Paper
3 December 1980 Computer Controlled Infrared Tunable Acousto-Optic Filter (TAOF) Spectrometer
P. Katzka, S. Estrin, I. C. Chang, G. Petrie
Author Affiliations +
Abstract
A TAOF combination radiometer/spectrometer covering the 2 to 5 micron range with a 7 cm-1 constant resolution across the band has been designed and constructed. The instrument is computer controlled, capable of acquiring and reducing data at a faster rate than conventional grating spectrometers and radiometers. It has been extensively evaluated in the laboratory, and has been used to acquire the spectra of a rocket in the field.
© (1980) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
P. Katzka, S. Estrin, I. C. Chang, and G. Petrie "Computer Controlled Infrared Tunable Acousto-Optic Filter (TAOF) Spectrometer", Proc. SPIE 0246, Contemporary Infrared Sensors and Instruments, (3 December 1980); https://doi.org/10.1117/12.959364