4 March 2013 Real-time photoacoustic imaging system for clinical burn diagnosis
Taiichiro Ida, Yasushi Kawaguchi, Satoko Kawauchi, Keiichi Iwaya, Daizoh Saitoh, Shunichi Sato, Toshiaki Iwai
Author Affiliations +
Abstract
We have developed a real-time (8~30 fps) photoacoustic (PA) imaging system with a linear-array transducer for burn diagnosis. In this system, PA signals originating from blood in the noninjured tissue layer located under the injured tissue layer are detected. The phantom study showed that thin light absorbers embedded in the tissue-mimicking scattering medium at depths of > 3 mm can be imaged with high contrast. The diagnostic experiments using rat burn models showed good agreements between the injury depths (zones of stasis) indicated by PA imaging and those by histological analysis. These results demonstrate the potential usefulness of the present system for clinical burn diagnosis.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Taiichiro Ida, Yasushi Kawaguchi, Satoko Kawauchi, Keiichi Iwaya, Daizoh Saitoh, Shunichi Sato, and Toshiaki Iwai "Real-time photoacoustic imaging system for clinical burn diagnosis", Proc. SPIE 8581, Photons Plus Ultrasound: Imaging and Sensing 2013, 858109 (4 March 2013); https://doi.org/10.1117/12.2002223