Paper
31 January 1994 Accuracy of an approximate equation used in the calculation of the attenuated total reflection (ATR)
John E. Bertie, Zhida Lan
Author Affiliations +
Proceedings Volume 2089, 9th International Conference on Fourier Transform Spectroscopy; (1994) https://doi.org/10.1117/12.166703
Event: Fourier Transform Spectroscopy: Ninth International Conference, 1993, Calgary, Canada
Abstract
We examine the accuracy of an approximate equation that is used to calculate IR multiple attenuated total reflection pATR spectra from the real and imaginary refractive index spectra, n(v) and k(v). Under typical experimental conditions, the results for benzene, methanol and water show that the approximate equation frequently gives errors greater than 10%.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John E. Bertie and Zhida Lan "Accuracy of an approximate equation used in the calculation of the attenuated total reflection (ATR)", Proc. SPIE 2089, 9th International Conference on Fourier Transform Spectroscopy, (31 January 1994); https://doi.org/10.1117/12.166703