90 s; instead, a simple first-order polynomial was used for detrending.

2.5.

Statistical Analysis

All data were assessed for normality through a visual inspection of a QQ plot as well as the Shapiro–Wilk test. All statistics were conducted in GraphPad Prism version 9 (GraphPad Software Inc., San Diego), and statistical significance was defined as p0.05.

2.5.1.

Protocol 1: digital carotid compression

A one-way analysis of variance (ANOVA) was used to compare regional differences in device sensitivity (four levels: frontal, sensorimotor, temporal, and occipital in the ipsilateral hemisphere) as measured by the percentage of good-quality, long-distance channels compared to the total number of expected channels (% of active channels). The impact of hair was also assessed by classifying participants as having “light” hair if they were either bald or had blonde or light brown hair and “dark” hair if they had brown or black hair. An ordinary two-way ANOVA was used to evaluate “light” hair versus “dark” hair (two levels) and differences in regional device sensitivity (four levels: frontal, sensorimotor, temporal, and occipital in the ipsilateral hemisphere). Additionally, a two-way repeated-measures ANOVA was used to compare the different depth-sensitive measurements (three levels: Nshort, Nlong, and Vlong) across time bins (four levels: baseline, 30 to 45 s, 45 to 60 s, and 60 to 75 s). For each time bin and moment, the 5-s average around the maximum change in Δ[HbO], Δ[Hb], ΔStO2, in the ipsilateral hemisphere was analyzed. The Nshort was selected to reflect scalp response, Nlong provides a signal similar to that of cwNIRS, and Vlong is the most sensitive to the brain.

2.5.2.

Protocol 2: Phenylephrine administration during shoulder surgery

A one-way ANOVA was used to compare regional differences in device sensitivity (four levels: frontal, sensorimotor, temporal, and occipital) as measured by the percentage of active channels. A Friedman test was used for comparisons if the data were not normally distributed. A paired t-test was run to determine differences in physiological parameters (MAP and HR) before and after PE administration. Additionally, a two-way repeated-measures ANOVA was used to compare the different depth-sensitive measurements (three levels: Nshort, Nlong, Vlong) across time bins (three levels: baseline, 1 to 3.5 min, 3.5 to 6 min). For each time bin and moment, the 5-s average around the maximum change in Δ[HbO], Δ[Hb], and ΔStO2 was analyzed. A one-way ANOVA was also used to assess the regional differences in Δ[HbO] and Δ[Hb], and the time of the minimum desaturation.

3.

Results

3.1.

Protocol 1: Digital Carotid Compression

Ten subjects were included in the study (28±5 years; five females and five males; Table 1). For the frontal region, 36%±23% of the long-distance channels had good-quality signal, among which 52%±16% were between 25 and 35 mm, 39%±8% between 35 and 45 mm, and 8%±12% between 45 and 60 mm. However, the number or good-quality channels decreased significantly (p0.004) for the other regions: 8%±12% for the sensorimotor cortex, 11%±17% for the temporal lobe, and 5%±10% for the occipital lobe [Fig. 3(a)]. Additionally, no significant interaction was found between hair type and region (p=0.3). However, significant main effects of hair type (p0.0001) and region (p0.0001) were found. Subjects with “light” hair (n=4; three males, one female) had a significantly higher percentage of good-quality channels [27%±25% versus 7%±10%; p0.001; Fig. 3(b)] when compared to those with “dark” hair (n=6; two males, four females).

Fig. 3

(a) Regional sensitivity of the device (n=10; mean age: 28±5; five females). Box-and-whisker plots show first and third quartiles as outlined by the box, with the middle line indicating the median and the ends of the line indicating minimum and maximum values. Crosses indicate outliers. Asterisks indicate that the frontal region was significantly different from all other regions. (b) “Light” (n=4) and “dark” (n=6) hair participants’ regional sensitivity are displayed as stacked bars of the four regions. An example of the active channel maps overlaid on a standard head model23 from a participant from the “light” hair group is shown in panel (c), and an example of the “dark” hair group is shown in panel (d).

JBO_29_S3_S33302_f003.png

Given that the frontal was the only region that had good-quality long-distance channels for all the participants [see Figs. 3(c) and 3(d)], it is the only region that could reliably be used to assess differences in the response to DCC across all participants. Figure 4(a) shows Δ[HbO] and Δ[Hb] responses to DCC obtained from the zero-order moment measured at the shortest SDD, Nshort (i.e., the measurement that is overwhelmingly sensitive to the extra-cerebral tissues). Figures 4(b)4(d) display the hemodynamic responses for all three moments extracted from data measured at SDD25  mm. The panels are ordered from the least to the most depth-sensitive measurements [moving from Figs. 4(b)4(d)]. The contralateral time courses are in Fig. S1 in the Supplementary Material but show negligible changes during DCC. Further analysis of the ipsilateral Δ[HbO], Δ[Hb], and ΔStO2 profiles across time revealed a significant interaction in all three parameters across the moments (p0.0001 for Δ[HbO], Δ[Hb], and StO2)—see Fig. S2 in the Supplementary Material for the complete analysis of significant differences. The most notable differences were the faster recovery during desaturation in the second half of the DCC and the hyperemic response after DCC for Vlong.

Fig. 4

(a)–(d) Average Δ[HbO] (red) and Δ[Hb] (blue) responses to DCC (indicated by the gray-shaded region) in the ipsilateral frontal region. Time courses are presented for the signals measured at short (a) and long (b)–(d) distances. Responses derived from (a) and (b) the total number of photons (N), (c) mean time of flight (t), and (d) variance (V) are shown. (e)–(g) Average long distance (black) and short distance (green) ΔStO2 derived from (e) N, (f) t, and (g) V are shown. All time courses were averaged across 10 subjects and shading surrounding each line represents the standard deviation.

JBO_29_S3_S33302_f004.png

3.2.

Protocol 2: Phenylephrine Administration During Shoulder Surgery

Data were acquired from 27 shoulder surgery patients. All patients were classified as light-skinned (I or II on the Fitzpatrick skin type scale) and had thinner lighter-colored hair due to age and demographics. Wearing the Kernel flow helmet resulted in a minor skin abrasion for one patient, but no complications were reported for the other subjects. The average age was 65±15 years (14 females and 13 males). Seventeen patients had right shoulder surgery, and ten underwent left shoulder surgery. Only 10 patients received at least one bolus injection of PE, and of these, 5 were excluded due to technical challenges, such as motion artifacts or corrupt files. For consistency, only the first PE injection from each patient was analyzed and compared. Table 2 provides the demographics of the five patients (68±5; two females, two males) included in the analysis and the number of PE injections they received.

To assess the device’s regional sensitivity, all 27 patients were considered. Four patients were excluded from this analysis due to technical issues where some modules could not be connected. For protocol 2, the signal was generally high as the subjects were older and had lower hair density. Overall, the frontal region had 67%±37% of the long-distance channels classified as good quality, the sensorimotor had 71%±32%, the temporal had 59%±35%, and the occipital had 62%±32%. No significant differences were found (p=0.5) in the percentage of good-quality channels between brain regions in this patient population. Resultantly, limiting long distance channels to those with SDD of 25 to 35 mm was suitable for the rest of the analyses.

An example of temperature-drift correction for each of the three moments is shown in Figs. 5(a)5(c). In this case, initial Nlong and Vlong time courses followed the same trend as temperature, whereas tlong showed the opposite trend. Figures 5(d) and 5(e) present the average Δ[HbO] and Δ[Hb] responses to phenylephrine injection before and after temperature-drift correction. The results demonstrated that temperature correction reduced the variability across subjects, as indicated by the standard deviation, and, more importantly, resulted in the Δ[HbO] and Δ[Hb] time courses returning to baseline.

Fig. 5

Example time courses of (a) Nlong, (b) tlong, and (c) Vlong from one module over the frontal region of a patient. The overlaid colored boxes indicate windows with different wavelet filter parameters that varied with the corresponding temperature time course (dashed black). For each moment, both the raw and filtered time courses are shown for [HbO] (magenta and red, respectively) and [Hb] (cyan and blue, respectively). Average [HbO] and [Hb] time courses in response to a phenylephrine injection are shown (d) before (e) and after temperature correction. These time courses were calculated from Vlong. The baseline is displayed in gray, and the response to the PE injection is in white.

JBO_29_S3_S33302_f005.png

Figure 6 shows the average time courses for the frontal region for patients included in the PE analysis (n=5). The PE injection occurred at the 1-min time point. Through qualitative observation, Fig. 6(a), predominantly the scalp response, showed large variability in the data; however, there appears to be a sharp decrease in Δ[HbO] and Δ[Hb] following PE injection. Qualitatively, in the more brain-sensitive measurements [Vlong, Fig. 6(d)], there is a pronounced desaturation with a simultaneous decrease in Δ[HbO] and Δ[Hb] increase. Further, the analysis of the time courses of Δ[HbO], Δ[Hb], and ΔStO2 (see Fig. S3 in the Supplementary Material) revealed a significant main effect of time in Δ[HbO] only (p0.04) as the baseline was significantly different from the injection period.

Fig. 6

(a)–(d) Average Δ[HbO] (red) and Δ[Hb] (blue) responses to the first phenylephrine bolus injection (indicated by the white region; the baseline is shown in gray) in the left frontal region. Time courses are presented for the signals measured at short (a) and long (b)–(d) Responses derived from (a) and (b) the total number of photons (N), (c) mean time of flight (t), and (d) variance (V) are shown. Average long distance (black) and short distance (green) ΔStO2 derived from (e) N, (f) t, and (g) V are shown. All time courses were averaged across five subjects and shading surrounding each line represents the standard deviation.

JBO_29_S3_S33302_f006.png

Moreover, MAP increased post-injection (61±10  mmHg to 80±12  mmHg; p0.01) versus pre-injection, but HR did not significantly change after PE injection (58±10  bpm to 54±11  bpm; p=0.08).

Figure 7 displays the average time courses (n=5) in all brain regions, calculated using Vlong. The curves show that all regions had a similar response to the PE injection, characterized by a simultaneous decrease in Δ[HbO] and increase in Δ[Hb]. There were no significant differences in the regional responses to PE injection for the magnitude of Δ[HbO] (p=0.7) or Δ[Hb] (p=0.7) or time of maximum change (p=0.99).

Fig. 7

Concentration changes (ΔC) of HbO (red), and Hb (blue), calculated using Vlong, showing the average (n=5) first response to phenylephrine bolus injection during surgery across the four brain regions (frontal, sensorimotor, temporal, and occipital). The shading indicates the standard deviation across participants.

JBO_29_S3_S33302_f007.png

4.

Discussion

This work investigated the feasibility of using a novel high-density full-head coverage time-resolved system for regional cerebral oxygenation monitoring. Two protocols were conducted: (1) DCCs in healthy volunteers and (2) PE administration during shoulder surgery. The two objectives were to investigate the system’s ability to measure oxygenation simultaneously across all brain regions and compare cerebral hemodynamic responses computed using time-resolved parameters with different depth sensitivities.

4.1.

Protocol 1: Digital Carotid Compression

In the DCC protocol, the device’s sensitivity was assessed in young healthy subjects by calculating the percentage of good-quality channels across the head. The analysis revealed that the frontal region had a significantly higher number of good-quality channels compared to the sensorimotor, temporal, and occipital regions. Additionally, the device had better sensitivity in subjects with light hair than those with dark hair. These findings indicate that signal quality depends on the type and density of hair. Consequently, using the device for full-head neuromonitoring in patients with darker and longer hair would be challenging. Further, the regional distribution of channel sensitivity reported in this work showed that the frontal region had the most signal while the occipital region had the lowest percentage of good-quality channels. This is consistent with the findings of Ortega-Martinez et al.,32 who reported a similar sensitivity distribution with the Kernel Flow. Due to the lower overall signals in the younger participants recruited for protocol 1, a large range of SDD values was used in the analysis. The similarity in the time courses for SDDs 25 to 35 mm and those 35 to 45 mm (see Fig. S4 in the Supplementary Material) indicates that this approach was reasonable. Ideally, improving the signal-to-noise ratio would enable a narrower range of SDD to be selected.

Since the frontal region (measured through the forehead) consistently had the best signal in all participants, this region was used to assess the device’s depth sensitivity in the DCC protocol. Additionally, only ipsilateral hemisphere time courses were analyzed, as the contralateral measurements showed negligible changes during DCC, as expected. The hemodynamic responses to the DCC challenge computed using the various depth-sensitive measurements differed significantly. The measurement known to have the most sensitivity to the brain (Vlong) showed a faster recovery from desaturation in the second half of the DCC, which may be attributed to collateral flow in the brain that is not possible in the scalp. The larger decrease in saturation observed in Nlong compared to Nshort agrees with our previous work18 using a cwNIRS system, which found a larger reduction in saturation in the long SDDs compared to the short SDDs in response to DCC. The long-distance cwNIRS measurements were previously attributed to the brain; however, the smaller decrease in saturation in Vlong observed in this study suggests that cerebral desaturation following PE administration is likely smaller than previously suggested.18 Additionally, Δ[Hb] increase was only visible in the long-distance measurements, which indicates greater oxygen extraction in the brain during DCC compared to the scalp, reflecting the lack of oxygen stores and higher energy metabolism in the brain.

Finally, a cerebral (Vlong) hyperemic response was observed during the recovery in the Δ[HbO] and ΔStO2 as blood flow increased following occlusion release. Similar hyperemia has been reported in our previous work using diffuse correlation spectroscopy18 but was not evident in the StO2 values using cwNIRS. This may be due to the additional depth-sensitivity of Vlong, as the hyperemia is also not evident in the Nlong in the current study. Diffuse correlation spectroscopy provides inherently better depth-sensitivity than cwNIRS due to the substantially higher blood flow in the brain compared to the scalp,33,34 which can explain this difference. Additionally, these findings agree with Sorteberg et al. who performed DCC on patients with neck neoplasms or traumatic carotid-cavernous fistulas,17 and found that the blood velocity in the middle cerebral artery (measured using transcranial Doppler ultrasound) showed a hyperemic response following DCC.

4.2.

Protocol 2: Phenylephrine Administration During Shoulder Surgery

There were no significant differences in device sensitivity between brain regions in the surgery data. This finding contrasts with the first protocol in which the frontal region had the best signal. This difference is likely due to the difference in demographics between the two studies. For the surgery study, subjects were older, had lighter-colored hair, and reduced hair density. Therefore, these results agree with the findings of the first protocol that those with “light” hair had better overall signal. The better sensitivity of the device in surgery patients is promising since the aim was to assess the device intraoperatively.35

An unexpected challenge with neuromonitoring over longer periods with the Kernel flow was overheating. An air-cooling device was used to blow cool air on the device; however, there were still temperature-induced signal drifts (see Fig. 5). A dynamic wavelet bandpass filtering approach to remove very low frequencies (below 0.008 Hz or 125 s) was implemented to correct for these signal drifts and very high frequencies (above 3.1 Hz) for motion artifacts. This wavelet filtering proved crucial, as evident by comparing Nlong, tlong, and Vlong time series before and after applying temperature correction (Fig. 5). It is possible that wavelet filtering may have removed physiological signals outside the bandpass filter. However, this study focused on the rapid cerebral hemodynamic response to a vasopressor, which occurred over a much shorter time frame than those caused by temperature instability.

After 5 min of PE injection, the MAP settles into a new elevated baseline; therefore, it is possible that the hemodynamic parameters settle into a new baseline as well. In this study, however, the filtered Δ[HbO] and Δ[Hb] returned to the pre-PE values, and not a new baseline [see Fig. 5(e)]. This finding agrees with the work of Meng et al.36 and Fassaert et al.37 who likewise demonstrated that while cerebral oxygenation returned to baseline within 5 min, the recovery of basal MAP typically took longer, likely reflecting differences in cerebral and systemic vascular regulation.

The injection of PE caused the expected systemic response, i.e., an increase in MAP consistent with other studies.38,39 The slight decrease in HR (although not significant) was also expected and has been reported in the literature.39 In terms of the NIRS measurements, Nshort, which is assumed to predominantly reflect scalp hemodynamics, showed a sharp decrease in Δ[HbO], potentially driven by the vasoconstrictive effects of PE on alpha receptors present in scalp tissue. However, a larger sample size would help investigate this further as the time courses had large variability. Desaturation events, characterized by a simultaneous increase in Δ[Hb] and decrease in Δ[HbO], were observed in the Vlong, the most brain-sensitive measurement. The cerebral desaturation agrees with previous studies using cwNIRS devices, which report decreases as large as 14% in cerebral oxygen saturation40; however, the desaturation reported in the current work are smaller (less than 2%) and likely not clinically significant. Previous studies have shown that the magnitude of StO2 changes in response to PE decreases with increased brain sensitivity.20,40,41 Additionally, Rosenthal et al.42 detected very small changes in cerebral blood flow in response to phenylephrine administration, which is consistent with the small changes in the time-resolved measurements in this work. Our findings also agree with other studies using more depth-sensitive NIRS techniques, such as frequency-domain NIRS (fdNIRS).21 It has been suggested that the previously reported cerebral desaturation following PE injection may be related to signal contamination due to extracranial vasoconstriction.43 Importantly, recent work utilizing direct measurements of cerebral tissue oxygen tension in open rat brains44 suggests that PE injection does not result in cerebral desaturation. Therefore, despite the well-known superior depth sensitivity of trNIRS,15 the small magnitude of desaturation observed in Vlong may be due to residual scalp contamination. Moreover, the small decreases in Δ[HbO] and increases in Δ[Hb] may be due to changes in cerebral artery vasomotor tone. That is, a small drop in arterial blood volume due to vasoconstriction could shift the measurements to the venous side, leading to the observed trends.

There were no statistically significant differences in the regional response to PE. It is important to note that although there were no significant differences, this may be due to the smaller sample size (only five patients received PE). A subsequent larger study could provide more conclusive answers to potential regional differences in response to PE administration.

4.3.

Limitations

There were a few limitations to this work. First, access to the data in the rawest form (before the automatic processing through the Kernel Flow pipeline) would be helpful to assess why channels were discarded within the automated pipeline and allow us to evaluate the stability and quality of the recorded DTOFs. Additionally, it remains unclear if potential regional differences in hemodynamic responses result from actual physiological differences. They could result from sensitivity differences across the head due to the data processing that uses sensitivity factors to estimate the hemodynamic response. Using magnetic resonance imaging, Wu et al. found that the skull thickness varies across the head;45 however, this was not accounted for in the sensitivity factors and could impact our conclusion on the lack of regional differences.

4.4.

Future Work

Overcoming the current hair limitation would help detect potential regional differences in the cerebral response to DCC when comparing regions supplied by different cerebral vasculatures. Due to the limited variability in skin tone in the current study, the analysis focused on hair and not skin tone. A larger and more diverse sample group would allow an analysis of the impact of both hair and skin types on device sensitivity. It would also be useful to conduct studies using tissue-mimicking phantoms to better understand the temporal stability of the Kernel Flow. A further consideration is that most surgeries are performed with the patient supine, which would require testing to ensure the device provides sufficient contact across the head when in this position.

5.

Conclusions

In summary, this study demonstrated the feasibility and potential of using a high-density, full-head coverage trNIRS device for adult neuromonitoring using two study protocols: DCC and PE administration during shoulder surgery. The device’s sensitivity depended on the presence and type of hair, particularly in the first protocol, which involved young volunteers. The second protocol showed improved signal when using the device on shoulder surgery patients who were older and had decreased hair density. Notably, the age of candidates for neurovascular surgery is typically higher than those recruited for the first protocol, which may alleviate this limitation. Analysis of the device’s depth-sensitivity showed distinct scalp and brain responses, which shows promise. The use of the device for intraoperative neuromonitoring demonstrated the feasibility of its clinical translation. A larger-scale study would shed more light on potential regional differences due to DCC and PE administration.

Disclosures

The Flow 50 device used in the current work was provided to our research group by Kernel for a free loan period, after which it was returned to the company. The company had no role in the study’s design, in the collection, analysis, or interpretation of data, in the writing of the manuscript, or in the decision to publish the results.

Code and Data Availability

Data can be made available by contacting the authors. Because of the participant consent obtained as part of the recruitment process, making these data publicly available is not possible.

Acknowledgments

We would like to kindly thank the Kernel Flow team for their technical support during this work. This work was supported by the Natural Sciences and Engineering Research Council of Canada (NSERC; Discovery Grant No. RGPIN-2023-05561) and the Canadian Institutes of Health Research (CIHR; Grant No. 180306). The authors would also like to acknowledge Ms. Lee-Anne Fochesato, Research Coordinator of the Department of Anesthesia and Perioperative Medicine, who facilitated the ethics board’s approval and conduct of this study.

References

1. 

A. Deschamps et al., “Reversal of decreases in cerebral saturation in high-risk cardiac surgery,” J. Cardiothorac. Vasc. Anesth., 27 (6), 1260 –1266 https://doi.org/10.1053/j.jvca.2013.01.019 JCVAEK 1053-0770 (2013). Google Scholar

2. 

J. Schoen et al., “Cognitive function after sevoflurane- vs propofol-based anaesthesia for on-pump cardiac surgery: a randomized controlled trial,” Br. J. Anaesth., 106 (6), 840 –850 https://doi.org/10.1093/bja/aer091 BJANAD 0007-0912 (2011). Google Scholar

3. 

H. B. Nielsen, “Systematic review of near-infrared spectroscopy determined cerebral oxygenation during non-cardiac surgery,” Front. Physiol., 5 93 https://doi.org/10.3389/fphys.2014.00093 FROPBK 0301-536X (2014). Google Scholar

4. 

G. S. Murphy, S. B. Greenberg and J. W. Szokol, “Safety of beach chair position shoulder surgery: a review of the current literature,” Anesth. Analg., 129 (1), 101 –118 https://doi.org/10.1213/ANE.0000000000004133 (2019). Google Scholar

5. 

G. S. Murphy et al., “Cerebral oxygen desaturation events assessed by near-infrared spectroscopy during shoulder arthroscopy in the beach chair and lateral decubitus positions,” Anesth. Analg., 111 (2), 496 –505 https://doi.org/10.1213/ANE.0b013e3181e33bd9 (2010). Google Scholar

6. 

J. M. Murkin et al., “Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study,” Anesth. Analg., 104 (1), 51 –58 https://doi.org/10.1213/01.ane.0000246814.29362.f4 (2007). Google Scholar

7. 

S. Goldman et al., “Optimizing intraoperative cerebral oxygen delivery using noninvasive cerebral oximetry decreases the incidence of stroke for cardiac surgical patients,” Heart Surg. Forum, 7 (5), E376 –E381 https://doi.org/10.1532/HSF98.20041062 (2004). Google Scholar

8. 

C. Lewis et al., “Cerebral neuromonitoring during cardiac surgery: a critical appraisal with an emphasis on near-infrared spectroscopy,” J. Cardiothorac. Vasc. Anesth., 32 (5), 2313 –2322 https://doi.org/10.1053/j.jvca.2018.03.032 JCVAEK 1053-0770 (2018). Google Scholar

9. 

B. Milne et al., “Neuromonitoring and neurocognitive outcomes in cardiac surgery: a narrative review,” J. Cardiothorac. Vasc. Anesth., 36 (7), 2098 –2113 https://doi.org/10.1053/j.jvca.2021.07.029 JCVAEK 1053-0770 (2022). Google Scholar

10. 

C. Rummel et al., “Monitoring cerebral oxygenation during balloon occlusion with multichannel NIRS,” J. Cereb. Blood Flow Metab., 34 (2), 347 –356 https://doi.org/10.1038/jcbfm.2013.207 (2014). Google Scholar

11. 

F. Kamar et al., “Sensitivity of a full-head coverage high-density time-resolved NIRS device to carotid compressions,” Proc. SPIE, 12376 123760H https://doi.org/10.1117/12.2646633 PSISDG 0277-786X (2023). Google Scholar

12. 

H. Y. Ban et al., “Kernel Flow: a high channel count scalable time-domain functional near-infrared spectroscopy system,” J. Biomed. Opt., 27 (7), 074710 https://doi.org/10.1117/1.JBO.27.7.074710 JBOPFO 1083-3668 (2022). Google Scholar

13. 

W. Becker, Advanced Time-Correlated Single Photon Counting Applications, 1 –63 Cham (2015). Google Scholar

14. 

A. Liebert et al., “Evaluation of optical properties of highly scattering media by moments of distributions of times of flight of photons,” Appl. Opt., 42 (28), 5785 https://doi.org/10.1364/AO.42.005785 APOPAI 0003-6935 (2003). Google Scholar

15. 

D. Milej et al., “Direct assessment of extracerebral signal contamination on optical measurements of cerebral blood flow, oxygenation, and metabolism,” Neurophotonics, 7 (4), 045002 https://doi.org/10.1117/1.NPh.7.4.045002 (2020). Google Scholar

16. 

D. Milej et al., “Incorporating early and late-arriving photons to improve the reconstruction of cerebral hemodynamic responses acquired by time-resolved near-infrared spectroscopy,” J. Biomed. Opt., 26 (5), 056003 https://doi.org/10.1117/1.JBO.26.5.056003 JBOPFO 1083-3668 (2021). Google Scholar

17. 

A. Sorteberg et al., “Varying impact of common carotid artery digital compression and internal carotid artery balloon test occlusion on cerebral hemodynamics,” Head Neck, 20 (8), 687 –694 https://doi.org/10.1002/(SICI)1097-0347(199812)20:8<687::AID-HED5>3.0.CO;2-1 (1998). Google Scholar

18. 

M. Suwalski et al., “Assessing the sensitivity of multi-distance hyperspectral NIRS to changes in the oxidation state of cytochrome C oxidase in the brain,” Metabolites, 12 (9), 817 https://doi.org/10.3390/metabo12090817 (2022). Google Scholar

19. 

M. Ferrari et al., “Effects of carotid compression, as assessed by near infrared spectroscopy, upon cerebral blood volume and haemoglobin oxygen saturation,” J. R. Soc. Med., 80 (2), 83 –87 https://doi.org/10.1177/014107688708000208 JRSMD9 0141-0768 (1987). Google Scholar

20. 

P. Brassard et al., “Phenylephrine decreases frontal lobe oxygenation at rest but not during moderately intense exercise,” J. Appl. Physiol., 108 (6), 1472 –1478 https://doi.org/10.1152/japplphysiol.01206.2009 (2010). Google Scholar

21. 

L. Meng et al., “Impact of phenylephrine administration on cerebral tissue oxygen saturation and blood volume is modulated by carbon dioxide in anaesthetized patients,” Br. J. Anaesth., 108 (5), 815 –822 https://doi.org/10.1093/bja/aes023 BJANAD 0007-0912 (2012). Google Scholar

22. 

L. Meng et al., “Effects of phenylephrine on systemic and cerebral circulations in humans: a systematic review with mechanistic explanations,” Anaesthesia, 79 71 –85 https://doi.org/10.1111/anae.16172 (2023). Google Scholar

23. 

R. Oostenveld et al., “FieldTrip: open source software for advanced analysis of MEG, EEG, and invasive electrophysiological data,” Comput. Intell. Neurosci., 2011 1 –9 https://doi.org/10.1155/2011/156869 (2011). Google Scholar

24. 

“Kernel flow home,” https://www.kernel.com (). Google Scholar

25. 

W. H. Ward et al., “Clinical presentation and staging of melanoma,” Cutaneous Melanoma: Etiology and Therapy, 79 –89 Codon Publications, pp.  (2017). Google Scholar

26. 

D. Milej et al., “Optimization of the method for assessment of brain perfusion in humans using contrast-enhanced reflectometry: multidistance time-resolved measurements,” J. Biomed. Opt., 20 (10), 106013 https://doi.org/10.1117/1.JBO.20.10.106013 JBOPFO 1083-3668 (2015). Google Scholar

27. 

A. Abdalmalak et al., “Can time-resolved NIRS provide the sensitivity to detect brain activity during motor imagery consistently?,” Biomed. Opt. Express, 8 (4), 2162 https://doi.org/10.1364/BOE.8.002162 BOEICL 2156-7085 (2017). Google Scholar

28. 

A. Abdalmalak et al., “Assessing time-resolved fNIRS for brain-computer interface applications of mental communication,” Front. Neurosci., 14 105 https://doi.org/10.3389/fnins.2020.00105 1662-453X (2020). Google Scholar

29. 

A. Abdalmalak et al., “Single-session communication with a locked-in patient by functional near-infrared spectroscopy,” Neurophotonics, 4 (4), 040501 https://doi.org/10.1117/1.NPh.4.4.040501 (2017). Google Scholar

30. 

H. Auger et al., “Quantification of extra-cerebral and cerebral hemoglobin concentrations during physical exercise using time-domain near infrared spectroscopy,” Biomed. Opt. Express, 7 (10), 3826 –3842 https://doi.org/10.1364/BOE.7.003826 BOEICL 2156-7085 (2016). Google Scholar

31. 

G. Giacalone et al., “Cerebral time domain-NIRS: reproducibility analysis, optical properties, hemoglobin species and tissue oxygen saturation in a cohort of adult subjects,” Biomed. Opt. Express, 8 (11), 4987 –5000 https://doi.org/10.1364/BOE.8.004987 BOEICL 2156-7085 (2017). Google Scholar

32. 

A. Ortega-Martinez et al., “How much do time-domain functional near-infrared spectroscopy (fNIRS) moments improve estimation of brain activity over traditional fNIRS?,” Neurophotonics, 10 (01), 013504 https://doi.org/10.1117/1.NPh.10.1.013504 (2022). Google Scholar

33. 

L. N. Shoemaker et al., “Using depth-enhanced diffuse correlation spectroscopy and near-infrared spectroscopy to isolate cerebral hemodynamics during transient hypotension,” Neurophotonics, 10 (2), 025013 https://doi.org/10.1117/1.NPh.10.2.025013 (2023). Google Scholar

34. 

J. Selb et al., “Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia,” Neurophotonics, 1 (1), 015005 https://doi.org/10.1117/1.NPh.1.1.015005 (2014). Google Scholar

35. 

A. J. Fowler et al., “Age of patients undergoing surgery,” British Journal of Surgery, 106 (8), 1012 –1018 https://doi.org/10.1002/bjs.11148 BJSUAM 0007-1323 (2019). Google Scholar

36. 

L. Meng et al., “Effect of phenylephrine and ephedrine bolus treatment on cerebral oxygenation in anaesthetized patients,” Br. J. Anaesth., 107 (2), 209 –217 https://doi.org/10.1093/bja/aer150 BJANAD 0007-0912 (2011). Google Scholar

37. 

L. Fassaert, G. Jan de Borst and R. Immink, “Effect of phenylephrine and ephedrine on cerebral (tissue) oxygen saturation during carotid endarterectomy (PEPPER): a randomized controlled trail,” Eur. J. Vasc. Endovasc. Surg., 58 (6), E92 https://doi.org/10.1016/j.ejvs.2019.06.625 (2019). Google Scholar

38. 

A. F. Kalmar et al., “Phenylephrine increases cardiac output by raising cardiac preload in patients with anesthesia induced hypotension,” J. Clin. Monit. Comput., 32 (6), 969 –976 https://doi.org/10.1007/s10877-018-0126-3 (2018). Google Scholar

39. 

R. Rajanathan et al., “Phenylephrine-induced cardiovascular changes in the anesthetized mouse: an integrated assessment of in vivo hemodynamics under conditions of controlled heart rate,” Front. Physiol., 13 831724 https://doi.org/10.3389/fphys.2022.831724 FROPBK 0301-536X (2022). Google Scholar

40. 

V. T. Foss et al., “Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial,” Front. Physiol., 5 81 https://doi.org/10.3389/fphys.2014.00081 FROPBK 0301-536X (2014). Google Scholar

41. 

A. T. Moerman et al., “Incongruous effect of phenylephrine on changes in cerebral blood volume measured by near-infrared spectroscopy (NIRS) indicating extracranial contamination,” J. Clin. Monit. Comput., 36 (3), 745 –750 https://doi.org/10.1007/s10877-021-00702-3 (2022). Google Scholar

42. 

G. Rosenthal et al., “Brain tissue oxygen tension is more indicative of oxygen diffusion than oxygen delivery and metabolism in patients with traumatic brain injury,” Crit. Care Med., 36 (6), 1917 –1924 https://doi.org/10.1097/CCM.0b013e3181743d77 CCMDC7 0090-3493 (2008). Google Scholar

43. 

S. Ogoh et al., “A decrease in spatially resolved near-infrared spectroscopy-determined frontal lobe tissue oxygenation by phenylephrine reflects reduced skin blood flow,” Anesth. Analg., 118 (4), 823 –829 https://doi.org/10.1213/ANE.0000000000000145 (2014). Google Scholar

44. 

S. F. Ku et al., “2023 CAS annual meeting abstracts,” Can. J. Anesth./J. Can. Anesth., 70 (S2), 224 –226 https://doi.org/10.1007/s12630-023-02651-8 (2023). Google Scholar

45. 

M. M. Wu et al., “Complete head cerebral sensitivity mapping for diffuse correlation spectroscopy using subject-specific magnetic resonance imaging models,” Biomed. Opt. Express, 13 (3), 1131 –1151 https://doi.org/10.1364/BOE.449046 BOEICL 2156-7085 (2022). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Farah Kamar, Leena Shoemaker, Rasa Eskandari, Daniel Milej, Darren Drosdowech, John M. Murkin, Keith St. Lawrence, Jason Chui, and Mamadou Diop "Assessing changes in regional cerebral hemodynamics in adults with a high-density full-head coverage time-resolved near-infrared spectroscopy device," Journal of Biomedical Optics 29(S3), S33302 (3 May 2024). https://doi.org/10.1117/1.JBO.29.S3.S33302
Received: 29 November 2023; Accepted: 12 February 2024; Published: 3 May 2024